[0] Title = Message = Indication for when no alarms is a possible situation = ALARM_ID_NO_ALARM [1] Title = Self Test Error Message = Call service to schedule a repair. Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up = ALARM_ID_STUCK_BUTTON_TEST_FAILED [2] Title = Self Test Error Message = Call service to schedule a repair. HD FPGA POST failure = ALARM_ID_HD_FPGA_POST_TEST_FAILED [3] Title = Self Test Error Message = Call service to schedule a repair. DG FPGA POST failure = ALARM_ID_DG_FPGA_POST_TEST_FAILED [4] Title = Self Test Error Message = Call service to schedule a repair. HD Watchdog POST failure = ALARM_ID_HD_WATCHDOG_POST_TEST_FAILED [5] Title = Self Test Error Message = Call service to schedule a repair. DG Watchdog POST failure = ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED [6] Title = Self Test Error Message = Call service to schedule a repair. UI communication POST failure. UI failed to communicate within a reasonable time after power up = ALARM_ID_UI_COMM_POST_FAILED [7] Title = Machine Error Message = Call service to schedule a repair. Real-time clock configuration error = ALARM_ID_HD_RTC_BATTERY_LOW [8] Title = Self Test Error Message = Call service to schedule a repair. HD accelerometer failed POST = ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE [9] Title = Self Test Error Message = Call service to schedule a repair. DG accelerometer failed POST = ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE [10] Title = Machine Error Message = Call service to schedule a repair. RTC or MCU timer inaccurate = ALARM_ID_HD_BLOOD_LEAK_FAULT [11] Title = Message = DG heaters failed POST = ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE [12] Title = Self Test Error Message = Call service to schedule a repair. HD firmware image integrity POST test failed = ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED [13] Title = Self Test Error Message = Call service to schedule a repair. DG firmware image integrity POST test failed = ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED [14] Title = Message = HD blood flow invalid calibration = ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION [15] Title = Message = HD dialysate flow invalid calibration = ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION [16] Title = Self Test Error Message = Call service to schedule a repair. HD alarm audio failed POST = ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE [17] Title = Self Test Error Message = Call service to schedule a repair. HD UI POST failed = ALARM_ID_HD_UI_POST_FAILED [18] Title = Calibration Error Message = Call service to schedule a repair. DG load cells invalid calibration record = ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD [19] Title = Calibration Error Message = Call service to schedule a repair. DG flow sensors invalid calibration record = ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD [20] Title = Calibration Error Message = Call service to schedule a repair. DG conductivity sensors invalid calibration record = ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD [21] Title = Calibration Error Message = Call service to schedule a repair. DG drain line volume invalid calibration record = ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD [22] Title = Calibration Error Message = Call service to schedule a repair. DG reservoirs invalid calibration record = ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD [23] Title = Calibration Error Message = Call service to schedule a repair. DG acid concentrate calibration record = ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD [24] Title = Calibration Error Message = Call service to schedule a repair. DG bicarb concentrate calibration record = ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD [25] Title = Calibration Error Message = Call service to schedule a repair. DG accelerometer invalid calibration record = ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD [26] Title = Calibration Error Message = Call service to schedule a repair. HD accelerometer invalid calibration record = ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD [27] Title = Message = HD blood flow invalid calibration record = ALARM_ID_HD_BLOOD_FLOW_INVALID_CAL_RECORD [28] Title = Message = HD dialyzer flow invalid calibration record = ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CAL_RECORD [29] Title = Calibration Error Message = Call service to schedule a repair. HD heparin force sensor invalid calibration record = ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD [30] Title = Machine Error Message = Call service to schedule a repair. HD Software fault. Software found itself in an unexpected state = ALARM_ID_HD_SOFTWARE_FAULT [31] Title = Blood Pump Error Message = Call service to schedule a repair. Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running = ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK [32] Title = Blood Pump Error Message = Call service to schedule a repair. Blood pump failed motor off check. Measured speed while commanded off = ALARM_ID_BLOOD_PUMP_OFF_CHECK [33] Title = Blood Pump Error Message = Call service to schedule a repair. Blood pump failed motor direction check. Measured vs. commanded = ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK [34] Title = Blood Pump Error Message = Call service to schedule a repair. Blood pump failed rotor speed check. Mismatch with rotor and motor speeds = ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK [35] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running = ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK [36] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis inlet pump failed motor off check. Measured speed while commanded off = ALARM_ID_DIAL_IN_PUMP_OFF_CHECK [37] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis inlet pump failed motor direction check. Measured vs. commanded = ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK [38] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds = ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK [39] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running = ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK [40] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis outlet pump failed motor off check. Measured speed while commanded off = ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK [41] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis outlet pump failed motor direction check. Measured vs. commanded = ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK [42] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds = ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK [43] Title = Machine Error Message = Call service to schedule a repair. Watchdog expired error. Watchdog was not "pet" within the appropriate time = ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAULT [44] Title = Machine Error Message = Call service to schedule a repair. User interface communication timeout = ALARM_ID_UI_COMM_TIMEOUT [45] Title = Machine Error Message = Call service to schedule a repair. Too many bad CRCs detected on received system messages = ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS [46] Title = Machine Error Message = Call service to schedule a repair. System message that required acknowledgment was not acknowledged = ALARM_ID_CAN_MESSAGE_NOT_ACKED [47] Title = Ultrafiltration Error Message = Call service to schedule a repair. Ultrafiltration rate is too high error during treatment = ALARM_ID_UF_RATE_TOO_HIGH_ERROR [48] Title = Ultrafiltration Error Message = Call service to schedule a repair. Ultrafiltration volume accuracy error during treatment = ALARM_ID_UF_VOLUME_ACCURACY_ERROR [49] Title = Machine Error Message = Call service to schedule a repair. HD FPGA communication down for too long = ALARM_ID_HD_FPGA_COMM_TIMEOUT [50] Title = Machine Error Message = Call service to schedule a repair. DG FPGA not accepting commanded valve states = ALARM_ID_VALVE_CONTROL_FAILURE [51] Title = Blood Pump Error Message = Call service to schedule a repair. Blood pump failed motor speed check. Measured vs. commanded = ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK [52] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysate inlet pump failed motor speed check. Measured vs. commanded = ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK [53] Title = Dialysate Pump Error Message = Call service to schedule a repair. Dialysate outlet pump failed motor speed check. Measured vs. commanded = ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK [54] Title = Machine Error Message = Call service to schedule a repair. HD critical data integrity check failed = ALARM_ID_HD_CRITICAL_DATA_ERROR [55] Title = Communication Failure Message = Call service to schedule a repair. DG critical data integrity check failed = ALARM_ID_DG_CRITICAL_DATA_ERROR [56] Title = Machine Error Message = Call service to schedule a repair. HD accelerometer error (no readings or FPGA reports error) = ALARM_ID_HD_ACCELEROMETER_FAILURE [57] Title = Machine Error Message = Call service to schedule a repair. DG accelerometer error (no readings or FPGA reports error) = ALARM_ID_DG_ACCELEROMETER_FAILURE [58] Title = Valve Error Message = Call service to schedule a repair. HD valve homing failed = ALARM_ID_HD_VALVE_HOMING_FAILED [59] Title = Valve Error Message = Call service to schedule a repair. HD valve transition time out = ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT [60] Title = Valve Error Message = Call service to schedule a repair. HD valve not functional = ALARM_ID_HD_VALVE_NOT_FUNCTIONAL [61] Title = Valve Error Message = Call service to schedule a repair. HD valve current out of range = ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE [62] Title = Valve Error Message = Call service to schedule a repair. HD valve position out of target = ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE [63] Title = Arterial Pressure Error Message = Call service to schedule a repair. FPGA reports failure reading arterial pressure sensor = ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT [64] Title = Venous Pressure Error Message = Call service to schedule a repair. FPGA reports failure reading venous pressure sensor = ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT [65] Title = Machine Error Message = Call service to schedule a repair. HD requests DG command with invalid parameter fault = ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT [66] Title = Message = HD blood leak sensor set point set failure = ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE [67] Title = Self Test Error Message = Call service to schedule a repair. HD blood pump occlusion self-test failure alarm = ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE [68] Title = Dialysate Recirculation Error Message = Dialysate has recirculated for too long. HD active reservoir recirculation out of range = ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE [69] Title = Message = HD blood leak sensor invalid calibration record = ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD [70] Title = Self Test Error Message = Call service to schedule a repair. HD arterial pressure self-test failure alarm = ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE [71] Title = Self Test Error Message = Call service to schedule a repair. HD venous pressure self-test failure alarm = ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE [72] Title = Message = HD blood flow meter status check self-test failure alarm = ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE [73] Title = Message = HD dialysate flow meter status check self-test failure alarm = ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE [74] Title = Self Test Error Message = Call service to schedule a repair. HD blood leak detector self-test failure alarm = ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_SEND_TIMEOUT [75] Title = Self Test Error Message = Call service to schedule a repair. HD syringe pump