Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-8145: Issues with UI Create Treatment screen

- Restricted the Patient ID entry to be max 20 characters and has to have at least 1 alphanumeric character without space, regarding the conversation with KeborMed.

  1. … 2 more files in changeset.
Code Cleanup

- DEN-11807: Treatment Confirmation - Rinseback Rate displays "min" instead of "mL/min"

- Renamed the Keyboard component name to VirtualKeyboard to resolve the InputPanel's Keyboard object/type conflict (There was no issue is was only a name conflict.)

- Preventing the Patient Id from being empty space string, and disabling the confirm in that case.

- Mode the create_update_folder.sh to the Scripts/local repository's folder.

- Added Qt logo and Diality Qt Colored log.

- Updated the alarmMapping.sh script to always checkout to common/staging and also print out the log on console.

-

  1. … 12 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

  1. … 8 more files in changeset.
DEN-10607: Screen

- Added the feature for the create treatment slider to be able to show OFF for zero values on label.

- Did a refactoring to clean up the SliderCreateTreatment.

  1. … 2 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 10 more files in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 8 more files in changeset.
DEN-8490: Update the End Treatment Screens with the new Instruction base

- Also added the milliseconds to the logs

  1. … 39 more files in changeset.
DEN-8317: Implementation: Merge the Wi-Fi into the code.

  1. … 11 more files in changeset.
DEN-7756: Addressed code review comments

  1. … 15 more files in changeset.
DEN-8137: Confirm button in Confirm Treatment screen becomes disabled after enabled

- fixed the confirm button enabled.

  1. … 1 more file in changeset.
DEN-7755: Added device information table.

  1. … 18 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Other than the comments bellow, we talked about the "--" was not correct which has been fixed.

1) The default value should not be shown on the slider that's not been configured yet

- Not sure that is has to be done but regardless, default values were not correct and fixed.

2) Dialyzer Temperature should show degree Celcius, not just C

- Dialyzer Temperature now has degree circle

3) Discuss whether to keep the manufacturing default on the slider since user needs to confirm each parameter

4) Concentrates and Dialyzer types should match with SRS /PRS

- Has been updated regarding the latest updates (BiCarb:PRS351, Dilyzer:PRS352)

5) Parameter value bounds back and forth on the slider during adjustment (display touch sensitivity issue)

6) Suggest to add + or - on the slider to provide ability to adjust precisely

7) Arterial Pressure and Venous Pressure should be consistent with other parameters, unit should be shown at the end

8) Suggest to display Arterial/Venous Pressure Limit as Arterial/Venous Low/High Pressure Limit and the current value as low/high mmHg

9) Suggest to display the current value of each parameter to the right of the slider

  1. … 7 more files in changeset.
DEN-7147: Fixed a bug on arterial/venous pressures

- which was causing user to move low/high other than current default to be able to confirm for FW validation.

  1. … 2 more files in changeset.
DEN-7377: preTx: Implementation [Dynamic Instruction Loading] ( 2 )

DEN-7457: Meetings & Support - General purpose

  1. … 51 more files in changeset.
DEN-7408: Code improvement for navigation

- VTreatmentCreate is cleaned up. (still needs a lot of work)

  1. … 11 more files in changeset.
DEN-7408: Code improvement for navigation

- added vCondition to the page function for the stack item to pass the state property value.

- added fault , service and initial placeholder pages.

  1. … 14 more files in changeset.
DEN-7461: Meetings & Support - Pre-Treatment Integration

- Integration modifications

- UF Modifications request

- Saline Modification request

- Updated the validation mAccepted to be 1=true and 0=false.

- Heparin Stopped stated modification request

  1. … 29 more files in changeset.
DEN-7518: preTx: Implementation [Screens] ( 3 )

- The first round of the pre-treatment flow works.

  1. … 65 more files in changeset.
DEN-7518: preTx: Implementation [Screens] ( 3 )

  1. … 11 more files in changeset.
DEN-7139: preTx: Implementation [Screens] ( 2 )

  1. … 37 more files in changeset.
DEN-7143: preTx: Implementation [Navigation] ( 2 )

- Added the pre-treatment Consumables states

- Made the Step bullet multi orientation state

- Made the Step indicator multi orientation state

- Added the component WaitDone which has a rotating circle for wait state and a check mark when set as done.

- Added the consumables stack and two screens [self test is done, installation needs work]

- Updated PreTreatmentMainStack to navigate for waterSample and consumables.

- Added the Water Sample fail screen

- Added the Water Sample result screen

- Added the Water Sample stack

- Added the Water Sample, Sample Screen

- Added the STATE property type for the State boolean types which will not initiated if the value has not been changed.

-

    • -0
    • +70
    ./PreTreatmentCreateStack.qml
  1. … 46 more files in changeset.
DEN-7373: preTx: Implementation [Component][Steps Indicator Improvement]

- Modified the StepIndicator to work with index instead of text.

- Modified StepBullet as well for coloring and style.

- Fixed the Saline bolus multiple call in MModel.

  1. … 9 more files in changeset.
DEN-7373: preTx: Implementation [Component][Steps Indicator Improvement] ( ? )

- Improved the TouchRect as part of StepIndicator improvement.

  1. … 28 more files in changeset.
DEN-7142: Implementation [Navigation] ( 1 )

- A Complete refactoring of the navigation management has been done.

    • -125
    • +0
    ./PreTreatmentUltrafiltration.qml
  1. … 33 more files in changeset.
DEN-7142: preTx: Implementation [Navigation] ( 1 )

- Added HD_OP_MODE to navigate to the treatment home screen.

- Note : there is an issue with the create treatment implementation which won't let to reset the stack and needs to be fixed.

- Added some Rinseback modification requests.

  1. … 7 more files in changeset.
Merge master into staging

  1. … 34 more files in changeset.
DEN-5759: Code Review 0

- Added a TODO for the DoubleSlider component to add a validation and active feature to be able to use it for PreTreatmentCreate.

6636: ptTx_Rinseback: Implementation

- Added the PostTreatmentRinseback.qml

- Added the PostTreatmentRecirculate.qml

- Improved the StackItem.qml

- Added Rinseback/Recirculate Navigation

- Removed Diagnostics screen since it's not one screen anymore and is too old.

-

  1. … 22 more files in changeset.
DEN-5757: Code Coverage