Merged DEN-13460

  1. … 8 more files in changeset.
DEN-13460 updated the code with DVT changes

  1. … 7 more files in changeset.
Check NEARLY_ZERO #define. Value should be 0.00000001F. I believe previous review commented indicated that all floating point #defines should have "F" at the end.

Check NEARLY_ZERO #define. Value should be 0.00000001F. I believe previous review commented indicated that all floating point #defines should have "F" at the end.

SW_CONFIG_ENABLE_VALUE should be first in the conditional.

SW_CONFIG_ENABLE_VALUE should be first in the conditional.

Need #define for minimum vRef

Need #define for minimum vRef

Is the data being published elsewhere?

Is the data being published elsewhere?

Is this supposed to have been removed.

Is this supposed to have been removed.

DEN-12931: Changed syringe pump calibration to expected DAC voltage. Added verification of calibration.

  1. … 2 more files in changeset.
indent of cases looks too small.

indent of cases looks too small.

Looks like exit normal request (below) will take precedence over this request. Is that intended?

Looks like exit normal request (below) will take precedence over this request. Is that intended?

Update input/outputs throughout.

Update input/outputs throughout.

Add blank line before return.

Add blank line before return.

DEN-12931: Added request for UI POST final result to ensure we get result before its needed in UI POST test.

  1. … 2 more files in changeset.
Do we want to keep this debug & sprintf?

Do we want to keep this debug & sprintf?

HD-DEN-12931_DG HD Dev HD DG Dvt Update
HD-DEN-12931_DG HD Dev HD DG Dvt Update
Merge remote-tracking branch 'origin/DEN-13141-miscellaneous_hd_issues_bb' into develop

  1. … 1 more file in changeset.
Merged DEN-12931

  1. … 5 more files in changeset.
DEN-13141, DEN-13067 Correct POST messaging for UI.

  1. … 2 more files in changeset.
DEN-12931 merged staging and resolved the conflicts

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
Merge branch 'DEN-12931-dg_hd_dev-hd_dg_dvt-update' into develop

  1. … 7 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-12845 addressed the code review comments

  1. … 2 more files in changeset.
Removed the message.

Removed the message.

We got recommendation to monitor arterial pressure even when not dialyzing if patient is still connected (so all other treatment mode states other than dialysis when not delivering saline bolus), b...

We got recommendation to monitor arterial pressure even when not dialyzing if patient is still connected (so all other treatment mode states other than dialysis when not delivering saline bolus), but didn't want to apply the user provided limits - just want to monitor for max pressure situation.

Im assuming that this if statement covers sub modes like Saline and Blood Prime. How do we know during blood prime, specifically, that we would not be dealing with an edge case in treatment paramet...

Im assuming that this if statement covers sub modes like Saline and Blood Prime. How do we know during blood prime, specifically, that we would not be dealing with an edge case in treatment parameters for the patient (Ex: -300 mmHg to -270 mmHg) and triggering a false positive alarm?

Note: I dont know actual patient bp values but these are currently acceptable treatment parameters for arterial pressure limits.

Merge branch 'staging' into DEN-12845-sw_dev_sprint_s71_steve

  1. … 7 more files in changeset.