ModeTreatment.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

This part of code has been removed.

This part of code has been removed.

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

HD-DEN-15788_Pbo Ramps UP TO 400 Mmhg During Dialysate Circuit Priming
HD-DEN-15788_Pbo Ramps UP TO 400 Mmhg During Dialysate Circuit Priming
Merge branch 'staging' into DEN-15660-test_configuration_treatment_recovery

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Done.

Done.

Done.

Done.

Add cartrideUsedTestRun

Add cartrideUsedTestRun

Update header

Update header

Add confirmInstallRequested to header

Add confirmInstallRequested to header

Fixed merge conflict.

  1. … 4 more files in changeset.
DEN-15556: Stabilize pressures after UF rate change.

  1. … 5 more files in changeset.
DEN-15612: Removed unnecessary change.

Fixed. Thanks!

Fixed. Thanks!

DEN-15621: code review comments.

  1. … 1 more file in changeset.
The 2 verify states are single pass. I don't think it makes sense to return to them. They either pass or you go to fault mode and so you wouldn't even be here - so I think you can assume you passed...

The 2 verify states are single pass. I don't think it makes sense to return to them. They either pass or you go to fault mode and so you wouldn't even be here - so I think you can assume you passed.
For first displacement verify state, I think you should go to second displacement setup state.
For second displacement verify state, I think you should go to complete state.

HD-DEN-15716_Skipped Treatment Start State While Transitioning From Pre Treatment TO Treatment
HD-DEN-15716_Skipped Treatment Start State While Transitioning From Pre Treatment TO Treatment