Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
Merge branch 'DEN-15804-blood-flow-rate-accuracy-results-fail-or-are-close-to-failing' into develop

DEN-15804 changed the blood flow coefficients

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14934: Updated blood flow estimation coefficients.

Done.

Done.

Add resumeRinsebackRequested.

Add resumeRinsebackRequested.

DEN-14934: Aligned rotor vs motor speed alarm threshold with requirements.

  1. … 2 more files in changeset.
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

HD-DEN-15844_Vbt Remains Open During Pressure Dry Self Test Triggering Alarm 23
HD-DEN-15844_Vbt Remains Open During Pressure Dry Self Test Triggering Alarm 23
DEN-15844: Fixed HD pump direction check to handle reverse w/r/t minimum speed.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-15489: Addressed issue with preventing pumps to be commanded to top speed.

  1. … 2 more files in changeset.
HD-DEN-15489_FW DG Disinfect Has Expired Check Does Not Clear With Valid Dates
HD-DEN-15489_FW DG Disinfect Has Expired Check Does Not Clear With Valid Dates