self-test failure alarm = ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE [76] Title = Machine Error Message = Call service to schedule a repair. HD monitored voltage is out of range = ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE [77] Title = Machine Error Message = Call service to schedule a repair. DG monitored voltage is out of range = ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE [78] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump direction (from encoder) error = ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR [79] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump direction (from controller) error = ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR [80] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump fault reported by FPGA = ALARM_ID_HD_SYRINGE_PUMP_FAULT [81] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump over travel error = ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR [82] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump DAC write failure = ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR [83] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump is running while the BP is off = ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR [84] Title = Venous Pressure Error Message = Call service to schedule a repair. HD venous pressure sensor not being read = ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR [85] Title = Pump Track Lever Not Fully Latched Message = Open front door and pull pump track lever all the way down. HD pump track lever is not fully latched = ALARM_ID_PUMP_TRACK_LATCH_OPENED [86] Title = Occlusion Sensor Error Message = Call service to schedule a repair. HD BP occlusion sensor not being read = ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR [87] Title = Message = This alarm ID is available for use = ALARM_ID_DG_HEATING_INVALID_CAL_RECORD [88] Title = Message = This alarm ID is available for use = ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE [89] Title = Occlusion Sensor Error Message = Call service to schedule a repair. HD BP occlusion sensor error = ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR [90] Title = Message = This alarm ID is available for use = ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR [91] Title = Dialysate Measurement Error Message = Call service to schedule a repair. HD No dialysate flow data receive in the last 3 seconds = ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE [92] Title = Message = This alarm ID is available for use = ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS [93] Title = Message = This alarm ID is available for use = ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE [94] Title = Message = This alarm ID is available for use = ALARM_ID_DG_PRESSURE_OUT_OF_RANGE [95] Title = Message = This alarm ID is available for use = ALARM_ID_DG_WATCHDOG_EXPIRED [96] Title = Message = This alarm ID is available for use = ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE [97] Title = Message = This alarm ID is available for use = ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE [98] Title = Battery Error Message = Call service to schedule a repair. HD battery communication fault = ALARM_ID_HD_BATTERY_COMM_FAULT [99] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump stall alarm = ALARM_ID_HD_SYRINGE_PUMP_STALL [100] Title = Self Test Error Message = Call service to schedule a repair. HD no cartridge self-test timeout = ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT [101] Title = Self Test Error Message = Call service to schedule a repair. HD dry self-test timeout = ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT [102] Title = Message = Real-time clock communication error = ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE [103] Title = Machine Error Message = Call service to schedule a repair. HD non-volatile calibration group invalid record CRC = ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID [104] Title = Air Trap Error Message = Call service to schedule a repair. Air trap level sensors reporting illegal combination of air/fluid = ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS [105] Title = Machine Error Message = Call service to schedule a repair. DG non-volatile system group invalid record CRC = ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID [106] Title = Message = This alarm ID is available for use = ALARM_ID_DG_DIALYSATE_FLOW_SENSOR_INVALID_CAL_RECORD [107] Title = Machine Error Message = Call service to schedule a repair. HD reports DG restarted fault = ALARM_ID_DG_RESTARTED_FAULT [108] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump ADC error = ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR [109] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump volume check error = ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR [110] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump speed check error = ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR [111] Title = Syringe Pump Error Message = Call service to schedule a repair. HD syringe pump not stopped in off state error = ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR [112] Title = Blood Leak Detector Error Message = Call service to schedule a repair. HD blood leak detector fault = ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_SEND_TIMEOUT [113] Title = Message = HD arterial air bubble detector self-test failure = ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_FAULT [114] Title = Air Bubble Detector Error Message = Call service to schedule a repair. HD venous air bubble detector self-test failure = ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE [115] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG temperature sensor out of range = ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE [116] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG temperature sensor ADC out of range = ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT [117] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG primary heater internal temperature out of range = ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED [118] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG primary heater cold junction temperature out of range = ALARM_ID_DG_RTC_CONFIG_ERROR [119] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG trimmer heater internal temperature out of range = ALARM_ID_DG_RTC_BATTERY_LOW [120] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG trimmer heater cold junction temperature out of range = ALARM_ID_PRE_TREATMENT_WET_PRIME_TEST_FAILURE [121] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG main primary heater voltage out of range = ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE [122] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG small primary heater voltage out of range = ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE [123] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG trimmer heater voltage out of range = ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE [124] Title = Blood Stagnant Message = Blood has sat too long in the blood tubing set. No rinseback allowed. HD end of treatment alarm (high priority - user still not acting to end treatment) = ALARM_ID_END_OF_TREATMENT_HIGH [125] Title = Blood Stagnant Message = Blood has sat too long in the blood tubing set. Treatment must end. Blood sitting too long after treatment stopped by user alarm (> 5 min) = ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK [126] Title = Dialyzer Blood Leak Detected Message = A blood leak has been detected. Treatment must end. HD blood leak detected alarm = ALARM_ID_HD_BLOOD_LEAK_DETECTED [127] Title = Venous Pressure Low Message = Venous pressure is low. Check to be sure your blue venous connections are secure and that your venous access is intact. Venous pressure too low during treatment = ALARM_ID_VENOUS_PRESSURE_LOW [128] Title = Venous Air Bubble Detected Message = Venous air bubble detected. Check for air bubbles in the venous line. If bubbles are present, tap the tubing to move the bubbles up to the air trap. If bubbles cannot be removed, treatment must end. HD venous air bubble detected alarm = ALARM_ID_HD_VENOUS_BUBBLE_DETECTED [129] Title = Venous Air Bubble Detected During Rinseback Message = Venous air bubble detected. Check for air bubbles in the venous line. If bubbles are present, tap the tubing to move the bubbles up to the air trap. If bubbles cannot be removed, treatment must end. HD venous air bubble detected rinseback alarm = ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK [130] Title = Venous Pressure High Message = Venous pressure is high. Check to be sure your blue venous connections are secure and that there are no kinks or clots in the tubing. Venous pressure too high during treatment = ALARM_ID_VENOUS_PRESSURE_HIGH [131] Title = Arterial Pressure Low Message = Arterial pressure is low. Check to be sure your red arterial connections are secure and that there are no kinks or clots in the tubing. Arterial pressure too low during treatment = ALARM_ID_ARTERIAL_PRESSURE_LOW [132] Title = Arterial Pressure High Message = Arterial pressure is low. Check to be sure your red arterial connections are secure and that your venous access is intact. Arterial pressure too high during treatment = ALARM_ID_ARTERIAL_PRESSURE_HIGH [133] Title = Fluid Leak in Dialysate Generator Message = Fluid leak detected in Dialysate Generation DG fluid leak detected alarm = ALARM_ID_DG_FLUID_LEAK_DETECTED [134] Title = Fluid Leak in Hemodialysis Delivery Device Message = Fluid leak detected in Hemodialysis Delivery device. Clean up the spill before continuing. HD fluid leak detected alarm = ALARM_ID_HD_FLUID_LEAK_DETECTED [135] Title = Excessive Acceleration Detected in the Hemodialysis Device Message = Excessive movement detected with Hemodialysis Delivery device. Confirm the device is stationary and level before continuing. HD experienced large acceleration = ALARM_ID_HD_SHOCK [136] Title = Excessive Acceleration Detected in the Dialysate Generation Device Message = Excessive movement detected with Dialysis Generation device. Confirm the device is stationary and level before continuing. DG experienced large acceleration = ALARM_ID_DG_SHOCK [137] Title = Excessive Tilt Detected in the Hemodialysis Device Message = Excessive tilt detected with Hemodialysis Delivery device. Confirm the device is stationary and level before continuing. HD tilt exceeds maximum = ALARM_ID_HD_EXCESSIVE_TILT [138] Title = Excessive Tilt Detected in the Dialysate Generation Device Message = Excessive tilt detected with Dialysis Generation device. Confirm the device is stationary and level before continuing. DG tilt exceeds maximum = ALARM_ID_DG_EXCESSIVE_TILT [139] Title = Power Lost Message = Power has been lost. Check all electrical connections. The display and CPU will run off of the battery until power returns. HD AC power lost alarm = ALARM_ID_HD_AC_POWER_LOST [140] Title = Communication Failure with Dialysate Generator Message = Communication failure with Dialysis Generator. Ensure all cords are securely connected. Dialysate generator communication timeout = ALARM_ID_DG_COMM_TIMEOUT [141] Title = Air Trap Unable to Fill Message = Failure to fill the Air Trap. Check the air trap vent line and the transducer protector. If the transducer protector is wet, replace it. Air trap fill timeout during treatment = ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT [142] Title = Blood Pump Occlusion Detected Message = Occlusion detected after the blood pump. Check for clots or kinks around the dialyzer and in the venous line. Blood pump occlusion detected = ALARM_ID_OCCLUSION_BLOOD_PUMP [143] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG dialysate temperature sensors drift timeout. = ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE [144] Title = Message = This alarm ID is available for use = ALARM_ID_DG_RO_FLOW_SENSOR_INVALID_CAL_RECORD [145] Title = Acid Concentrate