Approach simplified.

Approach simplified.

Update header.

Update header.

Update header

Update header

Update header

Update header

I reviewed DEN-15741 which is re: desire to stop DPi if user stopped with button. Everything I see changed in this module seems like a very complicated way to achieve that using properties that I d...

I reviewed DEN-15741 which is re: desire to stop DPi if user stopped with button.
Everything I see changed in this module seems like a very complicated way to achieve that using properties that I don't think were intended for this.
I think all we needed to do was check if user stop alarm is active and, in that special case, stop DPi instead of recirculating dialysate like it usually does in the recirc stop state.

Why was all of this necessary?

Why was all of this necessary?

Comment is copy. What is this flag for? Is it user requested?

Comment is copy. What is this flag for? Is it user requested?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Done. dryPressureTestsCompleted is only set when test config is active so I don't want to include that in doxygen documentation.

Done. dryPressureTestsCompleted is only set when test config is active so I don't want to include that in doxygen documentation.

Add pressureSelfTestPreNormalStartTime, dryPressureTestsCompleted

Add pressureSelfTestPreNormalStartTime, dryPressureTestsCompleted

Add occlusionBaseLineWasSet

Add occlusionBaseLineWasSet

HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2
HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2
HD-DEN-15858_NO Timer Countdown IS Visible For The Confirmation OF Chemical Flush Sample
HD-DEN-15858_NO Timer Countdown IS Visible For The Confirmation OF Chemical Flush Sample
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-16016_Bug Unable TO Start The Saline Bolus 3
HD-DEN-16016_Bug Unable TO Start The Saline Bolus 3
HD-DEN-15512_DG HD Failing TO Set Calibration Record
HD-DEN-15512_DG HD Failing TO Set Calibration Record