Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Finishing up the story for firmware

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

DEN-11098 Corrected delta volume calculationbased on the past 10 seconds value instead of the past 9 seconds

Merge branch 'DEN-10153_Treatment_Log_Prescribed_UF' into staging

  1. … 1 more file in changeset.
DEN-11098 Implemented steady state volume check

DEN-11098 Updated steady state volume detection logic

DEN-11114 implemented the dialysate heating feature

  1. … 8 more files in changeset.
DEN-11114: Fixed DEN-9288.

DEN-11098 Updated header

DEN-11114: Added reset status bits to startup event to aid investigation into spurious HD resets.

  1. … 1 more file in changeset.
DEN-11098 Corrected R1 and R2 in bypass state and dialyzer state

  1. … 1 more file in changeset.
DEN-11114 addressed the code review comments

  1. … 4 more files in changeset.
DEN-11114 updated the reservoirs code

  1. … 3 more files in changeset.
DEN-11114 tested the modified the reservoir management code

  1. … 3 more files in changeset.
DEN-11098: Addressed code review feedback.

DEN-11098 Added compiler switch RUN_BLOOD_CIRCUIT_PRIME_AGAIN

  1. … 1 more file in changeset.
DEN-11089 Added cases in switch statement to handle prime pause for new saline dialyzer state

DEN-11114 continued working on the resevoir management

  1. … 6 more files in changeset.
DEN-11089 Removed signalDialOutPumpHardStop since it has not been started

DEN-11089 the VDO pump is stopped before transition from dialyzer state to wet selftests state

DEN-11089 Correction - dialyzer dialysate bypass priming path uses R1 instead of R2 and dialyzer dialysate priming path uses R2 instead of R1

  1. … 1 more file in changeset.
Merge branch 'DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11114-dg_dev-dg-dialysate-temperature-control-2-of-2

DEN-11114 initial commit of the reservoirs module. Continued with implementing the reservoir management

  1. … 5 more files in changeset.
DEN-11114: Updated UF control to included estimated UF flow as input.

  1. … 3 more files in changeset.
DEN-11089 Incorporated peer review comments

DEN-11089 if priming for the 1st time, transition to the HD_PRIME_SALINE_SETUP_STATE

DEN-11089 In hadlePrimeDialysateBypassState, state should be HD_PRIME_RESERVOIR_ONE_FILL_COMPLETE_STATE

DEN-11089 Swapped the dialysate priming circuit - prime the R2 circuit first then prime the R1 circuit

DEN-11089 Implemented new saline dialyzer priming state to purge air from the dialyzer path

DEN-11114: Reset BP rotor count before each treatment starts. Allow BP rotor count override to work even if WORN_CARTRIDGE build flag is enabled.

  1. … 1 more file in changeset.
Disabled watchdog and safety shutdown POST tests for unit 2.