[0] Title = Indication for when no alarms is a possible situation Message = [1] Title = HD stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up Message = HD [2] Title = HD FPGA POST failure Message = HD [3] Title = DG FPGA POST failure Message = DG SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = HD Watchdog POST failure Message = HD [5] Title = DG Watchdog POST failure Message = DG SRSDG 53 = [6] Title = HD UI communication POST failure. UI failed to communicate within a reasonable time after power up Message = HD [7] Title = HD RTC battery low Message = HD [8] Title = HD accelerometer failed POST Message = HD [9] Title = DG accelerometer failed POST Message = DG SRSDG 323 = [10] Title = HD blood leak sensor setting embedded mode failure. Message = HD [11] Title = HD dialysate temperature below target temperature or below safety temp Message = HD [12] Title = HD firmware image integrity POST test failed Message = HD [13] Title = DG firmware image integrity POST test failed Message = DG SRSDG 813 = [14] Title = DG invalid usage record CRC Message = DG SRSDG 957 = [15] Title = HD dialysate temperature above high safety. Message = HD [16] Title = HD alarm audio failed POST Message = HD [17] Title = HD UI POST failed Message = HD [18] Title = HD didn't get ACK on message to DG that required acknowledgment Message = HD [19] Title = HD dialysate temperature above target temperature. Message = HD [20] Title = DG conductivity sensors invalid calibration record Message = DG SRSDG 16 = [21] Title = DG drain line volume invalid calibration record Message = DG SRSDG 936 = [22] Title = DG reservoirs invalid calibration record Message = DG SRSDG 935 = [23] Title = DG acid concentrate calibration record Message = DG SRSDG 934 = [24] Title = DG bicarb concentrate calibration record Message = DG SRSDG 933 = [25] Title = DG accelerometer invalid calibration record Message = DG SRSDG 817 = [26] Title = HD accelerometer invalid calibration record Message = HD [27] Title = HD blood leak sensor zero and self test sequence failed Message = HD [28] Title = DG two wire sensors FPGA fault Message = DG SRSDG 864 = [29] Title = HD heparin force sensor invalid calibration record Message = HD [30] Title = HD Software fault. Software found itself in an unexpected state Message = HD [31] Title = HD blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running Message = HD [32] Title = HD blood pump failed motor off check. Measured speed while commanded off Message = HD [33] Title = HD blood pump failed motor direction check. Measured vs. commanded Message = HD [34] Title = HD blood pump failed rotor speed check. Mismatch with rotor and motor speeds Message = HD [35] Title = HD dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running Message = HD [36] Title = HD dialysis inlet pump failed motor off check. Measured speed while commanded off Message = HD [37] Title = HD dialysis inlet pump failed motor direction check. Measured vs. commanded Message = HD [38] Title = HD dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds Message = HD [39] Title = HD dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running Message = HD [40] Title = HD dialysis outlet pump failed motor off check. Measured speed while commanded off Message = HD [41] Title = HD dialysis outlet pump failed motor direction check. Measured vs. commanded Message = HD [42] Title = HD dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds Message = HD [43] Title = DG main primary heater FPGA fault Message = DG SRSDG 954 = [44] Title = HD UI communication timeout Message = HD [45] Title = HD too many bad communications CRC Message = HD [46] Title = HD didn't get ACK on message to UI that required acknowledgment Message = HD [47] Title = Available for use Message = [48] Title = HD ultrafiltration volume accuracy error during treatment Message = HD [49] Title = HD FPGA communication down for too long Message = HD [50] Title = DG FPGA not accepting commanded valve states Message = DG SRSDG 26 = [51] Title = HD blood pump failed motor speed check. Measured vs. commanded Message = HD [52] Title = HD dialysate inlet pump failed motor speed check. Measured vs. commanded Message = HD [53] Title = HD dialysate outlet pump failed motor speed check. Measured vs. commanded Message = HD [54] Title = HD critical data integrity check failed Message = HD [55] Title = DG critical data integrity check failed Message = DG SRSDG 921 = [56] Title = HD accelerometer error (no readings or FPGA reports error) Message = HD [57] Title = DG accelerometer error (no readings or FPGA reports error) Message = DG SRSDG 869 SRSDG 870 = [58] Title = HD valve homing failed Message = HD [59] Title = HD valve transition time out Message = HD [60] Title = HD valve not functional Message = HD [61] Title = HD valve current out of range Message = HD [62] Title = HD valve position out of target Message = HD [63] Title = HD blood leak FPGA fault Message = HD [64] Title = DG barometric pressure sensor out of range Message = DG SRSDG 874 = [65] Title = HD DG requests DG command with invalid parameter fault Message = HD [66] Title = HD blood leak sensor set point set failure Message = HD [67] Title = HD blood pump occlusion self-test failure alarm Message = HD [68] Title = HD active reservoir recirculation out of range Message = HD [69] Title = HD blood leak sensor invalid calibration record Message = HD [70] Title = HD arterial pressure self-test failure alarm Message = HD [71] Title = HD venous pressure self-test failure alarm Message = HD [72] Title = HD No load cell data message receive at least once every 2 seconds Message = HD [73] Title = HD No dialysate temperature data message receive at least once every 2 seconds Message = HD [74] Title = DG inlet UV reactor is on with no flow Message = DG SRSDG 600 = [75] Title = HD syringe pump self-test failure alarm Message = HD [76] Title = HD monitored voltage is out of range Message = HD [77] Title = DG monitored voltage is out of range Message = DG SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = HD syringe pump direction (from encoder) error Message = HD [79] Title = HD syringe pump direction (from controller) error Message = HD [80] Title = HD syringe pump fault reported by FPGA Message = HD [81] Title = HD syringe pump over travel error Message = HD [82] Title = HD syringe pump DAC write failure Message = HD [83] Title = HD syringe pump is running while the BP is off Message = HD [84] Title = DG set RTC year is invalid Message = DG SRSDG 815 = [85] Title = HD pump track latch opened alarm Message = HD [86] Title = HD set RTC year is invalid Message = HD [87] Title = DG heating invalid calibration record Message = DG SRSDG 929 = [88] Title = DG concentrate pumps hall sensor out of range Message = DG [89] Title = DG outlet UV reactor on with no flow Message = DG SRSDG 599 = [90] Title = DG load cells A1/B1 FPGA fault Message = DG SRSDG 604 = [91] Title = HD No dialysate flow data receive in the last 3 seconds Message = HD [92] Title = Message = [93] Title = DG outlet primary conductivity out of range Message = DG SRSDG 430 = [94] Title = DG pressure out of range Message = DG SRSDG 629 = [95] Title = DG watchdog expired Message = DG SRSDG 907 = [96] Title = DG inlet water temperature in high range Message = DG SRSDG 410 = [97] Title = DG fill conductivity out of range Message = DG SRSDG 397 = [98] Title = HD battery communication fault Message = HD [99] Title = HD syringe pump stall alarm Message = HD [100] Title = HD no cartridge self-test timeout Message = HD [101] Title = UI POST HD communication failure Message = UI [102] Title = DG heat disinfect temperature gradient out of range Message = DG SRSDG 856 SRSDG 930 = [103] Title = HD invalid calibration CRC Message = HD [104] Title = HD air trap level sensors reporting illegal combination of air/fluid Message = HD [105] Title = DG invalid calibration CRC Message = DG SRSDG 608 = [106] Title = DG dialysate flow sensor invalid calibration