DEN-15278: Added rotor hall sensor state to BP, DPi and DPo broadcast messages.

  1. … 5 more files in changeset.
HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Oops. DeMorganized and done.

Oops. DeMorganized and done.

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Remove space between ().

Remove space between ().

Add blank line between declaration and code. Prefer we not assume bool and BOOL are the same. Use if statement and explicit assignment of TRUE to result if all conditions are true. First conditio...

Add blank line between declaration and code.

Prefer we not assume bool and BOOL are the same. Use if statement and explicit assignment of TRUE to result if all conditions are true.

First condition seems backward.

Second and third conditions seem problematic. You will never trigger an AC power loss alarm. Should remove them I think.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Restore blank line. Should be 2 blank lines above and below test support banner.

Restore blank line. Should be 2 blank lines above and below test support banner.

Remove blank line.

Remove blank line.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done. Error propagration.

Done. Error propagration.

Use power loss to reset the out of position counter for the valves. CPLD and isACPowerLost used to give the valves time to settle before checking.

Use power loss to reset the out of position counter for the valves. CPLD and isACPowerLost used to give the valves time to settle before checking.

Done.

Done.

Done.

Done.

Done.

Done.

Needs an else to zero out of range counter.

Needs an else to zero out of range counter.

This check is too far out. Each check function inside should check this with DG comm.

This check is too far out. Each check function inside should check this with DG comm.

Remove this condition.

Remove this condition.

Remove this condition.

Remove this condition.