DEN-15660 addressed the code review comments

  1. … 5 more files in changeset.
Nelson is not part of the commercial code.

Nelson is not part of the commercial code.

Done

Done

Done

Done

Done

Done

Nelson support is only in debug mode.

Nelson support is only in debug mode.

Done

Done

Done

Done

The un-averaged TDi and TRo are used for dialysate temperature checking and trimmer heater control I think we should keep them to be able to have a full picture if needed unless the CAN bus is real...

The un-averaged TDi and TRo are used for dialysate temperature checking and trimmer heater control I think we should keep them to be able to have a full picture if needed unless the CAN bus is really flooded.

This is a time and not timeout.

This is a time and not timeout.

Done

Done

Done

Done

Done

Done

Done

Done

Correct indentation

Correct indentation

Correct indentation

Correct indentation

Indentation needs to be corrected.

Indentation needs to be corrected.

handleEmptyBottleFlag is only an input

handleEmptyBottleFlag is only an input

DG-DEN-15702_Alarm 150 Inlet Water Conductivity High Does Not Disappear ON Standby Screen When
DG-DEN-15702_Alarm 150 Inlet Water Conductivity High Does Not Disappear ON Standby Screen When
DEN-15702 Check VPI Open for Inlet Conductivity, Pressure, temperature. If VPI Closed, clear alarms.

  1. … 1 more file in changeset.
What about acid?

What about acid?

This looks strange. If we have a empty bottle alarm, shouldn't we just not start a fill in the first place? With this approach, you're just stuck in fill start state until user clears the alarm.

This looks strange. If we have a empty bottle alarm, shouldn't we just not start a fill in the first place?
With this approach, you're just stuck in fill start state until user clears the alarm.

DEN-15660, DEN-15781, added code to not check the dialysate temperature sensorswhen the dialysate flow is zero. DEN-15808, Added code to process the empty jug alarm

  1. … 2 more files in changeset.
Can we move these declarations up to top of scope?

Can we move these declarations up to top of scope?

Replace "TIMEOUT" with "TIME"

Replace "TIMEOUT" with "TIME"

Is this really a timeout? Or is it a minimum duration?

Is this really a timeout? Or is it a minimum duration?

Do we still want un-averaged TDi and TRo in this broadcast? Is anybody using them? If not, why not replace the old temps with the new averages?

Do we still want un-averaged TDi and TRo in this broadcast? Is anybody using them? If not, why not replace the old temps with the new averages?

Swap first and TRo positions in enum. TRo should be first, then we add a first to mark it as first.

Swap first and TRo positions in enum. TRo should be first, then we add a first to mark it as first.