DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2
- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.
- A minor update on the POST progress display to start when 0 to max of complete enum -1.
- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]