SystemCommMessages.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merged DEN-8030

  1. … 9 more files in changeset.
DEN-8030 updated the valves controller

  1. … 7 more files in changeset.
Fixed merge conflicts.

  1. … 2 more files in changeset.
DEN-7605: Added build switches.

  1. … 5 more files in changeset.
Need to remove this comment.

Need to remove this comment.

Need to remove this comment.

Need to remove this comment.

Same as comment from DialinFlow.c

Same as comment from DialinFlow.c

We can't have 3 calls to isPersistentAlarmTriggered() on same alarm. If the first call starts the timer and the second or third call is good, it will reset the timer to 0 again. Suggest creating a ...

We can't have 3 calls to isPersistentAlarmTriggered() on same alarm. If the first call starts the timer and the second or third call is good, it will reset the timer to 0 again.
Suggest creating a BOOL variable to have all three conditions or with each other and call isPersistentAlarmTriggered() only once.

initBloodPrime() is called every time we transition to blood prime sub-mode.

initBloodPrime() is called every time we transition to blood prime sub-mode.

Blood prime last motor count does not seem to get updated except in initBloodPrime(). Should it be updated continuously somewhere?

Blood prime last motor count does not seem to get updated except in initBloodPrime(). Should it be updated continuously somewhere?

Fixed merge conflicts.

  1. … 4 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-7605: Added more published data and overrides to syringe pump module.

  1. … 5 more files in changeset.
Fixed.

Fixed.

Fixed.

Fixed.

Copy and paste error. Fixed.

Copy and paste error. Fixed.

Same as above.

Same as above.

Why do we need to subtract from MAX_ALARM_VOLUME_LEVEL?

Why do we need to subtract from MAX_ALARM_VOLUME_LEVEL?

These functions need doxygen description.

These functions need doxygen description.

I agree, but sometimes feel it helps readability.

I agree, but sometimes feel it helps readability.

Yes, you are right. I will use fwcommon code.

Yes, you are right. I will use fwcommon code.