Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
RESOLVED IN CODE WALKTHGROUGH

RESOLVED IN CODE WALKTHGROUGH

We did not check for the upper limit in the past, does this add new rejection scenarios

We did not check for the upper limit in the past, does this add new rejection scenarios

DEN-16693 removed the commented out code

DEN-16693 changed the upper and lower range get functions in the treatment parameters

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add minimum check here too.

Add minimum check here too.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

DEN-16693 addressed the code review comments

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Do we need to add check for minimum treatment duration?

Do we need to add check for minimum treatment duration?

Cast constant as (F32) as before.

Cast constant as (F32) as before.

DEN-16993 addressed the code review comments

  1. … 1 more file in changeset.
Done

Done

dialVolume is a volume (not a flow). Use constant (as before) for dialVolume check.

dialVolume is a volume (not a flow). Use constant (as before) for dialVolume check.

DEN-16693 updated treatment parameters and treament code

  1. … 1 more file in changeset.
HD-DEN-16713_Release11 Vectorcast Unit And
HD-DEN-16713_Release11 Vectorcast Unit And
DEN-16693 updated the institutional records units

DEN-16693 updated the check of the heaprin stop time to account for the value of the heaprin dispense rate

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16693 updated the code with the test results from the lab

Done

Done

DEN-16693 added code to check the in treatment parameters against the institutional data

  1. … 1 more file in changeset.
DEN-16693 updated the treatment parameters to accept 0 for heparin items even if it is below the minimum. Added code to check the sent value to be within min and maximum both