Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged staging

DEN-15183: Fixed DEN-15198 by not starting air trap control until treatment starts (patient connected).

    • -1
    • +0
    /firmware/App/Modes/PreTreatmentRecirc.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14325 addressed the code review comments

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

It is not needed. So I removed it.

It is not needed. So I removed it.

Needed?

Needed?

DEN-15183: Addressed code review comments. Fixed DEN-15194.

Fixed.

Fixed.

Fixed.

Fixed.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

    • -14
    • +0
    /firmware/App/Controllers/PresOccl.c
DEN-15183: Implemented updates to treatment recirculation workflow to include states for instructing user to disconnect and reconnected patient. Updated scalar for alarm audio current high gain signal. Updated alarm audio POST.

    • -4
    • +103
    /firmware/App/Modes/TreatmentRecirc.c
    • -29
    • +48
    /firmware/App/Services/AlarmMgmt.c
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adjusted Alarm limit

DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C

Add treatmentTimeBroadcastTimerCtr to header.

Add treatmentTimeBroadcastTimerCtr to header.

DEN-15024 Change syringe pump off error count to max time windowed count.

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
Add doorOpened to header.

Add doorOpened to header.

DEN-15183: indicate treatment over in treatment times broadcast if alarm preventing alarm resumption.

HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15196_Alarm 98 Rework
HD-DEN-15196_Alarm 98 Rework
DEN-15183: Removed pressure limit alarm exemption for pump ramping. Address sequence issue with invert dialyzer logic DEN-15195.

    • -14
    • +0
    /firmware/App/Controllers/PresOccl.c
DEN-15104 Minor correction

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adding an alarm blocking state to prevent cascade of errors during power fail.

    • -53
    • +75
    /firmware/App/Services/AlarmMgmt.c
HD-DEN-14980_Unable TO Resume After HD Unit Plugged Back IN Pre Treatment After Priming
HD-DEN-14980_Unable TO Resume After HD Unit Plugged Back IN Pre Treatment After Priming