Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-5674: Fixed drain reservoir cmd to DG to include indication of whether reservoir load cells should be tared.

  1. … 3 more files in changeset.
DEN-5674: Added synchronize of alarm audio/lamp control with published alarm state so that UI can flash alarm in sync w/ lamp. Added alarm audio.

  1. … 8 more files in changeset.
DEN-5887: Fixed name format for const data to all caps w/ underlines per coding standard.

  1. … 4 more files in changeset.
DEN-5887: Moved some load cell functions from dialysate outlet pump controller to DG Interface module. Added small and large moving average filters to load cell readings. Use small moving average for normal UF tracking/control and use large moving average for baseline and final reservoir weights to minimize reservoir transition errors due to load cell noise.

  1. … 3 more files in changeset.
DEN-5381: Removed test support macros

  1. … 2 more files in changeset.
DEN-4641: Updated persistent alarms to use ms instead of task interval counts.

  1. … 3 more files in changeset.
DEN-4641: Addressed code review feedback.

  1. … 3 more files in changeset.
DEN-4641: Merged with master. Fixed merge conflicts. Addressed code review feedback.

  1. … 17 more files in changeset.
DEN-4641: Addressed code review feedback.

  1. … 4 more files in changeset.
DEN-4641: Updates to UT.

  1. … 2 more files in changeset.
DEN-4641: Distributed remaining payload record definitions to associated modules and deleted MessagePayloads.h. Added messaging and verification of in-line pressure limit changes.

  1. … 10 more files in changeset.
DEN-4641: Fixed bug in last change.

DEN-4641: inline pressure range check only in treatment mode.

DEN-4641: Fixed safety line pull when pump running while off for dialysate outlet pump.

DEN-4640: Added inline pressure alarm checks.

  1. … 2 more files in changeset.
Fixed merge conflicts from master. Fixed some dev testing issues from merge.

  1. … 20 more files in changeset.
DEN-4641: Fixed PBA conversion formula.

DEN-4641: Pull safety line when cannot stop pumps.

DEN-4641: Changed HD pump Dialin set point override commands to simply set commands.

  1. … 1 more file in changeset.
DEN-4641: More fixes to HD pump motor speed measurement. Dialin cmd to clear all alarms will also restore op mode to prior if in fault mode.

  1. … 3 more files in changeset.
DEN-4641: Minor fixes to pump speed measurement and Dialin alarm features per verification.

  1. … 1 more file in changeset.
DEN-4641: Calculate HD pump motor speed more frequently. Added support for Dialin command to clear all alarms.

  1. … 7 more files in changeset.
DEN-4640: cosmetic changes.

DEN-4640: clear pump direction error counter if not in control state.

  1. … 1 more file in changeset.
DEN-4640: Added persistence to pump direction and blood pump max rotor speed alarms.

DEN-4640: Updated IT.

DEN-4640: Updated UT. Fixes per UT.

  1. … 1 more file in changeset.
DEN-4640: code review.

DEN-4640: Addressed code review comments.

  1. … 3 more files in changeset.
DEN-4641: Doxygen comment fixes.