DEN-11098 Removed code related to run blood circuit again because it no longer needed

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 98 more files in changeset.
Merge branch 'DEN-10602_UI_DEV_Add_PatientID_PreTreatment' into develop

DEN-11098 Addressed code review comments

  1. … 2 more files in changeset.
DEN-11114, DEN-11364 worked on the blood leak calibration and removed the fans clear the alarm if the RPM was in range again

  1. … 5 more files in changeset.
Fixed error in last.

Disabled entire BLD module. No t/o on no cartridge self-tests.

  1. … 1 more file in changeset.
Disabled venous pressure check. Increased prime max time to 15 min.

  1. … 1 more file in changeset.
Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

  1. … 1 more file in changeset.
Relaxed pressure recovery dry test requirement.

Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' into develop

  1. … 2 more files in changeset.
DEN-1111: Fix dry pressure test. Fix wet flow vs. res test. Adjust test code patch for PBA in BP flow estimator.

  1. … 2 more files in changeset.
Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

  1. … 1 more file in changeset.
DEN-11098 Updated prime time limit for the saline dialyzer state

DEN-11114: Fixed AC POST test.

Merged DEN-11114 and resolved merge conflict

DEN-11114 added header to the functions in FPGA

  1. … 1 more file in changeset.
DEN-11098 Converted from seconds to milliseconds fix

Fixed merge conflict

  1. … 1 more file in changeset.
DEN-11114: Fixed mid-treatment changes not immediately sending new ranges to UI problem.

DEN-11098 Corrected state transition and changed blood pump speed from 400 to 300 per system specification

DEN-11114: ifdef'd dead code in alarm lamp driver. Clear POR bit after logging reset in POST.

  1. … 1 more file in changeset.
Merge branch 'DEN-10602_UI_DEV_Add_PatientID_PreTreatment' into develop

DEN-11543: Unable proceed to the Confirm Treatment when "08-3251-9" Acid Concentrate is selected

The HD FW updated to accept up to 4 acid concentration type index.

Below is the UI DataList.conf for acid concentrate.

[Acid Concentrate]

Fres. Naturalyte

08-1251-1

08-2251-0

08-3251-9

DEN-11098 Incorporated peer review comments

  1. … 1 more file in changeset.
DEN-11098 Updated code per peer review comments

DEN-11098 Updated time out limits

Merge branch 'DEN-10602_UI_DEV_Add_PatientID_PreTreatment' into develop

  1. … 2 more files in changeset.
DEN-11098 Implemented code update due to requirements change