DEN-13946 Make monitor functions for pressure, temperature, thermistors and load cells consistent. Address code review comments.

DEN-14559 added functions to reset the POST states of the drivers that have a POST state machine for the command to request POST again

  1. … 7 more files in changeset.
DEN-13946 Do not process thermistor alarms if in DG_INIT

DEN-13946 Code review correction

DEN-14559 updated the cleaning modes with the SRS changes

  1. … 7 more files in changeset.
DEN-13946 Darren comments DEN-14075 Corrected thermistor override in previous commit

DEN-13946 Correction to thermistor timeout.

Merge branch 'DEN-13946-certain-temperature-sensors-trigger-wrong-alarm' into develop

# Conflicts:

# firmware/App/DGCommon.h

  1. … 4 more files in changeset.
DEN-13946 Complete and tested for termperatures, pressures, load cells and thermistors. DEN-14075 Corrected thermistor override limits.

Merge branch 'staging' into DEN-13946-certain-temperature-sensors-trigger-wrong-alarm

DEN-13946 Update thermistor init.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
DEN-14497 addressed the code review comment

DEN-13946 - Alarm generation is complete. Thermistors now configured based upon beta or DVT hardware configurations. Working with Noe to address FPGA thermistor values after reset.

  1. … 2 more files in changeset.
DEN-14497 added the latest VectorCAST from PiSquare

Merge branch 'staging' into DEN-14497-dg_hd_dev-hd_dg_dvt-update-part-7

  1. … 13 more files in changeset.
DEN-14497 addressed the code review comments

  1. … 3 more files in changeset.
Fixed merge conflicts.

  1. … 13 more files in changeset.
Merge branch 'staging' into DEN-13946-certain-temperature-sensors-trigger-wrong-alarm

DEN-13946 - Use typedefs for iterating over sensors.

DEN-14497 updated the drain pump flow control algorithm. Fixed the inlet water pressure, temperature, conductivity alarm recovery in treatment modes. Updated chemical disinfect flush mode

  1. … 6 more files in changeset.
DEN-13946 Use typedefs for itterating over sensors.

DEN-14497: Changed CPo and RR alarms to only occur at end of fill based on average conductivity.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14086-alarm-159-dialysate-inlet-pump-failed-flow-vs-motor-speed-check' into develop

DEN-14086: fixed logging issue with LC drift alarm.

Fixed merge conflicts.

  1. … 7 more files in changeset.
DEN-13946 - Correct monitor functions where multiple sensors are "or'd" into a single alarm.

DEN-14086: Increase loadcell drift allowance from +/- 60 to 80 grams - and for heat disinfect, +/- 200 grams.

DEN-14497 updated the drain pump control for the flow and updating chemical disinfect is in progress

  1. … 2 more files in changeset.