Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-17047 removed the UV reactors code from the firmware

  1. … 6 more files in changeset.
DEN-16970 added a check to not set any of the cleaning modes back to cancel water path if any of the cap alarms is active

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-16680 updated the VectorCAST test cases

Merge branch 'DEN-16680-p1b-inlet-water-shutdown-during-heat-disinfect-staging' into develop

  1. … 1 more file in changeset.
DEN-16680 updated the start/stop heat disinfect code

  1. … 1 more file in changeset.
DEN-16680 updated the heat disinfect code

DEN-16680 updated the code

DEN-16680 updated the heat disinfect mode with the latest changes on the passive active cool heat disinfect

  1. … 3 more files in changeset.
DEN-16680 fixed a typo

DEN-16680 addressed the code review comments

  1. … 1 more file in changeset.
DEN-16680 updated the code

DEN-16680 added the doxygen comments for the newly developed functions

DEN-16680 added code to support the nocturnal heat disinfect

  1. … 3 more files in changeset.
Changed the number of the chemical disinfect cycles to 3

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-236 added the code changes from staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-236 changed the maximum produce dialysate time in the mode fill to 45 seconds

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DENBUG-236 added the changes from staging

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DENBUG-232 added code to get the reservoirs empty weight from the calibration records and set the reservoirs tare offset to 0 upon transitioning to service mode

  1. … 2 more files in changeset.
DENBUG-236 changed the mode fill produce diaylsate timeouts to run the concentrate pumps at faster speed

DENBUG-236 added a blank line

DENBUG-236 changed the polarity of the concentrate pumps park progress

DENBUG-236 added code to check the progress of the concentrate pump's park

  1. … 2 more files in changeset.
DENBUG-236 added code to not transition from the fill modes's start state unless both concentrate pumps are parked

DENBUG-236 moved the local variables to the top in Gen Idle

DENBUG-236 changed the code set the reset park alarms to False once the pump started and not to park the concentrate pumps if they are already parked

  1. … 1 more file in changeset.