firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-5963 worked on RO pump and the drain states of heat disinfect

DEN-7091: Changed startup active reservoir to 1

DEN-5963 changed the RO pump control scheme from pressure control to flow control

DEN-7091: Added filter flush progress data broadcast

    • -0
    • +30
    ./App/Services/SystemCommMessages.c
DEN-5980 added a #define for disabling the CRC check comments

DEN-7091: Added water sample command handler

DEN-5980 added or removed some of the states

DEN-5980 removed old calibration commands

    • -27
    • +2
    ./App/Services/SystemCommMessages.c
DEN-6402: Fixed load cell tare reset between tare requests

DEN-5980 added more commands to send and receive NV data

    • -21
    • +315
    ./App/Services/SystemCommMessages.c
    • -2
    • +29
    ./App/Services/SystemCommMessages.h
DEN-6402: Addressed code review comments

DEN-6402: Added alarm for RO pump ramp up retry fail

DEN-6402: Added flow meter check and filters for load cells

DEN-6402: Added change valve setting command

    • -0
    • +26
    ./App/Services/SystemCommMessages.c
DEN-6402: Use reset persistent alarm timer

DEN-6402: Added control setup state and delay for RO pump

DEN-5980 disabled some functions for testing

    • -8
    • +102
    ./App/Services/SystemCommMessages.c
DEN-5963 added a few lines to stop the drain pump if RPM is 0

DEN-5980 started adding the structs to NVDatMgmt and modified VectorCAST's configuation file to use GNU compiler template

    • -0
    • +13
    ./App/Services/SystemCommMessages.c
DEN-6200: Revert heater module changes

DEN-6200: Added target temperature to trimmer heater command

    • -14
    • +5
    ./App/Services/SystemCommMessages.c
DEN-6200: Fixed doxygen comment

DEN-6200: Added command id to DG command response

DEN-6200: Return command request reject reason

    • -2
    • +31
    ./App/Services/SystemCommMessages.c
DEN-6200: Combined getAlarmActive and isAlarmActive

DEN-6200: Fixed spelling and doxygen comment

    • -2
    • +2
    ./App/Controllers/TemperatureSensors.c
DEN-6200: Added target flow rate to RO pump data record

DEN-6200: Added alarm condition cleared function

    • -7
    • +36
    ./App/Services/SystemCommMessages.c
DEN-5963 worked on the RO pump

DEN-6200: Updated persistent alarms

    • -19
    • +12
    ./App/Controllers/ConductivitySensors.c
    • -23
    • +6
    ./App/Controllers/TemperatureSensors.c