DEN-9418: Code Review 0

  1. … 6 more files in changeset.
DEN-9418: Code Review 0

  1. … 6 more files in changeset.
DEN-9614: Meetings & Support - General Support

Bug_Fixes :

- Fixed the issue in the log having null characters in the version message

- Fixing the issue with the bottom main menu not showing up after navigating back and forth from home screen.

- Removed the touchable property from the Treatment Review screen items.

New_Features :

-

Known_Bugs :

- N/A [Initial push]

Changes :

- Adding the acknowledge directly sending to the DG device by UI.

  1. … 13 more files in changeset.
DEN-9416: Implementations

- Added a newly added parameter of the BloodPrime (Safety Vol).

- UI POST is done.

- Out of 10 tests

1 - CANBus

2 - Touch Screen

3 - SD-Card

4 - RTC

tests are done.

- The

5 - File System integrity

will be done as part of CyberSecurity.

- The

6 - Bluetooth

will be added when the feature itself added to the application.

- The

7 - WiFi

is in progress and having issue correctly detect it.

- Some tests are listed to be added later in case required, like

8 - Display

9 - Ethernet

10- Sound

test couldn't find a way to test it yet.

- On each unsuccessful test an alarm will be triggered.

- UI alarms are in range of [250 - 259]

  1. … 18 more files in changeset.
DEN-9416: Implementation

- Added the Ethernet IP address on top of the screen for later easier test.

  1. … 4 more files in changeset.
DEN-9416: Implementation

DEN-9299: Adding the Syringe Pump data on Diagnostic

  1. … 24 more files in changeset.
DEN-9416: Implementations

- Most of the FW implementation is done.

- After the Check-in message two more messages needs to be sent to the HD to complete the POST:

- Version for the compatibility test.

- UI_POST to wrap the POST on HD side.

  1. … 14 more files in changeset.
DEN-9416: Implementations

- Added the UI Version values to the message 28.

  1. … 14 more files in changeset.
DEN-9142: Meetings & Support - General Support

- Added Dialin channels as "Debug" channels in the UI Log for V&V canbus testing (All were Unexpected before.)

- Added label for the Diagnostics screen

- The SD-CArd Indicator's mouse area made bigger to easier double touch it.

  1. … 9 more files in changeset.
DEN-8910: Meetings & Support - General Support

- Added a threading for the screenshot save

- Tested on the device and modified foldering.

  1. … 7 more files in changeset.
DEN-8910: Meetings & Support - General Support

- Active Alarm List toggle while the minimize is disables.

- Wrapped up the Serial information in settings

- Wrapped up the Service information in settings

- Fixed a minor issue in logging threading which was adding the log header in main thread and added more threading check for Logger.

- Added Screenshot functionality to the application by F12 on keyboard to save on usb drive "screenshot" folder with <yyyyMMdd-HHmmss>.png file format on device and on Desktop/Screenshot on desktop version.

  1. … 50 more files in changeset.
DEN-8693: Meetings & Support - General Support

- Updated the Alarm Mapping to add Alarm index on each line as comment.

- Added Flush States

- Fixed message 0x7700 length error.

  1. … 9 more files in changeset.
DEN-8693: Support for the disinfect

- Adding support for reading the brightness value from the Linux file system instead of having another file just to keep the same value.

- Fixed a bug in the Active Alarm List request.

- Adding the functionality of the Alarm Volume.

- Modified the Disinfection screens[Heat/chem] check marks with the new bundled states provided by FW for UI and the flush states are yet to come.

- Adjusted the Active Alarm List screen a little to be able to show all the current Alarm messages without cropping.

  1. … 26 more files in changeset.
DEN-8533 & DEN-8693

Merge and support

  1. … 44 more files in changeset.
DEN-8519: Implementations [MVC]

  1. … 15 more files in changeset.
DEN-8494: Implementation [ Service Log in & About ]

  1. … 19 more files in changeset.
DEN-8490: Meetings & Support - General Support

- Support : Instruction for the Usability test support

- Support : SW team support and adding the second DGOpMode parameter (SubMode) fro logging.

- Support : SW team support, changed the Alarm Status log to be data from event.

- Added another parameter for treatment log - alarm

- Removed the initialization logging from the event logs which are all CANBus messages, in to the debug log .

- Added a header for the event log to be easier export in the excel sheet (the header doesn't have timestamp)

-

  1. … 51 more files in changeset.
DEN-8559: TxLog: Implementation [The Average 30min period log]

  1. … 16 more files in changeset.
DEN-8559: TxLog: Implementation [The Average 30min period log]

- Implemented the Models

  1. … 16 more files in changeset.
DEN-8317: Implementation: Merge the Wi-Fi into the code.

- Mainly the WiFi integration

- Some other support tasks:

| Support : Instruction and Kishen images [ Calyton, Nick ]

| Support : Alarm Messaging Parser

| Support : Changed the Manager Screen into a Monitoring Dialog and can be shown by DblTap on SDCard Indication on top/right

| Support : Modified the Heat values to float on Monitoring Dialog

| Support/Bug : Fixed the Alarm Message Bug

  1. … 107 more files in changeset.
DEN-8316: TxLog: Implementation [Simulator/Testing]

  1. … 5 more files in changeset.
DEN-8316: TxLog: Implementation [Simulator/Testing]

  1. … 53 more files in changeset.
DEN-7835: Code Review 0

- Removed extra test steps in the instruction.conf

- Updated the SDItem.qml code dox.

- Updated the storage variable for Date/Time script location

  1. … 6 more files in changeset.
DEN-7931 UI Logging: Rotation and deletion do not trigger on disk usage.

- Fixed and gave to the V&V team, Michael, to test.

- First round of test was successful.

  1. … 13 more files in changeset.
DEN-7831: posTx: Implementation [Simulator/Testing]

- Disinfection

  1. … 39 more files in changeset.
DEN-7831: posTx: Implementation [Simulator/Testing]

- As part of the test and simulation treatment log was implemented at some point.

- It still needs to convert received values and map some other type values and dates.

  1. … 9 more files in changeset.
DEN-7946 Saline, Standby, Alarm, Keyboard, ....

  1. … 21 more files in changeset.
DEN-8052: posTx: Implementation [Messaging] (Disinfection)

DEN-8067: Meetings & Support - Post-Treatment Integration

  1. … 17 more files in changeset.
DEN-8052: posTx: Implementation [Messaging] (Disinfection)

- Disinfection MVC

- There is still a state message and also some parts in qml for the progress and states during the disinfection.

  1. … 38 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Other than the comments bellow, we talked about the "--" was not correct which has been fixed.

1) The default value should not be shown on the slider that's not been configured yet

- Not sure that is has to be done but regardless, default values were not correct and fixed.

2) Dialyzer Temperature should show degree Celcius, not just C

- Dialyzer Temperature now has degree circle

3) Discuss whether to keep the manufacturing default on the slider since user needs to confirm each parameter

4) Concentrates and Dialyzer types should match with SRS /PRS

- Has been updated regarding the latest updates (BiCarb:PRS351, Dilyzer:PRS352)

5) Parameter value bounds back and forth on the slider during adjustment (display touch sensitivity issue)

6) Suggest to add + or - on the slider to provide ability to adjust precisely

7) Arterial Pressure and Venous Pressure should be consistent with other parameters, unit should be shown at the end

8) Suggest to display Arterial/Venous Pressure Limit as Arterial/Venous Low/High Pressure Limit and the current value as low/high mmHg

9) Suggest to display the current value of each parameter to the right of the slider

  1. … 7 more files in changeset.