[ALARM_ID_NO_ALARM] Title = Indication for when no alarms is a possible situation ListTitle = Message = [ALARM_ID_TD_SOFTWARE_FAULT] Title = Service Required: Hemodialysis Device ListTitle = TD Software fault. Software found itself in an unexpected state Message = TD Fault: SW Invalid State [ALARM_ID_DD_SOFTWARE_FAULT] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state Message = DD Fault: Software Invalid State [ALARM_ID_RO_SOFTWARE_FAULT] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state Message = DD Fault: Software Invalid State [ALARM_ID_TD_FPGA_POST_TEST_FAILED] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA POST failure Message = TD POST: FPGA Compatibility [ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED] Title = Service Required: Hemodialysis Device ListTitle = TD Watchdog POST failure Message = TD POST: WD Safestate [ALARM_ID_TD_UI_COMM_POST_FAILED] Title = Service Required: Hemodialysis Device ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up Message = TD POST: Proc- UI Comm [ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE] Title = Service Required: Hemodialysis Device ListTitle = TD processor clock speed checks against FPGA clock failure Message = TD Fault: Proc- FPGA Clock [ALARM_ID_TD_FPGA_COMM_TIMEOUT] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA communication down for too long Message = TD Fault: FPGA Com [ALARM_ID_TD_UI_COMM_TIMEOUT] Title = Service Required: Hemodialysis Device ListTitle = User interface communication timeout Message = TD Fault: UI Comm [ALARM_ID_TD_DD_COMM_TIMEOUT] Title = Communication Error ListTitle = Dialysate generator communication timeout Message = Loss of DD Comm (Timeout) [ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS] Title = Service Required: Hemodialysis Device ListTitle = Too many bad CRCs detected on received system messages Message = TD Fault: CAN CRC Com [ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI] Title = Service Required: Hemodialysis Device ListTitle = System message that required acknowledgment was not acknowledged Message = TD Fault: CAN ACK UI Com [ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD] Title = Communication Error ListTitle = TD to DD comm error. Message = Loss of DD Comm (DD ACK) [ALARM_ID_TD_CRITICAL_DATA_ERROR] Title = Service Required: Hemodialysis Device ListTitle = TD critical data integrity check failed Message = TD Fault: Critical Data Err [ALARM_ID_DD_CRITICAL_DATA_ERROR] Title = Service Required: Dialysate Device ListTitle = DD critical data integrity check failed Message = DD Fault: Critical Data Err [ALARM_ID_TD_WATCHDOG_EXPIRED] Title = Service Required: Hemodialysis Device ListTitle = TD watchdog expired Message = TD Fault: WD Error [ALARM_ID_DD_WATCHDOG_EXPIRED] Title = Service Required: Dialysate Device ListTitle = DD watchdog expired Message = DD Fault: WD Error [ALARM_ID_DD_COMM_TOO_MANY_BAD_CRCS] Title = Service Required: Dialysate Device ListTitle = Too many bad CRCs detected on received system messages Message = DD Fault: CAN CRC Com [ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_RO] Title = Service Required: Dialysate Device ListTitle = DD to RO comm error Message = Loss of RO Comm (RO ACK) [ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD] Title = Service Required: Dialysate Device ListTitle = DD to TD comm error Message = Loss of TD Comm (TD ACK) [ALARM_ID_DD_FPGA_POST_TEST_FAILED] Title = Service Required: Dialysate Device ListTitle = DD FPGA POST failure Message = DD POST: FPGA Compatibility [ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE] Title = Service Required: Dialysate Device ListTitle = FPGA clock speed failure Message = DD Fault: Proc- FPGA Clock [ALARM_ID_DD_FPGA_COMM_TIMEOUT] Title = Service Required: Dialysate Device ListTitle = DD FPGA communication down for too long Message = DD Fault: FPGA Com [ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE] Title = Service Required: Hemodialysis Device ListTitle = TD alarm audio failed POST Message = TD POST: Primary Alarm Current [ALARM_ID_DD_VALVE_CONTROL_FAILURE] Title = Service Required: Dialysate Device ListTitle = DD FPGA not accepting commanded valve states Message = DD Fault: Valve Control [ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor reporting error or not being read successfully Message = TD Arterial Pressure Sensor Error [ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor reporting error or not being read successfully Message = TD Venous Pressure Sensor Error [ALARM_ID_DD_HYD_OUTLET_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully Message = DD Hyd outlet Pressure Sensor Error [ALARM_ID_DD_BIBAG_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD bibag pressure sensor reporting error or not being read successfully Message = DD Bibag Pressure Sensor Error [ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pressure sensor reporting error or not being read successfully Message = DD Spent Dial Pressure Sensor Error [ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully Message = DD Fresh Dial Pressure Sensor Error [ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD transmembrane sensor reporting error or not being read successfully Message = DD Transmemb Pressure Sensor Error [ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE] Title = Service Required: Hemodialysis Device ListTitle = TD monitored voltage is out of range Message = TD Fault: Voltage Monitor [ALARM_ID_TD_AC_POWER_LOST] Title = Power Loss ListTitle = TD AC power lost alarm Message = Loss of AC Power [ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT] Title = Power Loss ListTitle = TD AC power lost alarm Message = Blood Stop: Loss of AC Power [ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error [ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error [ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error [ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully Message = DD Conductivity Sensor Error [ALARM_ID_TD_VENOUS_BUBBLE_DETECTED] Title = Venous Air Detected ListTitle = TD venous air bubble detected alarm Message = Blood Stop: Venous Air Bubble [ALARM_ID_TD_CARTRIDGE_DOOR_OPENED] Title = Blood Pump Stopped: Door Open ListTitle = TD cartridge door opened alarm Message = Blood Stop: Door Open [ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED] Title = Stuck Button ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up Message = TD POST: Stuck Button [ALARM_ID_TD_ARTERIAL_PRESSURE_LOW] Title = Low Arterial Pressure ListTitle = Arterial pressure too low during treatment Message = Blood Stop: Arterial Pres. Low [ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH] Title = High Arterial Pressure ListTitle = Arterial pressure too high during treatment Message = Blood Stop: Arterial Pres. High [ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor is reading out of range Message = TD Fault: Arterial Press Range [ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE] Title = Self Test: Cartridge ListTitle = TD arterial pressure self-test failure alarm Message = Process: Pre- Tx Remove Cartridge [ALARM_ID_TD_VENOUS_PRESSURE_LOW] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment Message = Blood Stop: Venous Pres. Low [ALARM_ID_TD_VENOUS_PRESSURE_HIGH] Title = High Venous Pressure ListTitle = Venous pressure too high during treatment Message = Blood Stop: Venous Pres. High [ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor is reading out of range Message = TD Fault: Venous Press Range [ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE] Title = Self Test: Cartridge ListTitle = TD venous pressure self-test failure alarm Message = Process: Pre- Tx Remove Cartridge [ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE] Title = Self Test: Cartridge Check ListTitle = TD pre-treatment mode dry self-tests failure Message = Process: Pre- Tx Dry Self- Test [ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D11_PUMP speed control error Message = DD Fault: Pump Speed Fault [ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D10_PUMP speed control error Message = DD Fault: Pump Speed Fault [ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump hall sensor out of range. Message = DD Fault: Pump Speed Fault [ALARM_ID_DD_CONCENTRATE_PUMP_FAULT] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump fault Message = DD Fault: Pump Fault [ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD temperature sensor out of range Message = DD Fault: Temp Sensor Range [ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD temperature sensors not being read successfully Message = DD Fault: Temperature RTD Comm [ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD barometric sensor not being read successfully Message = DD Fault: Baro temp Comm [ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC] Title = Service Required: Dialysate Device ListTitle = DD barometric pressure sensor bad coefficients CRC Message = DD Fault: Baro temp Cal CRC [ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump RPM out of range Message = DD Fault: Fresh Dial Pump Speed Fault [ALARM_ID_DD_D12_PUMP_OFF_FAULT] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump off fault Message = DD Fault: Fresh Dial Pump Speed Not Off [ALARM_ID_DD_D12_PUMP_CURRENT_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump current out of range Message = DD Fault: Fresh Dial Pump Current [ALARM_ID_DD_D12_PUMP_DIRECTION_INVALID] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump invalid direction Message = DD Fault: Fresh Dial Pump Dir Fault [ALARM_ID_DD_D12_PUMP_DIRECTION_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump direction FPGA fault Message = DD Fault: Fresh Dial pump Dir Err Cnt [ALARM_ID_DD_D48_PUMP_RPM_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump RPM out of range Message = DD Fault: Spent Dial Pump Speed Fault [ALARM_ID_DD_D48_PUMP_OFF_FAULT] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump off fault Message = DD Fault: Spent Dial Pump Speed Not Off [ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump current out of range Message = DD Fault: Spent Dial Pump Current [ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump invalid direction Message = DD Fault: Spent Dial Pump Dir Fault [ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump direction FPGA fault Message = DD Fault: Spent Dial pump Dir Err Cnt [ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON] Title = Service Required: Dialysate Device ListTitle = DD D5 heater is On with low fluid level Message = DD Fault: Heater On With No Fluid [ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON] Title = Service Required: Dialysate Device ListTitle = DD D45 heater is On with low fluid level Message = DD Fault: Heater On With No Fluid [ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD water inlet input pressure sensor reporting error or not being read successfully Message = DD Water Inlet Input Press Sensor Error [ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT] Title = Service Required: Dialysate Device ListTitle = DD water inlet output pressure sensor reporting error or not being read successfully Message = DD Water Inlet Output Press Sensor Error [ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE] Title = Incoming Water Pressure Input Low ListTitle = DD Inlet water pressure input too low Message = DD: Inlet Water Pres Input Low [ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE] Title = Incoming Water Pressure Input High ListTitle = DD inlet water pressure input is too high Message = DD: Inlet Water Pres Input High [ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE] Title = Incoming Water Pressure Output Low ListTitle = DD Inlet water pressure output too low Message = DD: Inlet Water Pres Output Low [ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE] Title = Incoming Water Pressure Output High ListTitle = DD inlet water pressure output is too high Message = DD: Inlet Water Pres Output High [ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap fill timeout during treatment Message = Blood Stop: Air Trap Fill [ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS] Title = Service Required: Hemodialysis Device ListTitle = Air trap level sensors reporting illegal combination of air/fluid Message = TD Fault: Air Trap Sensor [ALARM_ID_TD_PINCH_VALVE_FAULT] Title = Service Required: Hemodialysis Device ListTitle = Pinch valve reporting a fault Message = TD Fault: Pinch Valve [ALARM_ID_TD_VALVE_HOMING_FAILED] Title = Service Required: Hemodialysis Device ListTitle = TD valve homing failed Message = TD Fault: Valve Home [ALARM_ID_TD_VALVE_TRANSITION_TIMEOUT] Title = Service Required: Hemodialysis Device ListTitle = TD valve transition time out Message = TD Fault: Valve Transition Err [ALARM_ID_RO_CAN_MESSAGE_NOT_ACKED_BY_DD] Title = Communication Error ListTitle = DD to RO comm error. Message = RO Loss of DD Comm (DD ACK) [ALARM_ID_RO_FPGA_COMM_TIMEOUT] Title = Service Required: Dialysate Device ListTitle = RO FPGA communication down for too long Message = RO Fault: FPGA Comm [ALARM_ID_RO_COMM_TOO_MANY_BAD_CRCS] Title = Service Required: Hemodialysis Device ListTitle = RO Too many bad CRCs detected on received system messages Message = RO Fault: CAN CRC Comm [ALARM_ID_RO_VALVE_CONTROL_FAILURE] Title = Service Required: Filtration Device ListTitle = RO FPGA not accepting commanded valve states Message = RO Fault: Valve Control [ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill start pressure is not in range Message = DD Fault: BC Pressure not in range [ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill end pressure is not in range Message = DD Fault: BC Pressure