Merge branch 'DEN-12441-sw_dev_sprint_67_darren' into develop

  1. … 3 more files in changeset.
DEN-12224 updated the blood leak driver

  1. … 4 more files in changeset.
DEN-12704: DEN-12344: Fixed alarm 142 to not be checked outside of treatment and pre-treatment mode. specific conditions apply to when the alarm is monitored for each op mode. pre-treatment wont check after pressure settle time. treatment wont check when in tx_stop

  1. … 1 more file in changeset.
DEN-12224: update pre-treatment reservoir mgmt handling of fill command.

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 3 more files in changeset.
DEN-12224, DEN-12642 updated the software configuration in post treatment.

  1. … 3 more files in changeset.
DEN-12635: Fixed init prime() to only initialize values. Removed duplicate init call for consumable self tests

DEN-12224: Split some pre-treatment self-test alarms to better differentiate.

  1. … 1 more file in changeset.
DEN-12224 fixed the switches enable high/low

  1. … 3 more files in changeset.
DEN-12224, DEN-12536 updated the post treatment mode

DEN-12224, DEN-12536 updated the post treatment's reservoir states

DEN-12224, DEN-12536 updated post treatment mode to drain both reservoirs

Merged DEN-12224

DEN-12224 updated the consumables self test

Disabled code for testing

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 1 more file in changeset.
DEN-12224: fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-12224: Explicitly set actuators on pre-treatment transition to functions. Testing longer reservoir settle time.

  1. … 1 more file in changeset.
DEN-12224 updated the code

  1. … 1 more file in changeset.
DEN-12224 updated the code

  1. … 3 more files in changeset.
DEN-12224 Fixed DEN-12346 HD prime out of time alarm [ 227] Systems agreed to increase MAX_PRIME_TIME from 10 min to 15 minutes

DEN-12224 Fixed ticket DEN-12533 HD should accepting only value for arterial pressure high from -270 to 0

DEN-12224 Added build switch to disable syringe pump removed alarm[219] on system without syringe pump installed

DEN-12441 Reset blood sitting timer when turning on blood circulation.

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 1 more file in changeset.
DEN-12224: Fixed merge error in ConsumableSelfTests.c

DEN-12224: Added explicit actuator sets to pre-treatment sub-mode transition functions. Renamed no cartridge self-test for pressure sensors. Removed no cartridge self-tests for blood and dialysate flow sensors and board temperature.

DEN-12441 Handle blood sitting timer when blood is not circulating. Update broadcast info.