Merge branch 'DIAL-281-requesting-heat-disinfect-timer-overrides' into develop

  1. … 2 more files in changeset.
DIAL-278: implementation of heat disinfection state timer overrides.

  1. … 4 more files in changeset.
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
DG-DIAL-281_Requesting Heat Disinfect Timer Overrides
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done.

Done.

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is redu...

Lets add a comment here that "we are insisting conductivity reach 100% of expected level since signal is very noisy and peak noise will reach 100% when average is about 93%. if sensor noise is reduced, this level should also be reduced accordingly."

100% now.

100% now.

take latest changes from staging

  1. … 3 more files in changeset.
Merge remote-tracking branch 'origin/DEN-16280-stuck-in-post-treatment-when-inlet-water-alarm-appears' into develop

  1. … 5 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add a blank line between declarations and code.

Add a blank line between declarations and code.

#includes are sorted. Put OperationModes.h in its proper place.

#includes are sorted. Put OperationModes.h in its proper place.

DEN-16093 merge in to develop

  1. … 3 more files in changeset.
required changes are done.

required changes are done.

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting...

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting). Function would also create and send the cmd response message and return TRUE/FALSE which would set local result here for the ack response message below.
We should reject cmd if pumps are busy/on and/or if current DG mode is fill or fault mode.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
Merge branch 'DIAL-280-reservoir-broadcast-override' into staging

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
Merge branch 'staging' into DIAL-280-reservoir-broadcast-override

  1. … 2 more files in changeset.