As we discussed, I think alarm 150 and related inlet water alarms are already set to block triggering during rinseback and post-treatment modes. So these changes are likely not needed.

As we discussed, I think alarm 150 and related inlet water alarms are already set to block triggering during rinseback and post-treatment modes. So these changes are likely not needed.

DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
DG-DEN-16309_Heat Disinfect Has Spikes IN Temperature Near 95C Which Will Cause Water TO Boil AT
This was a specific change for this alarm (150), and the associated alarms in this routine. It appeared there were similar mode limits for temperature as well. Pressure limits were more uniquely re...

This was a specific change for this alarm (150), and the associated alarms in this routine. It appeared there were similar mode limits for temperature as well. Pressure limits were more uniquely related to valve position, I thought.

I had seen the related changes in 59, and included them in my develop build.

We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature? FYI - getHDOperationMode()...

We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature?

FYI - getHDOperationMode() was not working - HD op mode handler was mishandling message - fixed in my DENBUG-59 branch, but not yet in staging.

DG-DEN-16308_Srsdg 53 Not Met
DG-DEN-16308_Srsdg 53 Not Met
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
DG-DIAL-276_DG Event Event ID 15 Prints Duplicate Events With Different State Time
Checking for alarm in this function means we will check when in control to flow or control to pressure states. Also checked only once per control interval which is maybe ok, but not typical for ala...

Checking for alarm in this function means we will check when in control to flow or control to pressure states. Also checked only once per control interval which is maybe ok, but not typical for alarm checks.
We do control to flow for treatment modes. Control to pressure is used in some cleaning modes. Not sure we want this alarm enforced in cleaning modes.

Remove blank line.

Remove blank line.

Put declarations at top of function. Use our BOOL type. Put ternary in ().

Put declarations at top of function. Use our BOOL type. Put ternary in ().

No point to having non-zero clear persistence - it's a fault.

No point to having non-zero clear persistence - it's a fault.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add a blank line between declarations and code.

Add a blank line between declarations and code.

#includes are sorted. Put OperationModes.h in its proper place.

#includes are sorted. Put OperationModes.h in its proper place.

required changes are done.

required changes are done.

DG-DEN-16280_Stuck IN Post Treatment When Inlet Water Alarm Appears
DG-DEN-16280_Stuck IN Post Treatment When Inlet Water Alarm Appears
Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting...

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting). Function would also create and send the cmd response message and return TRUE/FALSE which would set local result here for the ack response message below.
We should reject cmd if pumps are busy/on and/or if current DG mode is fill or fault mode.