SystemCommMessages.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
UT and IT updates. Minor fixes to messaging from DG.

  1. … 4 more files in changeset.
UT updates. Added UT for DGInterface. Misc fixes per UT.

  1. … 5 more files in changeset.
UT and IT updates.

  1. … 2 more files in changeset.
Added misc test/debug build flags. Added persistence to CAN buffer overflow faults so that we can try to recover from one-off spurts when a node is coming online.

  1. … 10 more files in changeset.
Made reservoir target volumes public. Set all UF volumes to F32 type for consistency.

  1. … 5 more files in changeset.
Fixed message payload sizing issue on UF option confirmation message handling.

  1. … 2 more files in changeset.
Changed max dialysate volume from 180 to 150 L. Split UF change response messages into 2 separate messages to prevent UI confusion.

  1. … 2 more files in changeset.
Added safety shutdown override command support. Zero treatment duration param in response to UF adjustment message if time adjustment should be disabled.

  1. … 7 more files in changeset.
Moved modes and sub-modes enums to common header. Added sub-mode to HD op mode broadcast. Added current UF rate to UF rate change response to UI.

  1. … 29 more files in changeset.
Misc UF fixes. Changed CAN from FIQ to IRQ to see if helps with intermittent lockup.

  1. … 10 more files in changeset.
Updated dialysis sub-mode to handle two reservoirs for UF.

  1. … 5 more files in changeset.
Implemented some test code of UF demo support in Standby mode. Started a state machine for treatment mode reservoir management.

  1. … 4 more files in changeset.
Added the micro variable to the build versions and updated handleFWVersionRequest function

  1. … 3 more files in changeset.
Added new HD pump alarms. Started implementing new HD pump alarms.

  1. … 7 more files in changeset.
Added DG messaging.

  1. … 4 more files in changeset.
Reordered a message handler.

  1. … 1 more file in changeset.
Added HD op mode broadcast.

  1. … 2 more files in changeset.
Continued implementation of DGInterface module.

  1. … 2 more files in changeset.
Added new DGInterface module. Messaging for HD/DG comm.

  1. … 8 more files in changeset.
Updates to UT/IT. Minor code fixes per UT/IT.

  1. … 4 more files in changeset.
Added treatment settings range update message. Addressed code review feedback.

  1. … 5 more files in changeset.
Minor messaging updates.

  1. … 5 more files in changeset.
Added messaging for blood/dialysate flow rate change and HD to DG dialysate temperature set points.

  1. … 9 more files in changeset.
Perparing for a merge with DEN-1906

  1. … 46 more files in changeset.
Minor fixes from unit testing. Updates to unit tests.

  1. … 3 more files in changeset.
Fixes for UF and treatment time changes. Updated pump factors.

  1. … 9 more files in changeset.
Finished implementation of UF settings change messaging and handling.

  1. … 3 more files in changeset.
Further implementation of UF settings change messaging and handling. Code review fixes.

  1. … 17 more files in changeset.
Changed breadboard build switch to be more biased toward the new HD board. Added UF pause resume functions and message handling. Updated internal ADC voltage and conversion factors for new HD board. Commented out RTC POST until RTC on new HD board is configured.

  1. … 12 more files in changeset.
Added HD specific common header. Streamlined includes for Common.h and HDCommon.h.

  1. … 55 more files in changeset.