Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Changed the minimum RO flow from 0.3 to 0.2 L/min

Merge branch 'DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2' into develop

DEN-13460 updated the definition of the RO pump

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Changed the heat disinfect control

DEN-12974 Updates per code review.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2' into develop

DEN-13460 updated the code

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-13460

    • -11
    • +11
    ./Controllers/ConcentratePumps.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Updated the heaters no flow value

DEN-13460 merged the remote branch

DEN-13460 updated the concentrate pump speed check and the increased the heaters on with no flow to 10 seconds

    • -11
    • +11
    ./Controllers/ConcentratePumps.c
  1. … 1 more file in changeset.
DEN-13460: Increased persistence for heater on with no flow alarm from 1 to 5 seconds.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge remote-tracking branch 'origin/develop' into develop

Merge branch 'DEN-12974-miscellaneous-flush-issues_bb' into develop

DEN-12974 DEN-13558 Update to fill status functions to support V&V

DEN-13460 Heater feedback primary channel changed to read from FPGA, secondary from internal ADC replaces old primary name.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-13460

    • -88
    • +140
    ./Controllers/TemperatureSensors.c
DEN-13460 updated the fault mode to be able to read the NV records in case it was not read correctly

DEN-13460 updated the drivers with the DVT features

    • -11
    • +11
    ./Controllers/TemperatureSensors.c
  1. … 1 more file in changeset.
DEN-13460: Changed RO pump control coefficients to be more agressive.