record Message = DG SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = HD reports DG restarted fault Message = HD [108] Title = HD syringe pump ADC FPGA fault Message = HD [109] Title = HD syringe pump volume check error Message = HD [110] Title = HD syringe pump speed check error Message = HD [111] Title = HD syringe pump not stopped in off state error Message = HD [112] Title = DG load cells A1/B1 FPGA fault Message = DG SRSDG 953 = [113] Title = DG Drain pump current out of range Message = DG SRSDG 890 = [114] Title = HD venous air bubble detector self-test failure Message = HD [115] Title = DG temperature sensor out of range Message = DG SRSDG 224 = [116] Title = HD UI SD card failure Message = UI HD [117] Title = DG CAN message not acked Message = DG SRSDG 909 = [118] Title = DG RTC config error Message = DG SRSDG 43 = [119] Title = DG RTC battery low error Message = DG SRSDG 908 = [120] Title = HD pre-treatment mode wet self-test prime check failure Message = HD [121] Title = DG main primary heater voltage out of range Message = DG SRSDG 611 = [122] Title = DG small primary heater voltage out of range Message = DG SRSDG 612 = [123] Title = DG trimmer heater voltage out of range Message = DG SRSDG 613 = [124] Title = HD end of treatment alarm (high priority - user still not acting to end treatment) Message = HD [125] Title = HD blood sitting too long after treatment stopped by user alarm (>5 min) Message = HD [126] Title = HD blood leak detected alarm Message = HD [127] Title = HD venous pressure too low during treatment Message = HD [128] Title = HD venous air bubble detected alarm Message = HD [129] Title = Available for use Message = [130] Title = HD venous pressure too high during treatment Message = HD [131] Title = HD arterial pressure too low during treatment Message = HD [132] Title = HD arterial pressure too high during treatment Message = HD [133] Title = DG fluid leak detected alarm Message = DG SRSDG 328 = [134] Title = HD fluid leak detected alarm Message = HD [135] Title = HD experienced large acceleration Message = HD [136] Title = DG experienced large acceleration Message = DG SRSDG 327 = [137] Title = HD tilt exceeds maximum Message = HD [138] Title = DG tilt exceeds maximum Message = DG SRSDG 326 = [139] Title = HD AC power lost alarm Message = HD [140] Title = HD DG communication timeout Message = HD [141] Title = HD air trap fill timeout during treatment Message = HD [142] Title = HD blood pump occlusion detected Message = HD [143] Title = DG dialysate temperature sensors out of range Message = DG SRSDG 912 = [144] Title = DG cleaning mode inlet water conductivity out of high range. Message = DG SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] Title = DG concentrate conductivity after adding acid out of range alarm Message = DG SRSDG 846 = [146] Title = DG RTC or timer accuracy failure. Message = DG SRSDG 312 = [147] Title = DG wait for the DG to produce dialysate Message = DG SRSDG 398 = [148] Title = DG Cleaning mode inlet water temperature too high Message = DG SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = HD watchdog expired Message = HD [150] Title = DG inlet water conductivity in high range Message = DG SRSDG 391 = [151] Title = DG inlet water conductivity in low range Message = DG SRSDG 826 = [152] Title = DG inlet water pressure in low range Message = DG SRSDG 417 = [153] Title = HD prime completed high priority alarm Message = HD [154] Title = DG cleaning mode inlet water conductivity out of low range. Message = DG SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] Title = HD No reservoirs data message receive at least once every 2 seconds Message = HD [156] Title = HD No DG operation mode message receive at least once every 2 seconds Message = HD [157] Title = DG chemical disinfect prime acid line timeout. Message = DG SRSDG 988 = [158] Title = DG inlet water temperature is in the low range Message = DG SRSDG 408 = [159] Title = DG chemical disinfect flush remove acid and close the concentrate cap. Message = DG SRSDG 1044 = [160] Title = HD blood pump rotor speed too high Message = HD [161] Title = DG comm too many bad CRCs Message = DG SRSDG 122 = [162] Title = DG FPGA clock speed check failure. Message = DG SRSDG 867 = [163] Title = HD sees primary load cell for reservoir 1 change too much too fast Message = HD [164] Title = HD sees primary load cell for reservoir 2 change too much too fast Message = HD [165] Title = HD in treatment rinseback operation timeout Message = HD [166] Title = Available for use Message = [167] Title = HD cartridge door opened alarm Message = HD [168] Title = Available for use Message = [169] Title = DG Dialysate flow rate out of maximum range Message = DG SRSDG 643 = [170] Title = HD syringe empty alarm Message = HD [171] Title = HD syringe pump occlusion alarm Message = HD [172] Title = HD syringe pump not enough Heparin alarm Message = HD [173] Title = HD RTC configuration error Message = HD [174] Title = HD RTC or timer accuracy failure Message = HD [175] Title = HD pump direction status error Message = HD [176] Title = HD dialysate temperature above high safety range Message = [177] Title = DG software fault. Software found itself in an unexpected state Message = DG SRSDG 919 = [178] Title = HD communication timeout Message = UI = SRSUI 1066 [179] Title = DG FPGA communication down for too long Message = DG SRSDG 605 = [180] Title = DG RO flow out of maximum range Message = DG SRSDG 883 = [181] Title = DG load cells weight out of range for tare Message = DG SRSDG 849 SRSDG 838 = [182] Title = DG load cells invalid calibration Message = DG SRSDG 162 = [183] Title = DG invalid load cell value Message = DG SRSDG 851 = [184] Title = DG inlet UV reactor not healthy Message = DG SRSDG 559 = [185] Title = DG fan RPM out of range Message = DG SRSDG 628 = [186] Title = DG concentrate pump fault Message = DG SRSDG 891 = [187] Title = DG concentrate pump CP1 speed control error Message = DG SRSDG 905 = [188] Title = DG concentrate pump CP2 speed control error Message = DG SRSDG 906 = [189] Title = DG drain pump RPM out of range Message = DG SRSDG 448 = [190] Title = DG drain pump off fault Message = DG [191] Title = DG flow rate out of upper range Message = DG SRSDG 885 = [192] Title = DG flow rate out of lower range Message = DG SRSDG 945 = [193] Title = HD dialysate temperature out of low safety range. Message = [194] Title = DG RO pump duty cycle out of range Message = DG SRSDG 881 = [195] Title = DG RO pump pressure out of range Message = DG SRSDG 925 = [196] Title = DG CPi/CPo sensors FPGA fault Message = DG SRSDG 955 = [197] Title = DG CD1/CD2 sensors FPGA fault Message = DG SRSDG 956 = [198] Title = DG RO flow too low while primary heater is on Message = DG SRSDG 872 = [199] Title = DG Dialysate flow too low while trimmer heater is on Message = DG SRSDG 871 = [200] Title = DG thermistors/sensors temperature out of range Message = DG SRSDG 178 = [201] Title = HD pre-treatment mode wet self-test lc vs FMD failure Message = HD [202] Title = HD pre-treatment mode dry pressure normal self-test failure Message = HD [203] Title = Available for use Message = [204] Title = DG RO rejection ratio out of range Message = DG SRSDG 429 = [205] Title = DG conductivity sensor fault Message = DG SRSDG 913 = [206] Title = DG dialysate fill runs out of time Message = DG SRSDG 437 = [207] Title = DG flow meter check failure alarm Message = DG SRSDG 240 = [208] Title = HD venous pressure occlusion alarm Message = HD [209] Title = DG drain circulation line timeout Message = DG SRSDG 980 = [210] Title = HD battery pack detected an error Message = HD [211] Title = HD blood sitting too long warning (>4 min) Message = HD [212] Title = HD end of treatment alarm (user not acting to end treatment) Message = HD [213] Title = HD prime completed medium priority alarm Message = HD [214] Title = Available for use Message = [215] Title = Available for use Message = [216] Title = HD end treatment sub-mode timeout alarm Message = HD [217] Title = Available for use Message = [218] Title = HD syringe detected alarm Message = HD [219] Title = HD syringe pump syringe removed alarm Message = HD [220] Title = DG THd sensors FPGA fault Message = DG SRSDG 947 = [221] Title = HD Saline bag is empty Message = HD [222] Title = HD occlusion sensor FPGA fault Message = HD [223] Title = HD arterial sensor FPGA fault Message = HD [224] Title = HD treatment stopped by user action - pressed stop button Message = HD [225] Title = HD end of treatment warning Message = HD [226] Title = HD prime completed low priority alarm Message = HD [227] Title = Available for use Message = [228] Title = HD time out on prime saline purge air state. Message = HD [229] Title = HD prime dialysate dialyzer time out alarm Message = HD [230] Title = HD prime dialysate bypass time out alarm Message = HD [231] Title = HD pre-treatment mode dry pressure self-test failure Message = HD [232] Title = HD pre-treatment mode wet self-test LC vs. LC failure Message = HD [233] Title = DG barometric sensor coefficients bad CRC Message = DG SRSDG 964 SRSDG 949 SRSDG 1089 = [234] Title = DG inlet water pressure in high range. Message = DG SRSDG 1037 = [235] Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) Message = HD [236] Title = HD needs new cartridge to be installed Message = HD [237] Title = HD prime saline dialyzer time out alarm Message = HD [238] Title = HD no cartridge loaded or installed improperly alarm Message = HD [239] Title = HD fail to remove cartridge alarm Message = HD [240] Title = DG bicarb conductivity out of range during bicarb pump check alarm Message = DG SRSDG 482 = [241] Title = DG reservoir drain time out Message = DG SRSDG 848 SRSDG 90 SRSDG 561 SRSDG 577 SRSDG 590 SRSDG 591 SRSDG 336 SRSDG 879 SRSDG 340 SRSDG 496 SRSDG 501 SRSDG 505 SRSDG 532 SRSDG 535 SRSDG 556 SRSDG 557 SRSDG 1048 SRSDG 1049 SRSDG 651 SRSDG 655 SRSDG 744 SRSDG 745 SRSDG 1004 SRSDG 1008 SRSDG 1019 SRSDG 1023 SRSDG 1033 SRSDG 1034 = [242] Title = DG reservoir fill time out Message = DG SRSDG 850 SRSDG 573 SRSDG 574 SRSDG 579 SRSDG 580 SRSDG 492 SRSDG 493 SRSDG 498 SRSDG 499 SRSDG 508 SRSDG 509 SRSDG 519 SRSDG 690 SRSDG 691 SRSDG 703 SRSDG 711 SRSDG 1020 SRSDG 1024 = [243] Title = DG reservoir leak time out Message = DG SRSDG 853 = [244] Title = DG cleaning mode temperature sensors difference out of range Message = DG SRSDG 852 = [245] Title = DG heat disinfect target temperature time out (could not reach to temperature) Message = DG SRSDG 854 = [246] Title = DG cleaning mode conductivity sensors out of range Message = DG [247] Title = Alarm Id available Message = [248] Title = DG chemical disinfect could not reach to target temperature out of range Message = DG [249] Title = DG chemical disinfect target conductivity out of range Message = DG SRSDG 992 = [250] Title = DG chemical disinfect insert acid and remove the concentrate cap Message = DG SRSDG 1046 = [251] Title = HD invalid system record Message = HD [252] Title = HD invalid service record Message = HD [253] Title = DG invalid system record Message = DG SRSDG 609 = [254] Title = DG invalid service record Message = DG SRSDG 610 = [255] Title = HD and UI software builds are not compatible Message = HD [256] Title = HD processor is in Disinfect Chemical Flush mode Message = HD = SRSUI 497 [257] Title = Alarm ID available for use Message = [258] Title = Alarm Id available Message = [259] Title = HD temperatures out of range Message = HD [260] Title = HD UI POST Application Integrity (Sha256Sum) failure Message = UI HD = SRSUI 1067 [261] Title = HD UI POST CANBus failure Message = UI HD = SRSUI 1068 [262] Title = HD UI POST Display failure Message = UI HD = SRSUI 1069 [263] Title = HD UI POST Touch failure Message = UI HD = SRSUI 1070 [264] Title = HD UI POST SD-Card failure Message = UI HD = SRSUI 1071 [265] Title = HD UI POST RTC failure Message = UI HD = SRSUI 1072 [266] Title = HD UI POST WiFi failure Message = UI HD = SRSUI 1073 [267] Title = HD UI POST Bluetooth failure Message = UI HD = SRSUI 1074 [268] Title = HD UI POST Ethernet failure Message = UI HD = SRSUI 1075 [269] Title = HD UI POST Sound failure Message = UI HD = SRSUI 1076 [270] Title = HD POST Safety Shutdown failure Message = HD [271] Title = DG POST Safety Shutdown failure Message = DG SRSDG 822 = [272] Title = HD Fan RPM out of range Message = HD [273] Title = Available for use Message = [274] Title = Available for use Message = [275] Title = HD arterial pressure sensor is reading out of range Message = HD [276] Title = HD venous pressure sensor is reading out of range Message = HD [277] Title = HD BP occlusion sensor is reading out of range Message = HD [278] Title = HD active reservoir weight out of range. Message = HD [279] Title = DG dialysate drain time out Message = DG SRSDG 752 = [280] Title = DG drain pump direction invalid Message = DG SRSDG 888 = [281] Title = Acid concentration bottle low volume alarm Message = DG SRSDG 836 = [282] Title = Bicarbonate concentration bottle low volume alarm Message = DG SRSDG 837 = [283] Title = DG load cells weight out of range Message = DG SRSDG 615 = [284] Title = DG load cells primary/back up drift out of range Message = DG SRSDG 383 = [285] Title = Available for use Message = [286] Title = DG concentrate cap not in proper position Message = DG [287] Title = HD processor clock speed checks against FPGA clock failure Message = HD [288] Title = HD load cells primary/back up drift out of range Message = HD [289] Title = DG dialysate cap not closed Message = DG SRSDG 928 = [290] Title = HD venous pressure sensor FPGA fault Message = HD [291] Title = HD processor is in Disinfect Flush mode Message = HD [292] Title = HD processor is in Disinfect Heat mode Message = HD [293] Title = HD processor is in Disinfect Chemical mode Message = HD [294] Title = Available for use Message = [295] Title = HD UI POST CloudSync failure Message = UI HD = SRSUI 1238 [296] Title = DG TDi sensors FPGA fault Message = DG SRSDG 951 = [297] Title = DG TRo sensors FPGA fault Message = DG SRSDG 950 = [298] Title = DG baro sensor FPGA fault Message = DG SRSDG 948 = [299] Title = DG invalid serial number Message = DG SRSDG 952 = [300] Title = HD invalid serial number Message = HD [301] Title = Alarm ID available Message = [302] Title = DG processor RAM error Message = DG SRSDG 126 = [303] Title = DG drain pump direction FPGA fault Message = DG SRSDG 958 = [304] Title = HD invalid usage record Message = HD [305] Title = HD processor RAM error Message = HD [306] Title = DG CD1 CD2 check fault Message = DG [307] Title = HD AC power lost second alarm Message = HD [308] Title = DG CPi conductivity sensor invalid character received Message = DG SRSDG 959 = [309] Title = DG CPo conductivity sensor invalid character received Message = DG SRSDG 960 = [310] Title = DG CD1 conductivity sensor invalid character received Message = DG SRSDG 962 = [311] Title = DG CD2 conductivity sensor invalid character received Message = DG SRSDG 961 = [312] Title = HD Dialysate In flow rate to too low. Message = HD [313] Title = DG cleaning mode inlet water temperature too low. Message = DG SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] Title = DG cleaning mode inlet water pressure too high. Message = DG SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] Title = DG cleaning mode inlet water pressure too low. Message = DG SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] Title = DG acid concentrate pump park fault. Message = DG SRSDG 983 = [317] Title = DG bicarb concentrate pump part fault. Message = DG SRSDG 984 = [318] Title = HD Air Pump timeout alarm Message = HD [319] Title = DG chemical disinfect flush flush sample. Message = DG SRSDG 1040 = [320] Title = DG chemical disinfect flush sample timeout. Message = DG SRSDG 1043 = [321] Title = DG outlet UV reactor not healthy Message = DG SRSDG 1069 = [322] Title = HD UI POST Year check failure Message = UI HD