Conductivity Out of Range Message = Dialysate conductivity is out of range. Check the acid dialysate concentrates and make sure the correct concentrate is connected. DG Concentrate conductivity after adding acid out of range alarm = ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE [146] Title = Message = Dialysate conductivity out of range = ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE [147] Title = Dialysate Conductivity Error Message = Call service to schedule a repair. DG dialysate generation conductivity fault alarm = ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT [148] Title = Inlet Water Temperature High Message = Inlet water temperature is high. Check the incoming water for issues with cold water supply. Inlet water temperature too high = ALARM_ID_INLET_WATER_HIGH_TEMPERATURE [149] Title = Inlet Water Temperature Low Message = Inlet water temperature is low. Check the incoming water for issues with hot water supply. Inlet water temperature too low = ALARM_ID_HD_WATCHDOG_EXPIRED [150] Title = Inlet Water Conductivity High Message = Inlet water conductivity is high. Check the incoming water for issues. Inlet water conductivity too high = ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE [151] Title = Inlet Water Conductivity Low Message = Inlet water conductivity is low. Check the incoming water for issues. Inlet water conductivity too low = ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE [152] Title = Inlet Water Pressure Low Message = Inlet water pressure is low. Check the incoming water supply. Inlet water pressure too low = ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE [153] Title = Message = HD prime completed high priority alarm = ALARM_ID_PRIME_COMPLETED_HIGH [154] Title = Communication Failure Message = Communication has been lost. Ensure all cords are securely connected. EEPROM operations (read write erase) failure write erase) failure = ALARM_ID_NVDATA_EEPROM_OPS_FAILURE [155] Title = Message = This alarm ID is available for use = ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE [156] Title = Message = HW usage data (treatment time in HD and total consumed water in DG) failure = ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT [157] Title = Message = This alarm ID is available for use = ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT [158] Title = Message = Blood pump failed flow vs. motor speed check. = ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE [159] Title = Dialysate Pump Failure Message = Dialysate pump failure. Check there are no kinks in the dialysate tubing. Dialysate inlet pump failed flow vs. motor speed check. = ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK [160] Title = Blood Pump Speed High Message = Blood pump failure. Check the pump rotor for secure connection. Blood pump rotor speed too high = ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH [161] Title = Message = Blood flow sensor signal strength too low TODO remove this alarm when FMB fully removed = ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS [162] Title = Message = Dialysate flow sensor signal strength too low TODO remove this alarm when FMD fully removed = ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE [163] Title = Ultrafiltration Measurement Error Message = Ultrafiltration error. HD sees primary load cell for reservoir 1 change too much too fast = ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM [164] Title = Ultrafiltration Measurement Error Message = Ultrafiltration error. HD sees primary load cell for reservoir 2 change too much too fast = ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM [165] Title = Rinseback Timeout Message = Maximum Rinseback time exceeded. Rinseback must end. HD in treatment rinseback operation timeout = ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM [166] Title = Recirculation Timeout Message = Maximum Recirculation time exceeded. Treatment must end. HD in treatment re-circ sub-mode for too long = ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM [167] Title = Door Open Message = Hemodialysis Delivery device door open. HD cartridge door opened alarm = ALARM_ID_CARTRIDGE_DOOR_OPENED [168] Title = Dialysate Recirculation Error Message = Dialysate recirculation error. HD active reservoir depletion timeout. = ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT [169] Title = Dialysate Generation Error Message = Call service to schedule a repair. Dialysate flow rate out of range = ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE [170] Title = Syringe Empty Message = Heparin syringe empty. Option to resume treatment without heparin. HD syringe empty alarm = ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY [171] Title = Heparin Occlusion Message = Heparin pump occlusion detected. Check heparin line is unclamped and there are no kinks in the heparin line. HD syringe pump occlusion alarm = ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION [172] Title = Not Enough Heparin Message = Not enough heparin is available for treatment. Check if syringe pump is touching syringe plunger. If syringe pump is touching syringe plunger, end treatment. If syringe pump is not touching syringe plunger, reload syringe with enough heparin. HD syringe pump not enough Heparin alarm = ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM [173] Title = Message = HD arterial air bubble detected alarm = ALARM_ID_HD_RTC_CONFIG_ERROR [174] Title = Message = HD arterial air bubble detected rinseback alarm = ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE [175] Title = Blood Pump Error Message = Blood pump eailure. Check the pump rotor for secure connection. HD pump direction status error = ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR [176] Title = Message = // TODO remove this alarm HD active reservoir is full during treatment and DG not ready to switch reservoirs. = ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH [177] Title = Machine Error Message = Call service to schedule a