Merged staging to DEN-15792.

  1. … 9 more files in changeset.
DEN-15792: Fixed some comments.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
HD-DEN-16408_Dialysate Flow Performance Decrease AT 600Qd
HD-DEN-16408_Dialysate Flow Performance Decrease AT 600Qd
DEN-16408: Updated DPi and DPo modules to increase speed range to 0..4000 RPM in order to allow pumps to reach 600 mL/min rate consistently.

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

Done.

Done.

Align declaration comments as a block on the right.

Align declaration comments as a block on the right.

Updates to dialysate pump drivers per motor controller RPM range change to 4,000 in order to allow sufficient range to always reach Qd=600 mL/min.

  1. … 1 more file in changeset.
HD-DEN-15792_HD Software Design Description
HD-DEN-15792_HD Software Design Description
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Fixed merge conflicts.

  1. … 7 more files in changeset.
DEN-14815: Switched to new CPLD based AC power loss check function.

  1. … 7 more files in changeset.
Add currentPostTreatmentState to header

Add currentPostTreatmentState to header

HD-DEN-14815_Cartridge Removal Screen IS Displayed Prematurely
HD-DEN-14815_Cartridge Removal Screen IS Displayed Prematurely
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Wouldn't we want to stop if an alarm occurred during recirculation, which affected the blood circuit? My inclination would be to manage which alarms were enabled in this state rather than not imple...

Wouldn't we want to stop if an alarm occurred during recirculation, which affected the blood circuit? My inclination would be to manage which alarms were enabled in this state rather than not implement the mechanism to stop the pumps.

Implementing the change.
Done.

Should blood recirc blocked be based on alarm properties (as you've written here) or based on whether user stopped (specifically)? My feeling is that since there is no blood in blood side of circui...

Should blood recirc blocked be based on alarm properties (as you've written here) or based on whether user stopped (specifically)? My feeling is that since there is no blood in blood side of circuit (because rinseback was completed before we got to recirc), the NoBloodRecirc alarm property maybe doesn't apply here and we should only stop BP if the user stop alarm is active.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.