not in range [ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill start pressure is not in range Message = DD Fault: BC Pressure not in range [ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill end pressure is not in range Message = DD Fault: BC Pressure not in range [ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD hydraulics chamber degassing pressure is not in range Message = DD Fault: Hyd Cham Neg Pres not in range [ALARM_ID_AVAILABLE_92] Title = ListTitle = Message = [ALARM_ID_AVAILABLE_93] Title = ListTitle = Message = [ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range Message = DD Fault: BC Pressure not in range [ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range Message = DD Fault: BC Pressure not in range [ALARM_ID_TD_TREATMENT_STOPPED_BY_USER] Title = Blood Pump Stopped: Treatment Paused ListTitle = TD treatment stopped by user action - pressed stop button Message = Blood Stopped: Treatment Pause [ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error [ALARM_ID_AVAILABLE_98] Title = ListTitle = Message = [ALARM_ID_AVAILABLE_99] Title = ListTitle = Message = [ALARM_ID_TD_COMM_TIMEOUT] Title = Service Required: Hemodialysis Device ListTitle = TD communication timeout Message = TD Fault: UI- Proc Comm [ALARM_ID_UI_POST_FAILURE_OS_VERSION] Title = Service Required: Hemodialysis Device ListTitle = UI POST OS Version compatibility failure Message = TD POST: OS Version [ALARM_ID_UI_POST_FAILURE_SHASUM] Title = Service Required: Hemodialysis Device ListTitle = UI POST Application Integrity (Sha256Sum) failure Message = TD POST: UI Image [ALARM_ID_UI_POST_FAILURE_CANBUS] Title = Service Required: Hemodialysis Device ListTitle = UI POST CANBus failure Message = TD POST: UI CAN Driver [ALARM_ID_UI_POST_FAILURE_DISPLAY] Title = Service Required: Hemodialysis Device ListTitle = If the UI POST reports display failure. Message = UI POST Display failure [ALARM_ID_UI_POST_FAILURE_TOUCH] Title = Service Required: Hemodialysis Device ListTitle = UI POST Touch failure Message = TD POST: UI Touch Driver [ALARM_ID_UI_POST_FAILURE_SDCARD] Title = Service Required: Hemodialysis Device ListTitle = UI POST SD-Card failure Message = TD POST: UI SD- Card [ALARM_ID_UI_POST_FAILURE_RTC] Title = Service Required: Hemodialysis Device ListTitle = UI POST RTC failure Message = TD POST: UI RTC [ALARM_ID_UI_POST_FAILURE_WIFI] Title = Service Required: Hemodialysis Device ListTitle = UI POST WiFi failure Message = TD POST: UI WiFi [ALARM_ID_UI_POST_FAILURE_BLUETOOTH] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bluetooth failure Message = TD POST: UI Bluetooth [ALARM_ID_UI_POST_FAILURE_ETHERNET] Title = Service Required: Hemodialysis Device ListTitle = UI POST Ethernet failure Message = TD POST: UI Ethernet [ALARM_ID_UI_POST_FAILURE_SOUND] Title = Service Required: Hemodialysis Device ListTitle = UI POST Sound failure Message = TD POST: UI Sound [ALARM_ID_UI_POST_FAILURE_CLOUDSYNC] Title = Service Required: Hemodialysis Device ListTitle = UI POST cloud sync POST failure Message = TD POST: UI CloudSync [ALARM_ID_UI_POST_FAILURE_INVALID_YEAR] Title = Service Required: Hemodialysis Device ListTitle = UI POST Invalid Year Message = TD POST Alarm: UI Invalid Year [ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bad Settings Message = TD POST Alarm: UI Bad Settings [ALARM_ID_UI_EVNT_FAILURE_SDCARD] Title = Service Required: Hemodialysis Device ListTitle = SD card is non-operational Message = TD Fault: UI- SD- Card Fault [ALARM_ID_UI_RESERVED_116] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_117] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_118] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_119] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_120] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_121] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_122] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_123] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_124] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_125] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_126] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_127] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_128] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_129] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_UI_RESERVED_130] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED [ALARM_ID_BATTERY_LOW] Title = Battery Capacity Low ListTitle = Insufficient battery charge to support 20 min without A/C power Message = TD Alarm (Low): Battery Low