repair. DG software fault. Software found itself in an unexpected state = ALARM_ID_DG_SOFTWARE_FAULT [178] Title = Message = HD communication timeout = ALARM_ID_HD_COMM_TIMEOUT [179] Title = Communication Failure Message = Call service to schedule a repair. DG FPGA communication down for too long = ALARM_ID_DG_FPGA_COMM_TIMEOUT [180] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. DG load cell ADC error = ALARM_ID_DG_LOAD_CELL_ADC_ERROR [181] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. DG load cells weight out of range for tare = ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE [182] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. DG load cells invalid calibration = ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD [183] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. DG invalid load cell value = ALARM_ID_DG_INVALID_LOAD_CELL_VALUE [184] Title = Machine Error Message = Call service to schedule a repair. Inlet or Outlet UV reactor not healthy = ALARM_ID_UV_REACTOR_NOT_HEALTHY [185] Title = Fan Error Message = Fan failure. Check that the fans inlets are not blocked or obstructed. DG fan RPM out of range = ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE [186] Title = Dialysate Concentrate Pump Error Message = Call service to schedule a repair. DG concentrate pump fault = ALARM_ID_DG_CONCENTRATE_PUMP_FAULT [187] Title = Dialysate Concentrate Pump Error Message = Call service to schedule a repair. DG concentrate pump CP1 speed control error = ALARM_ID_CP1_SPEED_CONTROL_ERROR [188] Title = Dialysate Concentrate Pump Error Message = Call service to schedule a repair. DG concentrate pump CP2 speed control error = ALARM_ID_CP2_SPEED_CONTROL_ERROR [189] Title = Dialysate Drain Pump Error Message = Call service to schedule a repair. DG drain pump RPM out of range = ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE [190] Title = Dialysate Drain Pump Error Message = Call service to schedule a repair. DG drain pump off fault = ALARM_ID_DRAIN_PUMP_OFF_FAULT [191] Title = Dialysate Flowrate Error Message = Call service to schedule a repair. DG flow rate out of upper range = ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE [192] Title = Dialysate Flowrate Error Message = Call service to schedule a repair. DG flow rate out of lower range = ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE [193] Title = Message = RO pump flow rate out of range = ALARM_ID__AVAILABLE_13 [194] Title = RO Pump Error Message = Call service to schedule a repair. DG RO pump duty cycle out of range = ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE [195] Title = RO Pump Error Message = Call service to schedule a repair. DG RO pump pressure out of range = ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE [196] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG temperature sensor error flag fault. = ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT [197] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG temperature sensors ADC fault = ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT [198] Title = Dialysate Heater Error Message = Call service to schedule a repair. DG heaters cold junction temperature out of range = ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON [199] Title = Message = DG heaters fault = ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON [200] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG thermistors/sensors temperature out of range = ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE [201] Title = Inlet Water Error Message = Inlet water quality is bad. Inlet water quality is bad (temperature conductivity pressure) conductivity pressure) = ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE [202] Title = Inlet Water Error Message = Inlet water quality is bad. Inlet water pressure fault = ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE [203] Title = Dialysate Pressure Error Message = Call service to schedule a repair. Pressure sensor fault = ALARM_ID_PRESSURE_SENSOR_FAULT [204] Title = RO Rejection Error Message = Call service to schedule a repair. RO rejection ratio out of range = ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE [205] Title = Dialysate Conductivity Error Message = Call service to schedule a repair. Conductivity sensor fault = ALARM_ID_CONDUCTIVITY_SENSOR_FAULT [206] Title = Dialysate Production Error Message = Call service to schedule a repair. DG dialysate fill runs out of time = ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME [207] Title = Dialysate Flowrate Error Message = Call service to schedule a repair. DG flow meter check failure alarm = ALARM_ID_DG_FLOW_METER_CHECK_FAILURE [208] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG primary heater on with no flow time out = ALARM_ID_HD_VENOUS_LINE_OCCLUSION [209] Title = Dialysate Production Error Message = Call service to schedule a repair. DG drain circulation line timeout = ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT [210] Title = Battery Error Message = Battery pack error. HD battery pack detected an error = ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED [211] Title = Blood Stagnant Message = Maximum time for blood stoppage has been exceeded. Blood sitting too long warning (> 4 min) = ALARM_ID_BLOOD_SITTING_WARNING [212] Title = Blood Stagnant Message = Maximum time for blood stoppage has been exceeded. Blood sitting too long warning (> 4 min) = ALARM_ID_END_OF_TREATMENT_ALARM [213] Title = Message = HD prime completed medium priority alarm = ALARM_ID_PRIME_COMPLETED_MEDIUM [214] Title = Saline Bolus Error Message = Inability to deliver saline bolus. Check for kinks or clamped lines in the saline line. HD Saline bolus volume check failure = ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE [215] Title = Rinseback Error Message = Unable to complete Rinseback. Check for kinks or clamped blood lines. HD rinseback volume check failure = ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE [216] Title = Treatment Ended Message = Treatment has ended HD end treatment sub-mode timeout alarm = ALARM_ID_END_TREATMENT_TIMEOUT_ALARM [217] Title = Prime Error Message = Blood prime error. Check for kinks or clamped blood lines. HD blood prime volume check failure = ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE [218] Title = Syringe Not Detected Message = Heparin syringe not detected. HD syringe detected alarm = ALARM_ID_HD_SYRINGE_DETECTED [219] Title = Syringe Removed Message = Heparin syringe removed. HD syringe pump syringe removed alarm = ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED [220] Title = Battery Error Message = Battery pack error. HD battery pack SOC is too low. = ALARM_ID_HD_BATTERY_PACK_CHARGE_TOO_LOW [221] Title = Saline Bag Empty Message = Saline back appears empty. Please check saline bag and line and replace if saline bag is empty. HD Saline bag is empty = ALARM_ID_EMPTY_SALINE_BAG [222] Title = Dialysate Temperature High Message = Dialysate temperature high. HD dialysate temperature too high alarm = ALARM_ID_DIALYSATE_TEMPERATURE_HIGH [223] Title = Dialysate Temperature Low Message = Dialysate temperature low. HD dialysate temperature too low alarm = ALARM_ID_DIALYSATE_TEMPERATURE_LOW [224] Title = Treatment Stopped Message = Treatment paused. Treatment stopped by user action - pressed stop button = ALARM_ID_TREATMENT_STOPPED_BY_USER [225] Title = End of Treatment Message = Treatment has ended HD end of treatment warning = ALARM_ID_END_OF_TREATMENT_WARNING [226] Title = Prime Complete Message = Prime of the bloodlines is complete. HD prime completed low priority alarm = ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY [227] Title = Maximum Prime Time Exceeded Message = Maximum time for blood prime has been exceeded. HD prime out of time alarm = ALARM_ID_PRIME_OUT_OF_TIME [228] Title = Maximum Prime Time Exceeded Message = Priming failure. Check blood and dialysate lines for clamps and kinks. HD time out on prime sub-mode purge air state. = ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT [229] Title = Maximum Prime Time Exceeded Message = Priming failure. Check blood and dialysate lines for clamps and kinks. HD prime dialysate dialyzer time out alarm = ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT [230] Title = Maximum Prime Time Exceeded Message = Priming failure. Check blood and dialysate lines for clamps and kinks. HD prime dialysate bypass time out alarm = ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT [231] Title = Self Test Error Message = Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. HD pre-treatment mode dry self-tests failure = ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE [232] Title = Self Test Error Message = Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. Check for closed clamps or kinks in the tubing. HD pre-treatment mode wet self-tests failure = ALARM_ID_PRE_TREATMENT_WET_LC_TEST_FAILURE [233] Title = Battery Error Message = Battery pack error. RTC battery low = ALARM_ID__AVAILABLE_8 [234] Title = Communication Failure Message = Communication failure. Ensure all cords are securely connected. RTC (i.e read time) or RAM operations failure (read or write) = ALARM_ID_RTC_RAM_OPS_ERROR [235] Title = Rinseback Complete Message = Rinseback has been completed. HD in treatment stopped sub-mode after rinseback completed (no escalation) = ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK [236] Title = New Cartridge Needed Message = New cartridge needs to be installed. HD needs new cartridge to be installed = ALARM_ID_INSTALL_NEW_CARTRIDGE [237] Title = Message = HD prime saline dialyzer time out alarm = ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT [238] Title = Cartridge Needed Message = Cartridge not loaded or loaded improperly. Check that a cartridge is properly installed. HD no cartridge loaded or installed improperly alarm = ALARM_ID_NO_CARTRIDGE_LOADED [239] Title = Cartridge Removal Error Message = Cartridge not unloaded, or unloaded improperly. Check that the cartridge has been removed. HD fail to remove cartridge alarm = ALARM_ID_CARTRIDGE_REMOVAL_FAILURE [240] Title = Dialysate Conductivity Error Message = Dialysate conductivity out of range. Check the bicarbonate dialysate concentrates and make sure the correct concentrate is connected. Bicarb conductivity out of range during bicarb pump check alarm = ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE [241] Title = Dialysate Drain Pump Error Message = Call service to schedule a repair. DG reservoir drain time out = ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT [242] Title = Dialysate Production Error Message = Call service to schedule a repair. DG reservoir fill time out = ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT [243] Title = Dialysate Production Error Message = Call service to schedule a repair. DG reservoir leak time out = ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT [244] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG temperature sensors difference out of range = ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE [245] Title = Heat Disinfection Error Message = Call service to schedule a repair. DG heat disinfect target temperature time out (could not reach to temperature) = ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT [246] Title = Heat Disinfection Error Message = Call service to schedule a repair. DG heat disinfect inlet pressure and temperature sensors out of range = ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT [247] Title = Heat Disinfection Error Message = Call service to schedule a repair. DG heat disinfect inlet conductivity and temperature out of range = ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT [248] Title = Chemical Disinfection Error Message = Call service to schedule a repair. DG chemical disinfect could not reach to target temperature = ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT [249] Title = Chemical Disinfection Error Message = Call service to schedule a repair. DG chemical disinfect