DEN-13049 - UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- There was a bug in the fix which was related to the validation and caused to fail the treatment parameters validation.

- also removed the disable if the SD is not detected correctly. (QStorageInfo is not sending DeviceReady correctly)

- enabled the float rounding to 3 in the log

- Enabled the Create Treatment button rejection notification, since it seems it is not being handled by alarms as presumed.

- Fixed the BloodPressure interval value to display OFF when set to zero.

  1. … 4 more files in changeset.
DEN-9129: Unable to proceed past "Create a Custom Treatment" when Arterial and Venous Pressure (High) Limits maximum values are initially selected

Also modified the SliderCreateTreatment.qml to show by the default the zero value not the OFF since it is only two Heparin Bolus/Dispensing which display OFF for zero value.

  1. … 3 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

  1. … 5 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 44 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

  1. … 30 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 159 more files in changeset.
DEN-12512: INT: Support/Meetings/Planning

DEN-09712 Date format between system date/time and BP/HP "Last Read" is not consistent

DEN-09929 Treatment Log shows "?C" instead of "°C"

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

  1. … 14 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 77 more files in changeset.
DEN-12106: Heparin Delivery Active Text Does Not Appear

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 582 more files in changeset.
UI-BUG-S62: UI-BUG-S62-1

Addressed code reviews.

  1. … 5 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Restricted the Patient ID entry to be max 20 characters and has to have at least 1 alphanumeric character without space, regarding the conversation with KeborMed.

  1. … 2 more files in changeset.
Code Cleanup

- DEN-11807: Treatment Confirmation - Rinseback Rate displays "min" instead of "mL/min"

- Renamed the Keyboard component name to VirtualKeyboard to resolve the InputPanel's Keyboard object/type conflict (There was no issue is was only a name conflict.)

- Preventing the Patient Id from being empty space string, and disabling the confirm in that case.

- Mode the create_update_folder.sh to the Scripts/local repository's folder.

- Added Qt logo and Diality Qt Colored log.

- Updated the alarmMapping.sh script to always checkout to common/staging and also print out the log on console.

-

  1. … 12 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

  1. … 8 more files in changeset.
DEN-10607: Screen

- Added the feature for the create treatment slider to be able to show OFF for zero values on label.

- Did a refactoring to clean up the SliderCreateTreatment.

  1. … 2 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 10 more files in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 8 more files in changeset.
DEN-8490: Update the End Treatment Screens with the new Instruction base

- Also added the milliseconds to the logs

  1. … 39 more files in changeset.
DEN-8317: Implementation: Merge the Wi-Fi into the code.

  1. … 11 more files in changeset.
DEN-7756: Addressed code review comments

  1. … 15 more files in changeset.
DEN-8137: Confirm button in Confirm Treatment screen becomes disabled after enabled

- fixed the confirm button enabled.

  1. … 1 more file in changeset.
DEN-7755: Added device information table.

  1. … 18 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Other than the comments bellow, we talked about the "--" was not correct which has been fixed.

1) The default value should not be shown on the slider that's not been configured yet

- Not sure that is has to be done but regardless, default values were not correct and fixed.

2) Dialyzer Temperature should show degree Celcius, not just C

- Dialyzer Temperature now has degree circle

3) Discuss whether to keep the manufacturing default on the slider since user needs to confirm each parameter

4) Concentrates and Dialyzer types should match with SRS /PRS

- Has been updated regarding the latest updates (BiCarb:PRS351, Dilyzer:PRS352)

5) Parameter value bounds back and forth on the slider during adjustment (display touch sensitivity issue)

6) Suggest to add + or - on the slider to provide ability to adjust precisely

7) Arterial Pressure and Venous Pressure should be consistent with other parameters, unit should be shown at the end

8) Suggest to display Arterial/Venous Pressure Limit as Arterial/Venous Low/High Pressure Limit and the current value as low/high mmHg

9) Suggest to display the current value of each parameter to the right of the slider

  1. … 7 more files in changeset.
DEN-7147: Fixed a bug on arterial/venous pressures

- which was causing user to move low/high other than current default to be able to confirm for FW validation.

  1. … 2 more files in changeset.
DEN-7377: preTx: Implementation [Dynamic Instruction Loading] ( 2 )

DEN-7457: Meetings & Support - General purpose

  1. … 51 more files in changeset.
DEN-7408: Code improvement for navigation

- VTreatmentCreate is cleaned up. (still needs a lot of work)

  1. … 11 more files in changeset.
DEN-7408: Code improvement for navigation

- added vCondition to the page function for the stack item to pass the state property value.

- added fault , service and initial placeholder pages.

  1. … 14 more files in changeset.