UI-LEAH-1884_Leahi/Features/Leah 1884 UI BN Administrativescripts
UI-LEAH-1884_Leahi/Features/Leah 1884 UI BN Administrativescripts
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 619 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 619 more files in changeset.
LEAH-1884: [WIP] DeviceController Improvement

- got a potentially solid idea on how to design the models and controllers

- the idea is completely implemented in brightness.

- valid implementations are usbmount, brightness, wifiscan.

- other than brighness other ones need a final review.

- Note: if the requested value is not valid in toByteArray application would be better to rever like the one is done in fromByteArray. to test set the range of the brightness slider to 0 to 6.

- extra setting models need to be removed like the device controller.

  1. … 24 more files in changeset.
LEAH-1884: DeviceController improvement

- Fixed an issue in the VTreatmentRanges if the config file has not read successfully and the dialysate flow rate is set to zero.

- Added the WiFiList device model to the Device management.

- Added MVC for the WiFi List scan.

- Added the readyReadStandard{out/err} to get the executed in-progress script output, to help communicate with the script during execution. [ Update progress to be sent to the application during execution.]

- Added the mCompleted to the DeviceModels to detect when the execution of the progress is complete and it is not the intermittent communication during the execution.

  1. … 14 more files in changeset.
UI-DEN-SUNRISE
UI-DEN-SUNRISE
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 31 more files in changeset.
UI-DEN-16899_Release/Stagingmerge Den 16899 UI BN R13 Build
UI-DEN-16899_Release/Stagingmerge Den 16899 UI BN R13 Build
DEN-16851: New Alarm for Increased Occlusion Sensitivity - Added the following fields to mssage 0x0009: arterial long filter, venous long filter, blood pump occl long filter, partial occl base line

DEN-16204: UI: FW: Pure Water mode turning on intermittently.

- Added a new payload to send the current firmware value with the response to a set from UI

- Changed the RO only mode code to not flip the UI RO only status if the firmware rejected the set value and just get the status in the firmware

DEN-16862: Add a command to reset HD during UI software install

- Added a command to request an HD reset during the UI software and while HD is in service mode.

DEN-16864: UI Bloodflow message is incomplete

- Updated the payload of the following messages: 0x002C, 0x0005, 0x0008, 0x006C.

DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Changed the order of the UI response message from firmware so the accept is always the first field.

DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Addressed the code review comments.

  1. … 30 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 31 more files in changeset.
RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

Done

Done

Done

Done

Done

Done

Done

Done

Done

Done

add the newly added parameter here as well.

add the newly added parameter here as well.