components

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-13489 IMP: General Implementation

DEN-13035: Update alarm body messages

- Removed the exclamation mark icon form the Alarm Dialog.

- Removed the exclamation mark icon form the Alarm Bar.

- Made the whole AlarmDialog title bar doubleClick and diagnostics opener if the alarm is not minimizable.

- Moved the Alarm ID from title to the bottom left of the dialog with format: "ID:<id>".

- Clipped the AlarmDialon body message area, crop the text if longer than dialog area.

- Changed the "settings" property name of the VSettings to "instructions".

- Changed the date/time formats keys to simplify the code and conf file.

- Implemented the Alarm Body messages, to load from Alams/Alarms.conf, instead of the FW.

- The FW texts/documents will still being used if the Alarms.conf doesn't have a definition.

- A instruction base capability has also been added to the MVC, but UI still not supporting it.

  1. … 18 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- A bug found on multiple, ON,OFF toggle, slider sends minimum instead of '0' while ON.

- Slider improved.

- Added float rounding of the request message logs to 3.

  1. … 4 more files in changeset.
DEN-9129: Unable to proceed past "Create a Custom Treatment" when Arterial and Venous Pressure (High) Limits maximum values are initially selected

Also modified the SliderCreateTreatment.qml to show by the default the zero value not the OFF since it is only two Heparin Bolus/Dispensing which display OFF for zero value.

  1. … 2 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 4 more files in changeset.
DEN-13127: LTS: Development Support

- [Export Button Enabling in VM #61|https://github.com/Diality/testsuites/issues/61]

  1. … 8 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 40 more files in changeset.
DEN-12558: HD does not transition to service mode

- implemented the service mode confirmation

- decreased the dialog animation duration

- renamed the service dates MVC to differentiate with service mode.

  1. … 32 more files in changeset.
DEN-12541: Get the SD-Card size from the system

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 158 more files in changeset.
DEN-12512: INT: Support/Meetings/Planning

DEN-09712 Date format between system date/time and BP/HP "Last Read" is not consistent

DEN-09929 Treatment Log shows "?C" instead of "°C"

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

  1. … 14 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

- Reorganized the Diagnostics screen a little more.

  1. … 2 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

  1. … 11 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

- Reorganized the Diagnostics screen a little more.

  1. … 3 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

  1. … 12 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 570 more files in changeset.
UI-BUG-S63: UI-BUG-S63-2: Addressed the code review comments.

DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 11 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- Added a script to later be used for the log export on the device, but it now only works on the VM since the device doesn't have the zip tool.

- Added a mimic function to simulate the BCuff measurement received and assigned F11 to it and it will only be enabled if the macro DEBUG_BCUFF_MIMIC is defined.

- Added opened and closed signals to the EntryDialog.qml to make it do the auto focus on the firstInput object when opened.

- Added setFocus() function in the BPHREntry.qml

- Removed the enabled option from doReset of the VTreatmentVitals

- And created two separate enableDialog for vital entry, and enable update for the auto update the properties.

- Added a constructor with systolic, diastolic, heartRate parameter to set the data on creation to be used in the message mimic.

- Added systolic, diastolic, heartRate parameters to the doTrigger of the VCommonAdjustmentVitals to open the vitals entry dialog with default values.

  1. … 15 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

  1. … 7 more files in changeset.
DEN-10607: Screen

- Added the feature for the create treatment slider to be able to show OFF for zero values on label.

- Did a refactoring to clean up the SliderCreateTreatment.

  1. … 2 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 11 more files in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 8 more files in changeset.
DEN-10399: UI: Alarms block Shutdown screen

DEN-9271: Dialysate Temperature implemented resolution does not meet PRS

  1. … 6 more files in changeset.
DEN-9639: Code Review 0

- Addressed code reviews

- Reoved LOG_DEBUG_UI since the DEBUG log is all about UI and adding a UI to it was useless.

- Removed the time stamp from the POST start/stop log since the log itself has timestamp.

- Added more documentation to the ApplicationPOST class.

- Renamed the FileSystem to ShaSum check in the ApplicationPOST.

- Updated the WiFi POST check funciton.

- Updated the initial Screen to pause the animation when the UI navigates out of that screen when we are in test (usual case should not happen since we fail or pass the POST and either stops the animation.)

- Updated the AlarmMapping to reflect the FILESYSTEM error change to SHASUM error.

  1. … 19 more files in changeset.
[ DEN-9653 , DEN9655 ] : Implementation - Settings [ Screen , MVC ]

  1. … 17 more files in changeset.
DEN-9418: Code Review 0

- Cherry picked the code review updated from post branch

- Correcting the message IDs enum comments.

  1. … 3 more files in changeset.
DEN-9418: Code Review 0

  1. … 4 more files in changeset.
DEN-9649: Implementation - Yocto configuration

- Addressed some frequent code review comments

- Fixed the UI POST AlarmTriggered message

- Added the deleteLater for the QQuickViewer pointer _viewer.

  1. … 113 more files in changeset.