DEN-12558: HD does not transition to service mode

- implemented the service mode confirmation

- decreased the dialog animation duration

- renamed the service dates MVC to differentiate with service mode.

  1. … 32 more files in changeset.
DEN-12541: Get the SD-Card size from the system

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 158 more files in changeset.
DEN-12512: INT: Support/Meetings/Planning

DEN-09712 Date format between system date/time and BP/HP "Last Read" is not consistent

DEN-09929 Treatment Log shows "?C" instead of "°C"

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

  1. … 14 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

- Reorganized the Diagnostics screen a little more.

  1. … 2 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

  1. … 11 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

- Reorganized the Diagnostics screen a little more.

  1. … 3 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

  1. … 12 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 570 more files in changeset.
UI-BUG-S63: UI-BUG-S63-2: Addressed the code review comments.

DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 11 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- Added a script to later be used for the log export on the device, but it now only works on the VM since the device doesn't have the zip tool.

- Added a mimic function to simulate the BCuff measurement received and assigned F11 to it and it will only be enabled if the macro DEBUG_BCUFF_MIMIC is defined.

- Added opened and closed signals to the EntryDialog.qml to make it do the auto focus on the firstInput object when opened.

- Added setFocus() function in the BPHREntry.qml

- Removed the enabled option from doReset of the VTreatmentVitals

- And created two separate enableDialog for vital entry, and enable update for the auto update the properties.

- Added a constructor with systolic, diastolic, heartRate parameter to set the data on creation to be used in the message mimic.

- Added systolic, diastolic, heartRate parameters to the doTrigger of the VCommonAdjustmentVitals to open the vitals entry dialog with default values.

  1. … 15 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

  1. … 7 more files in changeset.
DEN-10607: Screen

- Added the feature for the create treatment slider to be able to show OFF for zero values on label.

- Did a refactoring to clean up the SliderCreateTreatment.

  1. … 2 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 11 more files in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 8 more files in changeset.
DEN-10399: UI: Alarms block Shutdown screen

DEN-9271: Dialysate Temperature implemented resolution does not meet PRS

  1. … 6 more files in changeset.
DEN-9639: Code Review 0

- Addressed code reviews

- Reoved LOG_DEBUG_UI since the DEBUG log is all about UI and adding a UI to it was useless.

- Removed the time stamp from the POST start/stop log since the log itself has timestamp.

- Added more documentation to the ApplicationPOST class.

- Renamed the FileSystem to ShaSum check in the ApplicationPOST.

- Updated the WiFi POST check funciton.

- Updated the initial Screen to pause the animation when the UI navigates out of that screen when we are in test (usual case should not happen since we fail or pass the POST and either stops the animation.)

- Updated the AlarmMapping to reflect the FILESYSTEM error change to SHASUM error.

  1. … 19 more files in changeset.
[ DEN-9653 , DEN9655 ] : Implementation - Settings [ Screen , MVC ]

  1. … 17 more files in changeset.
DEN-9418: Code Review 0

- Cherry picked the code review updated from post branch

- Correcting the message IDs enum comments.

  1. … 3 more files in changeset.
DEN-9418: Code Review 0

  1. … 4 more files in changeset.
DEN-9649: Implementation - Yocto configuration

- Addressed some frequent code review comments

- Fixed the UI POST AlarmTriggered message

- Added the deleteLater for the QQuickViewer pointer _viewer.

  1. … 113 more files in changeset.
DEN-9614: Meetings & Support - General Support

Bug_Fixes :

- Fixed the issue in the log having null characters in the version message

- Fixing the issue with the bottom main menu not showing up after navigating back and forth from home screen.

- Removed the touchable property from the Treatment Review screen items.

New_Features :

-

Known_Bugs :

- N/A [Initial push]

Changes :

- Adding the acknowledge directly sending to the DG device by UI.

  1. … 13 more files in changeset.
DEN-9416: Implementation

DEN-9299: Adding the Syringe Pump data on Diagnostic

  1. … 24 more files in changeset.
DEN-9416: Implementations

- Most of the FW implementation is done.

- After the Check-in message two more messages needs to be sent to the HD to complete the POST:

- Version for the compatibility test.

- UI_POST to wrap the POST on HD side.

  1. … 14 more files in changeset.
DEN-9142: Meetings & Support - General Support

- Made the Diagnostic screen labels updated with the given labels from FW and increased the font size.

  1. … 1 more file in changeset.
Code Review Addressed

  1. … 1 more file in changeset.
DEN-9142: Meetings & Support - General Support

- Added Dialin channels as "Debug" channels in the UI Log for V&V canbus testing (All were Unexpected before.)

- Added label for the Diagnostics screen

- The SD-CArd Indicator's mouse area made bigger to easier double touch it.

  1. … 9 more files in changeset.