Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-12224 fixed the switches enable high/low

  1. … 3 more files in changeset.
DEN-12224, DEN-12536 updated the post treatment mode

DEN-12224, DEN-12536 updated the post treatment's reservoir states

DEN-12224, DEN-12536 updated post treatment mode to drain both reservoirs

Merged DEN-12224

DEN-12224 updated the consumables self test

Disabled code for testing

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 1 more file in changeset.
DEN-12224: fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-12224: Explicitly set actuators on pre-treatment transition to functions. Testing longer reservoir settle time.

  1. … 1 more file in changeset.
DEN-12224 updated the code

  1. … 1 more file in changeset.
DEN-12224 updated the code

  1. … 3 more files in changeset.
DEN-12224 Fixed DEN-12346 HD prime out of time alarm [ 227] Systems agreed to increase MAX_PRIME_TIME from 10 min to 15 minutes

DEN-12224 Fixed ticket DEN-12533 HD should accepting only value for arterial pressure high from -270 to 0

DEN-12224 Added build switch to disable syringe pump removed alarm[219] on system without syringe pump installed

DEN-12441 Reset blood sitting timer when turning on blood circulation.

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 1 more file in changeset.
DEN-12224: Fixed merge error in ConsumableSelfTests.c

DEN-12224: Added explicit actuator sets to pre-treatment sub-mode transition functions. Renamed no cartridge self-test for pressure sensors. Removed no cartridge self-tests for blood and dialysate flow sensors and board temperature.

DEN-12441 Handle blood sitting timer when blood is not circulating. Update broadcast info.

DEN-12441 Added Recirc Dialysate Only and Blood Only sub states to Treatment End.

DEN-12224: Command DG to standby when HD faults. Removed redundant code.

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-12224-dg_hd_dev-switches-monitor

DEN-12224: only allow 3 options for acid treatment param.

Merged DE-12224

  1. … 9 more files in changeset.
Merged DEN-11928

  1. … 2 more files in changeset.
DEN-12224 pulled the remote and resolved the merge conflicts

  1. … 2 more files in changeset.
DEN-12224 merged staging and added the switches monitor softwar configuration

  1. … 23 more files in changeset.