DENBUG-318 addressed the code review comments

DENBUG-318 added code to request the DG institutional values in mode standby until they are received

  1. … 1 more file in changeset.
Merge branch 'DENBUG-318-rejection-ratio-setting-feature-implementation-in-manufacturing-configuration-for-treatment-staging' into develop

  1. … 1 more file in changeset.
DENBUG-318 added code to keep requesting DG institutional values once start DG has been sent until the values are received

  1. … 1 more file in changeset.
DENBUG-318 added the request command to get the DG institutional record from HD upon starting the DG

DENBUG-318 added an RO rejection ratio event

DENBUG-318 changed the max RO rejection ratio to min RO rejection ratio in percent and updated the mode fill calculations to check the RR

  1. … 1 more file in changeset.
DENBUG-320 removed the acid and bicarb used volumes in mode fill broadcast

Merged DENBUG-318

  1. … 4 more files in changeset.
DENBUG-318 added a new event for min inlet water conductivity alarm from HD institutional record

  1. … 1 more file in changeset.
DENBUG-318 updated the code to receive all the DG institutional record items. Added code to use the min inlet water conductivity alarm from the institutional record

  1. … 5 more files in changeset.
DENBUG-319 added code to receive a command from dialing to enable the mode fill chemicals test

  1. … 3 more files in changeset.
DENBUG-319 added code to skip the acid and bicarb check in mode fill if it is not the first fill

DENBUG-318 changed the location of requesting RO rejection ratio upon starting DG

DENBUG-320 addressed a code review comment

Resolved merge conflicts from DENBUG-318

  1. … 5 more files in changeset.
DENBUG-318 added the word max to the RO rejection ratio

  1. … 2 more files in changeset.
DENBUG-320 added the word max to the RO rejection ratio

  1. … 3 more files in changeset.
DENBUG-320 removed the code that checked for the acid and bicarb low volume and triggered alarms

  1. … 5 more files in changeset.
DENBUG-318 added an event for RO rejection ratio from HD institutional record

DENBUG-318 added code to request and receive the RO rejection ratio from the HD institutional record and apply it in the mode fill

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merged release/p1b/staging

  1. … 2 more files in changeset.
DENBUG-298 updated the code from staging

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DENBUG-307 updated the active cool mode to set the actuators prior to transitioning to complete state

DENBUG-307 updated the active cool to run the concentrate pumps for a few seconds at the end of the mode

DENBUG-307 changed the heat disinfect active cool to run the concentrate pumps during the final reservoir drains

DENBUG-304 uncommeted the code that synchronizes the DG RTC time with HD upon starting a treatment

  1. … 1 more file in changeset.
DENBUG-307 updated the heat disinfect to run the concentration pumps for the duration of heat disinfect from the time that the reservoirs are filled up until the end of disinfect R2 to R1