DENBUG-318: Rejection Ratio Setting Feature Implementation in Manufacturing Configuration For Treatment - Updated the code to add max RO rejection ratio and min inlet water conductivity alarm limit to the HD institutional record and display it on the UI
DENBUG-318: Rejection Ratio Setting Feature Implementation in Manufacturing Configuration For Treatment - Updated the code to add max RO rejection ratio and min inlet water conductivity alarm limit to the HD institutional record and display it on the UI
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same - Made UF Rate related values in VTreatmentAdjustmentUltrafiltrationEdit to be rounded up prior to displaying on the UI.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same - Updated the current UF rate in the VTreatmentUltrafiltration to be rounded up. - Removed commented code.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same - Updated the code to round up the UF rate that is displayed on the UF rate change screen.
DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating - Added code to clear the saline bolus reject reason upon changing the treatment states.