inlet pressure and temperature sensors out of range = ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT [250] Title = Chemical Disinfection Error Message = Call service to schedule a repair. DG chemical disinfect inlet conductivity and temperature out of range = ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT [251] Title = Machine Error Message = Call service to schedule a repair. HD invalid system record = ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC [252] Title = Machine Error Message = Call service to schedule a repair. HD invalid service record = ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC [253] Title = Machine Error Message = Call service to schedule a repair. DG invalid system record = ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC [254] Title = Machine Error Message = Call service to schedule a repair. DG invalid service record = ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC [255] Title = Machine Error Message = Call service to schedule a repair. HD and UI software builds are not compatible = ALARM_ID_HD_UI_COMPATIBILITY_ERROR [256] Title = Message = HD and DG software builds are not compatible = ALARM_ID_HD_DG_COMPATIBILITY_ERROR [257] Title = Machine Error Message = Call service to schedule a repair. DG FPGA power out timeout = ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT [258] Title = Message = HD FPGA power out timeout = ALARM_ID_HD_FPGA_POWER_OUT_TIMEOUT [259] Title = Dialysate Temperature Error Message = Call service to schedule a repair. HD temperatures out of range = ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE [260] Title = Message = UI POST Application Integrity (Sha256Sum) failure = ALARM_ID_UI_POST_FAILURE_SHASUM [261] Title = Message = UI POST CANBus failure = ALARM_ID_UI_POST_FAILURE_CANBUS [262] Title = Message = UI POST Display failure = ALARM_ID_UI_POST_FAILURE_DISPLAY [263] Title = Message = UI POST Touch failure = ALARM_ID_UI_POST_FAILURE_TOUCH [264] Title = Message = UI POST SD-Card failure = ALARM_ID_UI_POST_FAILURE_SDCARD [265] Title = Message = UI POST RTC failure = ALARM_ID_UI_POST_FAILURE_RTC [266] Title = Message = UI POST WiFi failure = ALARM_ID_UI_POST_FAILURE_WIFI [267] Title = Message = UI POST Bluetooth failure = ALARM_ID_UI_POST_FAILURE_BLUETOOTH [268] Title = Message = UI POST Ethernet failure = ALARM_ID_UI_POST_FAILURE_ETHERNET [269] Title = Message = UI POST Sound failure = ALARM_ID_UI_POST_FAILURE_SOUND [270] Title = Self Test Error Message = Call service to schedule a repair. HD POST Safety Shutdown failure = ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED [271] Title = Self Test Error Message = Call service to schedule a repair. DG POST Safety Shutdown failure = ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED [272] Title = Message = Fan failure. Check that the fans inlets are not blocked or obstructed. HD Fan RPM out of range = ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE [273] Title = Blood Pump Error Message = Call service to schedule a repair. HD measured blood pump flow rate is out of range = ALARM_ID_HD_BLOOD_FLOW_OUT_OF_RANGE [274] Title = Dialysate Pump Error Message = Call service to schedule a repair. HD measured dialysate inlet pump flow rate is out of range = ALARM_ID_HD_DIAL_IN_FLOW_OUT_OF_RANGE [275] Title = Arterial Pressure Error Message = Call service to schedule a repair. HD arterial pressure sensor is reading out of range = ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE [276] Title = Venous Pressure Error Message = Call service to schedule a repair. HD venous pressure sensor is reading out of range = ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE [277] Title = Occlusion Sensor Error Message = Call service to schedule a repair. HD BP occlusion sensor is reading out of range = ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE [278] Title = Dialysate Production Error Message = Ultrafiltration error. HD active reservoir weight out of range. = ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE [279] Title = Dialysate Drain Pump Error Message = DG dialysate drain time out = ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT [280] Title = Arterial Pressure Error Message = Call service to schedule a repair. HD arterial pressure sensor read timeout error = ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR [281] Title = Acid Concentrate Low Message = Acid concentrate low. Replace acid concentrate bottle. DG Acid concentration bottle low volume alarm = ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME [282] Title = Bicarbonate Concentrate Low Message = Bicarbonate concentrate low. Replace bicarbonate concentrate bottle. DG Bicarbonate concentration bottle low volume alarm = ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME [283] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. DG load cells weight out of range = ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE [284] Title = Message = DG load cells primary/back up drift out of range = ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE [285] Title = Dialysate Recirculation Error Message = Dialysate recirculation error. HD treatment recirculate timeout warning = ALARM_ID_HD_TREATMENT_RECIRC_TIMEOUT_WARNING [286] Title = Rinseback Complete Message = Rinseback complete. HD treatment rinseback complete timeout warning = ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_WARNING [287] Title = Machine Error Message = Call service to schedule a repair. HD processor clock speed checks against FPGA clock failure = ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE [288] Title = Dialysate Temperature Error Message = Call service to schedule a repair. DG trimmer heater on with flow timeout. = ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE [289] Title = DG Cap State Error Message = Check dialysate/concentrate caps. DG cap in wrong state = ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION [290] Title = Message = = ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON