[0] Title = Indication for when no alarms is a possible situation ListTitle = Message = [1] Title = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up ListTitle = HD POST: Stuck Button Message = Stop or Off button are pressed for more than 1 sec during POST. [2] Title = HD FPGA POST failure ListTitle = HD POST: FPGA Compatibility Message = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. [3] Title = DG FPGA POST failure ListTitle = DG POST: FPGA Compatibility Message = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = HD Watchdog POST failure ListTitle = HD POST: WD Safestate Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA [5] Title = DG Watchdog POST failure ListTitle = DG POST: WD Safestate Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V SRSDG 53 = [6] Title = UI communication POST failure. UI failed to communicate within a reasonable time after power up ListTitle = HD POST: Proc- UI Comm Message = If the UI failed to checkin via the CAN bus within 40 seconds [7] Title = Real-time clock configuration error ListTitle = HD Alarm (Low): RTC Bat Low Message = If the RTC is not functional as indicated by the bits in the first 3 control registers. [8] Title = HD accelerometer failed POST ListTitle = HD POST: Acc Vector Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. [9] Title = DG accelerometer failed POST ListTitle = DG POST: Acc Vector Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. SRSDG 323 = [10] Title = HD blood leak sensor embedded mode communication failed ListTitle = HD POST: BLD Config Emb Mode Message = If the blood leak sensor embedded mode is unresponsive. [11] Title = Dialysate temperature is below the target temperature ListTitle = Dial. Gen: Dialysate Temp Low Message = If the dialysate temperature is below the target temperature for a certain period of time. [12] Title = HD firmware image integrity POST test failed ListTitle = HD POST: HD Image Message = If the calculated CRC of the HD image does not match the expected CRC. [13] Title = DG firmware image integrity POST test failed ListTitle = DG POST: DG Image Message = If the calculated CRC of the DG firmware image does not matched the expected CRC SRSDG 813 = [14] Title = DG non-volatile usage information group invalid record CRC ListTitle = DG POST: Usage Rec NVRAM CRC Message = If the DG usage information record's CRC does not match the calculated CRC in POST. SRSDG 957 = [15] Title = Dialysate temperature is out of high safety range ListTitle = Dial. Gen: Dialysate Temp High Message = If the dialysate temperature is > 46 C for a certain period of time. [16] Title = HD alarm audio failed POST ListTitle = HD POST: Primary Alarm Current Message = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output [17] Title = HD UI POST failed ListTitle = HD POST: UI POST Not Complete Message = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. [18] Title = HD to DG comm error. ListTitle = Loss of DG Comm (DG ACK) Message = HD message to DG that requires acknowledge was not acknowledged within a certain time. [19] Title = Dialysate temperature is above the target temperature ListTitle = Dial. Gen: Dialysate Temp High Message = If the dialysate temperature is above the target temperature for a certain period of time. [20] Title = DG conductivity sensors invalid calibration record ListTitle = DG POST: Conductivity Cal Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 16 = [21] Title = DG drain line volume invalid calibration record ListTitle = DG POST: Drain Line Vol Cal Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 936 = [22] Title = DG reservoirs invalid calibration record ListTitle = DG POST: Reservoir Vol Cal Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 935 = [23] Title = DG acid concentrate calibration record ListTitle = DG POST: Acid Conc Cal Message = If the non-volatile data is valid and if not raise the alarm SRSDG 934 = [24] Title = DG bicarb concentrate calibration record ListTitle = DG POST: Bicarb Conc Cal Message = If the non-volatile data is valid and if not raise the alarm SRSDG 933 = [25] Title = DG accelerometer invalid calibration record ListTitle = DG POST: Acc Cal Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 817 = [26] Title = HD accelerometer invalid calibration record ListTitle = HD POST: Acc Cal Message = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. [27] Title = HD BLD sensor failed zero sequence ListTitle = Process: Pre- Tx Wet Self- Test Message = BLD fails to complete zero sequence. [28] Title = DG two wire sensors FPGA fault ListTitle = DG Fault: Conductivity RTD Comm Message = If the FPGA does not report fresh data for a certain period of time SRSDG 864 = [29] Title = HD heparin force sensor invalid calibration record ListTitle = HD POST: SP DAC Cal Message = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. [30] Title = HD Software fault. Software found itself in an unexpected state ListTitle = HD Fault: SW Invalid State Message = If the specific software faults listed in alarm management SW faults has been detected [31] Title = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ListTitle = HD Fault: BP Current Message = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. [32] Title = Blood pump failed motor off check. Measured speed while commanded off ListTitle = HD Fault: BP Speed Message = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM [33] Title = Blood pump failed motor direction check. Measured vs. commanded ListTitle = HD Fault: BP Direction Message = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. [34] Title = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds ListTitle = HD Fault: BP Rotor Message = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. [35] Title = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ListTitle = HD Fault: DPi Current Message = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. [36] Title = Dialysis inlet pump failed motor off check. Measured speed while commanded off ListTitle = HD Fault: DPi Speed Message = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. [37] Title = Dialysis inlet pump failed motor direction check. Measured vs. commanded ListTitle = HD Fault: DPi Direction Message = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. [38] Title = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds ListTitle = HD Fault: DPi Rotor Message = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. [39] Title = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ListTitle = HD Fault: DPo Current Message = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. [40] Title = Dialysis outlet pump failed motor off check. Measured speed while commanded off ListTitle = HD Fault: DPo Speed Message = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. [41] Title = Dialysis outlet pump failed motor direction check. Measured vs. commanded ListTitle = HD Fault: DPo Direction Message = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. [42] Title = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds ListTitle = HD Fault: DPo Rotor Message = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed [43] Title = Main primary heater FPGA fault ListTitle = DG Fault: Iso ADC Comm Message = If the FPGA does not report fresh data for a certain period of time SRSDG 954 = [44] Title = User interface communication timeout ListTitle = HD Fault: UI Comm Message = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. [45] Title = Too many bad CRCs detected on received system messages ListTitle = HD Fault: CAN CRC Com Message = If too many bad messages CRCs are received within a set period of time [46] Title = System message that required acknowledgment was not acknowledged ListTitle = HD Fault: CAN ACK UI Com Message = If no retries left in the pending acknowlegement list. [47] Title = ListTitle = DG Fault: RO Pump High Message = [48] Title = Ultrafiltration volume accuracy error during treatment ListTitle = HD Fault: UF Rate Tare Err Message = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. [49] Title = HD FPGA communication down for too long ListTitle = HD Fault: FPGA Com Message = If there are more than 3 comm errors in a 1 min period. [50] Title = DG FPGA not accepting commanded valve states ListTitle = DG Fault: Valve Control Message = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. SRSDG 26 = [51] Title = Blood pump failed motor speed check. Measured vs. commanded ListTitle = HD Fault: BP Speed Message = If measured motor speed versus commanded rotor speed exceeded limits. [52] Title = Dialysate inlet pump failed motor speed check. Measured vs. commanded ListTitle = HD Fault: DPi Speed Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. [53] Title = Dialysate outlet pump failed motor speed check. Measured vs. commanded ListTitle = HD Fault: DPo Speed Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. [54] Title = HD critical data integrity check failed ListTitle = HD Fault: Critical Data Err Message = If the critical data failed range check. [55] Title = DG critical data integrity check failed ListTitle = DG Fault: Critical Data Err Message = If the critical data failed range check. SRSDG 921 = [56] Title = HD accelerometer error (no readings or FPGA reports error) ListTitle = HD Fault: Acc Com Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. [57] Title = DG accelerometer error (no readings or FPGA reports error) ListTitle = DG Fault: Acc Com Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. SRSDG 869 SRSDG 870 = [58] Title = HD valve homing failed ListTitle = HD Fault: Valve Home Message = If the number of failed homing > = 3 counts. [59] Title = HD valve transition time out ListTitle = HD Fault: Valve Transition Err Message = if the valve's transition time has timed out (3000msec) [60] Title = HD valve not functional ListTitle = HD Fault: Valve Control Message = If the valve bit masks are not as expected in different valve control mode. [61] Title = HD valve current out of range ListTitle = HD Fault: Valve Current Message = If the valve current is over the threshold of 1.0 A for 1000 msec. [62] Title = HD valve position out of target ListTitle = HD Fault: Valve Position Message = If the current position has deviated from the expected position for more than 1 second. [63] Title = ListTitle = Message = [64] Title = Barometric pressure is out of range ListTitle = DG Fault: Baro Range Message = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. SRSDG 874 = [65] Title = HD requests DG command with invalid parameter fault ListTitle = Loss of DG Comm (Reject) Message = If the latest DG command response indicated invalid parameter [66] Title = HD blood leak sensor set point set failure ListTitle = HD POST: BLD Config Set Point Message = If the set point of the blood leak sensor does not match the set point in the calibration record. [67] Title = HD blood pump occlusion self-test failure alarm ListTitle = Process: Pre- Tx Remove Cartridge Message = If the blood pump pressure > 5000 [68] Title = HD active reservoir recirculation out of range ListTitle = Dial. Gen: Dialysate not ready Message = If the recirculation level has exceeded the max allowed [69] Title = HD blood leak sensor invalid calibration record ListTitle = HD POST: BLD Cal Message = If the BLD data that was received from NVDataMgmt is NOT legitimate. [70] Title = HD arterial pressure self-test failure alarm ListTitle = Process: Pre- Tx Remove Cartridge Message = If arterial pressure < -300 or > 100 mmHg [71] Title = HD venous pressure self-test failure alarm ListTitle = Process: Pre- Tx Remove Cartridge Message = If the venous pressure < -100 or > 600 mmHg [72] Title = HD No load cell data message receive at least once every 2 seconds ListTitle = Loss of DG Comm (LC) Message = No new load cell data for 2 seconds. [73] Title = HD No dialysate temperature data message received at least once every 2 seconds ListTitle = Loss of DG Comm (TDI/TRO) Message = No new dialysate temperature data for 2 seconds. [74] Title = DG inlet UV reactor is on while there is no flow ListTitle = DG Fault: UVi On No Flow Message = If the inlet UV reactor is on while VPi is closed for a certain period of time. SRSDG 600 = [75] Title = HD syringe pump self-test failure alarm ListTitle = HD Fault: Syringe DAC Message = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. [76] Title = HD monitored voltage is out of range ListTitle = HD Fault: Voltage Monitor Message = If the HD monitored voltages are out of range for a certain period of time. [77] Title = DG monitored voltage is out of range ListTitle = DG Fault: Voltage Monitor Message = If the DG monitored voltages are out of range for a certain period of time. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = HD syringe pump direction (from encoder) error ListTitle = HD Fault: Syringe Pump Enc Dir Message = If the measured encoder direction of the syringe pump does not agree with the expected direction [79] Title = HD syringe pump direction (from controller) error ListTitle = HD Fault: Syringe Pump Drv Dir Message = If the measured direction of the syringe pump motor controller does not agree with the expected direction [80] Title = HD syringe pump fault reported by FPGA ListTitle = HD Fault: Syringe Pump Driver Message = If the FPGA reported the syringe pump fault [81] Title = HD syringe pump over travel error ListTitle = HD Fault: Syringe Pump OT Message = If the syringe pump position is deemed to have overtraveled [82] Title = HD syringe pump DAC write failure ListTitle = HD Fault: Syringe Pump DAC Err Message = If the syringe pump DAC write error is reported by the FPGA [83] Title = HD syringe pump is running while the BP is off ListTitle = HD Fault: Syringe Pump On Err Message = If the syringe pump is on while the blood pump is off [84] Title = DG RTC set year is invalid ListTitle = DG POST: DG RTC Time Message = If DG RTC set year is < 2021. SRSDG 815 = [85] Title = HD pump track lever is not fully latched ListTitle = Blood Stop: Cartridge Lever Open Message = If the pump track lever is detected as unlatched when it should be. [86] Title = HD RTC set year is invalid ListTitle = HD POST: HD RTC Time Message = If HD RTC set year is < 2021. [87] Title = DG heating parameters invalid calibration record ListTitle = DG POST: Heat Cal Message = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. SRSDG 929 = [88] Title = DG concentrate pump hall sensor out of range. ListTitle = DG Fault: CP1 CP2 Speed Range Message = Hall sensor pulse to pulse timing is < 1ms (speed too fast). [89] Title = DG outlet UV reactor is on while there is no flow ListTitle = DG Fault: UVo On No Flow Message = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. SRSDG 599 = [90] Title = DG load cells A1/B1 FPGA fault ListTitle = DG Fault: LC A1/B1 Comm Message = If the FPGA does not report fresh data for a certain period of time. SRSDG 604 = [91] Title = HD No dialysate flow data received in the last 3 seconds ListTitle = Loss of DG Comm (FMD) Message = If not receiving new dialysate flow readings in expected interval (3 seconds) [92] Title = ListTitle = Message = [93] Title = DG outlet primary conductivity sensor out of range ListTitle = DG Fault: RO Permeate Cond High Message = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. SRSDG 430 = [94] Title = DG pressure sensor out of range ListTitle = DG Fault: Pressure Sensor Range Message = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. SRSDG 629 = [95] Title = DG watchdog expired ListTitle = DG Fault: WD Error Message = If the watchdog signal is low while not in POST. SRSDG 907 = [96] Title = DG inlet water temperature out of range ListTitle = Dial. Gen: Inlet Water Temp High Message = If the inlet water temperature is > 37 C. SRSDG 410 = [97] Title = DG fill conductivity out of range ListTitle = Dial. Gen: Dialysate Cond. Message = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. SRSDG 397 = [98] Title = HD battery or battery charger communication fault ListTitle = HD Alarm (Low): Battery Com Message = If battery I2C communication fault persistence count > 5. [99] Title = HD syringe pump stall alarm ListTitle = HD Fault: Syringe Pump Stall Message = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. [100] Title = ListTitle = Message = [101] Title = UI POST FW Communication failure ListTitle = HD POST: UI- Proc Comm Message = If the HD is not communicating with the UI during POST. [102] Title = DG heat disinfection temperature gradient out of range ListTitle = Cleaning: Tpo THD Temp Delta Message = If temperature gradient between Tpo and Thd exceeds maximum 15°C. SRSDG 856 SRSDG 930 = [103] Title = HD non-volatile calibration group invalid record CRC ListTitle = HD POST: Cal NVRAM CRC Message = If the computed CRC of the calibration group record is not equal to the record CRC. [104] Title = Air trap level sensors reporting illegal combination of air/fluid ListTitle = HD Fault: Air Trap Sensor Message = If the air trap level readings are not valid for more than 2 seconds. [105] Title = DG non-volatile system group invalid record CRC ListTitle = DG POST: Cal NVRAM CRC Message = If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. SRSDG 608 = [106] Title = DG flow sensors calibration error ListTitle = DG POST: Flow Sensor Cal Message = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = HD reports DG restarted fault ListTitle = DG Fault: DG Restarted Message = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. [108] Title = HD syringe pump ADC error ListTitle = HD Fault: Syringe Pump ADC Com Message = If the ADC read is not fresh after POST init and the ADC read counter has not changed. [109] Title = HD syringe pump volume check error ListTitle = HD Fault: Syringe Pump Vol Message = If the volume delivered is off by more than 5% or 0.1mL. [110] Title = HD syringe pump speed check error ListTitle = HD Fault: Syringe Pump Speed Message = If the syringe measured rate is off by more than 5% or 0.1mL. [111] Title = HD syringe pump not stopped in off state error ListTitle = HD Fault: Syringe Pump Speed Message = If the syringe pump position is not equal to its last position. [112] Title = DG load cells A2/B2 FPGA fault ListTitle = DG Fault: LC A2/B2 Comm Message = If the FPGA does not report fresh data for a certain period of time. SRSDG 953 = [113] Title = DG drain pump current out of range ListTitle = DG Fault: DRP Current Message = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. SRSDG 890 = [114] Title = HD venous air bubble detector self-test failure ListTitle = Process: Pre- Tx Wet Self- Test Message = If the venous air bubble self test did not detect bubble for more than 500 msec. [115] Title = DG temperature sensor out of range ListTitle = DG Fault: Temp Sensor Range Message = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. SRSDG 224 = [116] Title = SD card is non-operational ListTitle = HD Alarm (Low): SD- Card Error Message = SD card is non-operational [117] Title = System message that required acknowledgment was not acknowledged ListTitle = Loss of HD Comm (HD ACK) Message = If no retries left in the pending acknowlegement list. SRSDG 909 = [118] Title = DG RTC configuration error ListTitle = DG Fault: RTC Config Message = If the RTC configuration registers report an error. SRSDG 43 = [119] Title = DG RTC battery low ListTitle = DG Alarm (Low): RTC Bat Low Message = If the RTC battery bit report a low alarm. SRSDG 908 = [120] Title = HD pre-treatment wet prime test failure ListTitle = Process: Pre- Tx Wet Self- Test Message = If the HD pre-treatment wet self test failed. [121] Title = DG main primary heater voltage out of range ListTitle = DG Fault: Prim Heater Lg Control Message = If the main primary heater's voltage is out of range. SRSDG 611 = [122] Title = DG small primary heater voltage out of range ListTitle = DG Fault: Prim Heater Sm Control Message = If the small primary heater's voltage is out of range. SRSDG 612 = [123] Title = DG trimmer heater voltage out of range ListTitle = DG Fault: Trimmer Heater Control Message = If the trimmer voltage is out of range. SRSDG 613 = [124] Title = HD end of treatment alarm (high priority - user still not acting to end treatment) ListTitle = Blood Stop: Blood Clot Risk Message = In treatment end paused state for > 5 minutes. [125] Title = Blood sitting too long after treatment stopped by user alarm (> 5 min) ListTitle = Blood Stop: Blood Clot Risk Message = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. [126] Title = HD blood leak detected alarm ListTitle = Blood Stop: Dialyzer Blood Leak Message = If the blood leak is detected based on the blood leak status reported by the sensor. [127] Title = Venous pressure too low during treatment ListTitle = Blood Stop: Venous Pres. Low Message = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. [128] Title = HD venous air bubble detected alarm ListTitle = Blood Stop: Venous Air Bubble Message = If bubbles are detected during the treatment. [129] Title = HD blood leak detected and it is in the process of attempting to remove the bubbles/foam ListTitle = Blood Stop: Dialyzer Blood Leak Bypass Message = If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. [130] Title = Venous pressure too high during treatment ListTitle = Blood Stop: Venous Pres. High Message = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. [131] Title = Arterial pressure too low during treatment ListTitle = Blood Stop: Arterial Pres. Low Message = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. [132] Title = Arterial pressure too high during treatment ListTitle = Blood Stop: Arterial Pres. High Message = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. [133] Title = DG fluid leak detected alarm ListTitle = DG Fault: Fluid Leak in DG Message = If the fluid leak is detected as reported by the fluid leak detector. SRSDG 328 = [134] Title = HD fluid leak detected alarm ListTitle = Blood Stop: Fluid Leak in HD Message = If the fluid leak is detected as reported by the fluid leak detector. [135] Title = HD experienced large acceleration ListTitle = Blood Stop: HD Shock Detected Message = If the system experienced excessive shock > 2.5 g. [136] Title = DG experienced large acceleration ListTitle = Dial. Gen: Shock- Vibration Message = If the system experienced excessive shock > 2.5 g. SRSDG 327 = [137] Title = HD tilt exceeds maximum ListTitle = Blood Stop: HD Tilted Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. [138] Title = DG tilt exceeds maximum ListTitle = Dial. Gen: Tilt Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. SRSDG 326 = [139] Title = HD AC power lost alarm ListTitle = Loss of AC Power Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. [140] Title = Dialysate generator communication timeout ListTitle = Loss of DG Comm (Timeout) Message = If DG communication timeout is detected for more than 2 seconds. [141] Title = Air trap fill timeout during treatment ListTitle = Blood Stop: Air Trap Fill Message = If the air trap fill timeout during treatment. [142] Title = Blood pump occlusion detected ListTitle = Blood Stop: Blood Pump Occlusion Message = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall [143] Title = DG dialysate temperature sensors drift timeout. ListTitle = DG Fault: TRo TDi Delta Message = If TDi and TRo have drifted for more 1 C for a certain period of time. SRSDG 912 = [144] Title = DG cleaning mode inlet water conductivity out of high range ListTitle = Cleaning: Inlet Water Cond High Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] Title = DG Concentrate conductivity after adding acid out of range alarm ListTitle = Dial. Gen: Acid Conc. Cond. Message = If the conductivity value for the acid concentrate is out of range for more than 45 secs. SRSDG 846 = [146] Title = DG RTC vs. Timer accuracy error ListTitle = DG POST: DG RTC Timer Message = If RTC and timer are not aligned on 1 second elapsed time. SRSDG 312 = [147] Title = DG dialysate generation conductivity fault alarm ListTitle = Dial. Gen: Dialysate not ready Message = If dialysate conductivity is not with 5% of the TCD SRSDG 398 = [148] Title = Cleaning Mode Inlet water temperature too high ListTitle = Cleaning: Inlet Water Temp High Message = Inlet temperature > 37 C. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = HD watchdog expired ListTitle = HD Fault: WD Error Message = If the watchdog signal is low while not in POST. [150] Title = Inlet water conductivity too high ListTitle = Dial. Gen: Inlet Water Cond High Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). SRSDG 391 = [151] Title = Inlet water conductivity too low ListTitle = Dial. Gen: Inlet Water Cond Low Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). SRSDG 826 = [152] Title = Inlet water pressure too low ListTitle = Dial. Gen: Inlet Water Pres Low Message = Inlet water pressure < 25 PSI. SRSDG 417 = [153] Title = HD prime completed high priority alarm ListTitle = Process: Pre- Tx Terminated Message = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) [154] Title = Cleaning mode inlet water conductivity too low ListTitle = Cleaning: Inlet Water Cond Low Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] Title = HD No reservoir data message receive at least once every 2 seconds ListTitle = Loss of DG Comm (Reservoir) Message = No new reservoir data for 2 seconds. [156] Title = HD No DG op mode message receive at least once every 2 seconds ListTitle = Loss of DG Comm (Op Mode) Message = No DG op mode for 2 seconds. [157] Title = Chemical disinfect priming the acid line timeout ListTitle = Cleaning: Chem Disinf PAA Prime Message = If the disinfectant acid line is not primed within a certain period of time. SRSDG 988 = [158] Title = Inlet water temperature low ListTitle = Dial. Gen: Inlet Water Temp Low Message = Inlet temperature < 24 °C for 5 seconds. SRSDG 408 = [159] Title = Close concentrate line be able to flush it ListTitle = Cleaning: Chem Disinf Connect Cap Message = Start of chemical disinfect flush. SRSDG 1044 = [160] Title = Blood pump rotor speed too high ListTitle = HD Fault: BP Rotor Max Speed Message = If blood pump motor speed exceeded 100.0 RPM. [161] Title = Too many bad CRCs detected on received system messages ListTitle = DG Fault: HD- DG Comm Message = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. SRSDG 122 = [162] Title = FPGA clock speed failure ListTitle = DG Fault: Proc- FPGA Clock Message = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. SRSDG 867 = [163] Title = HD sees primary load cell for reservoir 1 change too much too fast ListTitle = Dial. Gen: UF Rate Check Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). [164] Title = HD sees primary load cell for reservoir 2 change too much too fast ListTitle = Dial. Gen: UF Rate Check Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). [165] Title = HD in treatment rinseback operation timeout ListTitle = Blood Stop: Blood Clot Risk Message = If the maximum time inrinseback operation has been exceeded. [166] Title = ListTitle = Message = [167] Title = HD cartridge door opened alarm ListTitle = Blood Stop: Door Open Message = If the cartridge door is opened as reported by the FPGA. [168] Title = ListTitle = Message = [169] Title = Dialysate flow rate out of range ListTitle = DG Fault: Dialysate Flow Sens Rg Message = If the dialysate flow rate > 2.00 L/min. SRSDG 643 = [170] Title = HD syringe empty alarm ListTitle = Heparin Syringe Empty Message = If the syringe pump position is near empty position. [171] Title = HD syringe pump occlusion alarm ListTitle = Process: Pre- Tx Hep. Pump Occ. Message = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. [172] Title = HD syringe pump not enough Heparin alarm ListTitle = Process: Pre- Tx Not Enough Hep. Message = If the estimated syringe volume needed for treatment is not sufficient [173] Title = HD RTC configuration error ListTitle = HD Fault: RTC Config Message = If the RTC configuration registers report an error. [174] Title = HD RTC vs. Timer accuracy error ListTitle = HD POST: HD RTC Timer Message = If RTC and timer are not aligned on 1 second elapsed time. [175] Title = HD pump commutation error ListTitle = HD Fault: Pump Commutation Err Message = If the set pump direction does not agree with the hall sensor direction. [176] Title = ListTitle = Message = [177] Title = DG software fault. Software found itself in an unexpected state ListTitle = DG Fault: Software Invalid State Message = If the specific software faults listed in alarm management SW faults has been detected SRSDG 919 = [178] Title = HD communication timeout ListTitle = HD Fault: UI- Proc Comm Message = If the HD stops communicating with the UI. = SRSUI 1066 [179] Title = DG FPGA communication down for too long ListTitle = DG Fault: FPGA Com Message = If retries for commands exceeds limit or FPGA reports comm error. SRSDG 605 = [180] Title = RO flow rate out of range ListTitle = DG Fault: RO Flow Sensor Range Message = If the RO flow rate > 2.00 L/min. SRSDG 883 = [181] Title = DG load cells weight out of range for tare ListTitle = DG Fault: Tare Out of Range Message = If the load cell weight is out of range for tare. SRSDG 849 SRSDG 838 = [182] Title = DG load cells invalid calibration ListTitle = DG POST: LC Cal Message = If the CRC of load cells calibration data does not match the calculated CRC. SRSDG 162 = [183] Title = DG invalid load cell value ListTitle = Cleaning: Invalid Resv Transfer Message = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. SRSDG 851 = [184] Title = Inlet reactor not healthy ListTitle = DG Fault: UVi Error Message = If the inlet reactor are not healthy. SRSDG 559 = [185] Title = DG fan RPM out of range ListTitle = DG Alarm (Low): Fan Speed Message = If the fan RPM is out of Min/Max range. SRSDG 628 = [186] Title = DG concentrate pump fault ListTitle = DG Fault: CP1/CP2 Fault Message = If the concentrate pumps fault is reported by the FPGA. SRSDG 891 = [187] Title = DG concentrate pump CP1 speed control error ListTitle = DG Fault: CP1 Speed Fault Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. SRSDG 905 = [188] Title = DG concentrate pump CP2 speed control error ListTitle = DG Fault: CP2 Speed Fault Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. SRSDG 906 = [189] Title = DG drain pump RPM out of range ListTitle = DG Fault: DRP Speed Fault Message = If the drain pump RPM > 100.0 of the target RPM. SRSDG 448 = [190] Title = DG drain pump off fault ListTitle = DG Fault: DRP Speed Not Off Message = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. [191] Title = DG flow rate out of upper range ListTitle = DG Fault: RO Flow High Message = If the RO flow rate > 10% from the target flow rate. SRSDG 885 = [192] Title = DG flow rate out of lower range ListTitle = DG Fault: RO Flow Low Message = If the RO flow rate < 10% from the target flow rate. SRSDG 945 = [193] Title = ListTitle = Message = [194] Title = DG RO pump duty cycle out of range ListTitle = DG Fault: RO Driver Feedback Message = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. SRSDG 881 = [195] Title = DG RO pump pressure out of range ListTitle = DG Fault: RO Pump High Pressure Message = If the DG RO pump pressure > 135 psi for a certain period of time. SRSDG 925 = [196] Title = DG CPi/CPo conductivity sensors FPGA fault ListTitle = DG Fault: CPi/CPo Comm Message = If the FPGA reports error for a certain period of time. SRSDG 955 = [197] Title = DG CD1/CD2 conductivity sensors FPGA fault ListTitle = DG Fault: CD1/CD2 Comm Message = If the FPGA reports error for a certain period of time. SRSDG 956 = [198] Title = DG primary heater while RO flow is low ListTitle = DG Fault: HP On With No Flow Message = If the RO flow is < 0.2 L/min for a certain period of time. SRSDG 872 = [199] Title = DG trimmer heater while dialysate flow is low ListTitle = DG Fault: HR On With No Flow Message = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. SRSDG 871 = [200] Title = DG thermistor or temperature sensor out of range ListTitle = DG Fault: Thermistor Range Message = If thermistor > 80.0 C or < 0.0 C. SRSDG 178 = [201] Title = HD pre-treatment wet flow test failure ListTitle = Process: Pre- Tx Wet Self- Test Message = If the pre-treatment wet flow test has failed. [202] Title = HD pre-treatment dry pressure normal test failure ListTitle = Process: Pre- Tx Dry Self- Test Message = If the pre-treatment dry pressure normal test failed. [203] Title = ListTitle = Message = [204] Title = RO rejection ratio out of range ListTitle = DG Fault: RO Rejection Message = If RO rejection ratio > 0.1 SRSDG 429 = [205] Title = Conductivity sensor fault ListTitle = DG Fault: Cond Sensors Error Message = If conductivity sensor reports 0x11 for more than 2 consecutive reads. SRSDG 913 = [206] Title = DG dialysate fill runs out of time ListTitle = DG Fault: Reservoir Fill Timeout Message = If DG dialysate fill > 5 minutes. SRSDG 437 = [207] Title = DG flow meter check failure alarm ListTitle = DG Fault: FMP Flow Rate Error Message = If integrated Volume To Load Cell Reading Percent > 0.1. SRSDG 240 = [208] Title = HD venous line occlusion fault ListTitle = Blood Stop: Venous Pres. High Message = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. [209] Title = DG drain circulation line timeout ListTitle = Cleaning: Flush Circ Error Message = If the integrated volume < 0.1 for more than 60 secs. SRSDG 980 = [210] Title = HD battery pack detected an error ListTitle = HD Alarm (Low): Battery Fault Message = If the battery pack status error is detected from the slave device I2C address 0x0B. [211] Title = Blood sitting too long warning (> 4 min) ListTitle = Blood Stopped: Warning Message = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. [212] Title = Blood sitting too long warning (> 4 min) ListTitle = Blood Stopped: Warning Message = During treatment end paused state, if the blood is in the line for more than 4 minutes. [213] Title = HD prime completed medium priority alarm ListTitle = Process: Pre- Tx Prime Complete Message = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) [214] Title = ListTitle = Message = [215] Title = ListTitle = Message = [216] Title = HD end treatment sub-mode timeout alarm ListTitle = Process: End of Dialysis Tx Message = If the treatment ended successfully. [217] Title = ListTitle = Message = [218] Title = HD syringe detected alarm ListTitle = Process: Remove Syringe Message = If the syringe detection switch > 2.0 volts. [219] Title = HD syringe pump syringe removed alarm ListTitle = Blood Stop: Syringe Not Detected Message = If the syringe detection switch < 2.0 volts. [220] Title = DG THd sensor FPGA fault ListTitle = DG Fault: THD RTD Comm Message = If the FPGA does not report fresh data for a certain period of time SRSDG 947 = [221] Title = HD Saline bag is empty ListTitle = Blood Stop: Saline Bag Empty Message = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. [222] Title = HD occlusion sensor FPGA fault ListTitle = HD Fault: BP Occ Com Message = If the FPGA does not report fresh data or indicates an error for a certain period of time. [223] Title = HD arterial sensor FPGA fault ListTitle = HD Fault: Art. Pres. Com Message = If the FPGA does not report fresh data for a certain period of time [224] Title = Treatment stopped by user action - pressed stop button ListTitle = Blood Stopped: Treatment Pause Message = If the user has pressed the stop button. [225] Title = HD end of treatment warning ListTitle = Process: End of Treatment Message = If the treatment duration has exceeded the prescribed treatment time entered by the user. [226] Title = HD prime completed low priority alarm ListTitle = Process: Pre- Tx Priming Complete Message = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection [227] Title = ListTitle = Message = [228] Title = HD time out on prime sub-mode purge air state. ListTitle = Process: Pre- Tx Prime Timeout Message = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. [229] Title = HD prime dialysate dialyzer time out alarm ListTitle = Process: Pre- Tx Prime Timeout Message = If the priming of the dialyzer dialysate flow path has exceeded the time limit. [230] Title = HD prime dialysate bypass time out alarm ListTitle = Process: Pre- Tx Prime Timeout Message = If the priming of the dialyzer bypass flow path has exceeded the time limit. [231] Title = HD pre-treatment mode dry self-tests failure ListTitle = Process: Pre- Tx Dry Self- Test Message = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState [232] Title = HD pre-treatment mode wet self-tests failure ListTitle = Process: Pre- Tx Wet Self- Test Message = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState [233] Title = DG barometric pressure sensor bad coefficients CRC ListTitle = DG Fault: Baro Cal CRC Message = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] Title = DG inlet water pressure is too high ListTitle = Dial. Gen: Inlet Water Pres High Message = If the inlet water pressure is > 80 psig. SRSDG 1037 = [235] Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) ListTitle = Process: Rinseback Complete Message = If there is no active alarm and rinseback has completed. [236] Title = HD needs new cartridge to be installed ListTitle = Process: Pre- Tx Dry Self- Test Message = If the air trap sensor reported AIR and bubble sensor detected bubbles. [237] Title = ListTitle = Message = [238] Title = HD no cartridge loaded or installed improperly alarm ListTitle = Process: Pre- Tx Dry Self- Test Message = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. [239] Title = HD fail to remove cartridge alarm ListTitle = Process: Post- Tx Remove Cartridge Message = In the post treatment state, if the cartridge has not been removed. [240] Title = Bicarb conductivity out of range during bicarb pump check alarm ListTitle = Dial. Gen: Bicarb Conc. Cond. Message = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. SRSDG 482 = [241] Title = DG reservoir drain time out ListTitle = Cleaning: Resv Drain Timeout Message = If the DG reservoir drain is not complete in 120 secs. SRSDG 848 SRSDG 90 SRSDG 561 SRSDG 577 SRSDG 590 SRSDG 591 SRSDG 336 SRSDG 879 SRSDG 340 SRSDG 496 SRSDG 501 SRSDG 505 SRSDG 532 SRSDG 535 SRSDG 556 SRSDG 557 SRSDG 1048 SRSDG 1049 SRSDG 651 SRSDG 655 SRSDG 744 SRSDG 745 SRSDG 1004 SRSDG 1008 SRSDG 1019 SRSDG 1023 SRSDG 1033 SRSDG 1034 = [242] Title = DG reservoir fill time out ListTitle = Cleaning: Resv Fill Timeout Message = If the DG reservoir fill is not complete in 120 secs. SRSDG 850 SRSDG 573 SRSDG 574 SRSDG 579 SRSDG 580 SRSDG 492 SRSDG 493 SRSDG 498 SRSDG 499 SRSDG 508 SRSDG 509 SRSDG 519 SRSDG 690 SRSDG 691 SRSDG 703 SRSDG 711 SRSDG 1020 SRSDG 1024 = [243] Title = DG reservoir leak time out ListTitle = Cleaning: Fluid Loss Detected Message = If the reservoir level is drifted by 100 mL for more than 30 seconds. SRSDG 853 = [244] Title = DG temperature sensors difference out of range ListTitle = Cleaning: Temp Sensor Check Message = If any of the temperature sensors are out of the average of the sensors by more than 3 C. SRSDG 852 = [245] Title = DG heat disinfect target temperature time out (could not reach to temperature) ListTitle = Cleaning: Heat Disinfect Temp TO Message = If heating up to minimum temperature for heat disinfect failed after 4 hours. SRSDG 854 = [246] Title = DG chemical disinfect inlet conducitivty out of range ListTitle = Cleaning: RO Water Cond Error Message = If CPo and CD2 conductivity values are above 100 uS/cm. [247] Title = ListTitle = Message = [248] Title = DG chemical disinfect out of target temperature ListTitle = Cleaning: Chem Disinfect Proc Message = If TPo > 50 C after 1 second. [249] Title = DG chemical disinfect out of target conductivity ListTitle = Cleaning: Chem Disinf PAA Cond Message = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. SRSDG 992 = [250] Title = Insert the acid to start chemical disinfect ListTitle = Cleaning: Chem Disinf Connect PAA Message = Start of chemical disinfect SRSDG 1046 = [251] Title = HD invalid system record ListTitle = HD POST: Sys Rec NVRAM CRC Message = If the calculated CRC of the HD system record image does not match the expected CRC. [252] Title = HD invalid service record ListTitle = HD POST: Serv Rec NVRAM CRC Message = If the calculated CRC of the HD service record image does not match the expected CRC. [253] Title = DG invalid system record ListTitle = DG POST: Sys Rec NVRAM CRC Message = If the calculated CRC of the DG system record image does not match the expected CRC. SRSDG 609 = [254] Title = DG invalid service record ListTitle = DG POST: Serv Rec NVRAM CRC Message = If the calculated CRC of the DG service record image does not match the expected CRC. SRSDG 610 = [255] Title = HD and UI software builds are not compatible ListTitle = HD POST: UI Compatibility Message = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. [256] Title = DG flush started ListTitle = Cleaning: Chem Flush In Progress Message = If DG flush successfully started from the UI. = SRSUI 497 [257] Title = ListTitle = Message = [258] Title = ListTitle = Message = [259] Title = HD temperatures out of range ListTitle = HD Fault: Temp Sensors Error Message = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. [260] Title = UI POST Application Integrity (Sha256Sum) failure ListTitle = HD POST: UI Image Message = If the UI POST reports a shasum faillure. = SRSUI 1067 [261] Title = UI POST CANBus failure ListTitle = HD POST: UI CAN Driver Message = If the UI POST reports a can bus failure. = SRSUI 1068 [262] Title = UI POST Display failure ListTitle = Message = If the UI POST reports a display failure. [263] Title = UI POST Touch failure ListTitle = HD POST: UI Touch Driver Message = If the UI POST reports a touch failure. = SRSUI 1070 [264] Title = UI POST SD-Card failure ListTitle = HD POST: UI SD- Card Message = If the UI POST reports SD card failure. = SRSUI 1071 [265] Title = UI POST RTC failure ListTitle = HD POST: UI RTC Message = If the UI POST reports RTC failure. = SRSUI 1072 [266] Title = UI POST WiFi failure ListTitle = HD POST: UI WiFi Message = If the UI POST reports WiFi failure. = SRSUI 1073 [267] Title = UI POST Bluetooth failure ListTitle = HD POST: UI Bluetooth Message = If the UI POST reports bluetooth failure. = SRSUI 1074 [268] Title = UI POST Ethernet failure ListTitle = HD POST: UI Ethernet Message = If the UI POST reports ethernet failure. = SRSUI 1075 [269] Title = UI POST Sound failure ListTitle = HD POST: UI Sound Message = If the UI POST reports sound failure. = SRSUI 1076 [270] Title = HD POST Safety Shutdown failure ListTitle = HD POST: Proc Safestate Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. [271] Title = DG POST Safety Shutdown failure ListTitle = DG POST: Proc Safestate Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. SRSDG 822 = [272] Title = HD Fan RPM out of range ListTitle = HD Alarm (Low): Fan Speed Message = If the fan RPM is out of Min/Max range. [273] Title = ListTitle = Message = [274] Title = ListTitle = Message = [275] Title = HD arterial pressure sensor is reading out of range ListTitle = HD Fault: Arterial Press Range Message = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. [276] Title = HD venous pressure sensor is reading out of range ListTitle = HD Fault: Venous Press Range Message = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. [277] Title = HD BP occlusion sensor is reading out of range ListTitle = HD Fault: Occlusion Sens Range Message = If the blood pump occlusion count is < 2000. [278] Title = HD active reservoir weight out of range. ListTitle = DG Fault: Reservoir Overfill Message = If the active reservoir weight is > 1900.0 g. [279] Title = DG dialysate drain time out ListTitle = DG Fault: Reservoir Drain Message = If the DG dialysate drain exceeded 120 seconds. SRSDG 752 = [280] Title = DG drain pump invalid direction ListTitle = DG Fault: DRP Direction Fault Message = If the drain pump's direction is not 1 when draining. SRSDG 888 = [281] Title = DG Acid concentration bottle low volume alarm ListTitle = Dial. Gen: Acid Conc. Low Message = If the acid concentration bottle volume > 3240 (3340 - 100). SRSDG 836 = [282] Title = DG Bicarbonate concentration bottle low volume alarm ListTitle = Dial. Gen: Bicarb Conc. Low Message = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). SRSDG 837 = [283] Title = DG load cells weight out of range ListTitle = DG Fault: LC Range Message = If the weight of the DG load cells > 4500.0 g or < 0.0 g. SRSDG 615 = [284] Title = DG load cells primary/back up drift out of range ListTitle = DG Fault: Redundant LC Check Message = If redundant load cells deviate by more than 60g after tare for 5 seconds. SRSDG 383 = [285] Title = DG RO permeate sample started ListTitle = Cleaning: RO Permeate Sample In Progress Message = If DG RO permeate sample successfully started from UI. [286] Title = DG cap in wrong state ListTitle = Cleaning: Conc Cap Removed Message = If the concentrate cap is not in the right position in the cleaning modes. [287] Title = HD processor clock speed checks against FPGA clock failure ListTitle = HD Fault: Proc- FPGA Clock Message = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. [288] Title = HD load cell inconsistency ListTitle = DG Fault: LC Drift Check Message = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. [289] Title = DG cap in wrong state ListTitle = Cleaning: Dialysate Cap Removed Message = If the dialysate cap is not in the right position in the cleaning modes. SRSDG 928 = [290] Title = HD venous sensor FPGA fault ListTitle = HD Fault: Ven. Pres. Com Message = If the FPGA does not report fresh data for a certain period of time [291] Title = DG flush started ListTitle = Cleaning: Flush In Progress Message = If DG flush successfully started from the UI. [292] Title = DG heat disinfect started ListTitle = Cleaning: Heat Disinf In Progress Message = If DG heat disinfect successfully started from the UI. [293] Title = DG chem disinfect started ListTitle = Cleaning: Chem Disinfect In Progress Message = If DG flush successfully started from the UI. [294] Title = DG active cool started ListTitle = Cleaning: Active Cool In Progress Message = If DG active cool successfully started. [295] Title = UI cloud sync POST failure ListTitle = HD POST: UI CloudSync Message = If the UI POST reports cloud sync failure. = SRSUI 1238 [296] Title = DG TDi sensor FPGA fault ListTitle = DG Fault: TDi RTD Comm Message = If the FPGA does not report fresh data for a certain period of time SRSDG 951 = [297] Title = DG TRo sensor FPGA fault ListTitle = DG Fault: TRo RTD Comm Message = If the FPGA does not report fresh data for a certain period of time SRSDG 950 = [298] Title = DG barometric sensor FPGA fault ListTitle = DG Fault: Baro Pres Comm Message = If the FPGA does not report fresh data for a certain period of time SRSDG 948 = [299] Title = DG invalid serial number ListTitle = DG POST: Serial Number Check Message = If the DG serial number is empty and does not have any characters. SRSDG 952 = [300] Title = HD invalid serial number ListTitle = HD POST: Serial Number Check Message = If the HD serial number is empty and does not have any characters. [301] Title = Open dialysate cap to be able to collecte the dispensing fluid ListTitle = Cleaning: RO Permeate Sample Connect Cap Message = RO permeate sample collection state [302] Title = DG_CPU RAM error ListTitle = DG Fault: Processor RAM error Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. SRSDG 126 = [303] Title = DG drain pump direction FPGA fault ListTitle = DG Fault: DRP Hall Commutation Message = If the FPGA indicates an error for a certain period of time. SRSDG 958 = [304] Title = HD invalid usage reocord ListTitle = HD POST: Usage Rec NVRAM CRC Message = If the CRC of the HD usage record does not match the calculated CRC. [305] Title = HD_CPU RAM error ListTitle = HD Fault: Processor RAM error Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. [306] Title = ListTitle = DG Fault: CD1 CD2 Check Message = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other [307] Title = HD AC power lost alarm ListTitle = Blood Stop: Loss of AC Power Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. [308] Title = DG CPi conductivity sensor transmits invalid characters ListTitle = DG Fault: CPi Comm Message = If the CPi conductivity sensor transmits invalid characters for a certain period of time. SRSDG 959 = [309] Title = DG CPo conductivity sensor transmits invalid characters ListTitle = DG Fault: CPo Comm Message = If the CPo conductivity sensor transmits invalid characters for a certain period of time. SRSDG 960 = [310] Title = DG CD1 conductivity sensor transmits invalid characters ListTitle = DG Fault: CD1 Comm Message = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. SRSDG 962 = [311] Title = DG CD2 conductivity sensor transmits invalid characters ListTitle = DG Fault: CD2 Comm Message = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. SRSDG 961 = [312] Title = The dialysate inlet flow is too low. ListTitle = Dial. Gen: Dialysate In Flow Message = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. [313] Title = DG cleaning mode inlet water temperature too low ListTitle = Cleaning: Inlet Water Temp Low Message = If inlet water temperature is < 24 C. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] Title = DG cleaning mode inlet water pressure too high ListTitle = Cleaning: Inlet Water Press High Message = If inlet water temperature is > 80 psig. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] Title = DG cleaning mode inlet water pressure too low ListTitle = Cleaning: Inlet Water Press Low Message = If inlet water pressure is < 14 psig. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] Title = Acid concentrate pump did not park successfully ListTitle = DG Fault: Acid CP Pump Park Message = If the park took more than 1 second. SRSDG 983 = [317] Title = Bicarb concentrate pump did not park successfully ListTitle = DG Fault: Bicarb CP Pump Park Message = If the park took more than 1 second. SRSDG 984 = [318] Title = HD Air Pump operation timeout ListTitle = HD Fault: Air Pump Timeout Message = If air pump has been turned on for greater than 10s. [319] Title = Collect sample to check the chemical disinfectant acid ListTitle = Cleaning: Chem Residual Sampling Message = Chemical disinfect flush sample alarm is prompted. SRSDG 1040 = [320] Title = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes ListTitle = Cleaning: Chem Residual Timeout Message = Chemical disinfect flush sample was not collected form more than 30 minutes. SRSDG 1043 = [321] Title = Outlet reactor not healthy ListTitle = DG Fault: UVo Error Message = If the outlet reactor are not healthy. SRSDG 1069 = [322] Title = UI POST Invalid Year ListTitle = HD POST Alarm: UI Invalid Year Message = If the UI POST reports a invalid year. [323] Title = Service Required: Hemodialysis Device ListTitle = HD POST Alarm: UI Bad Settings Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.