Done

Why are these changed back to ACK_NOT_REQUIRED?

Why are these changed back to ACK_NOT_REQUIRED?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
FALSE was correct (throughout) for last param of sendAckResponseMsg() if message/request found to be invalid in some way, TRUE if message is ok.

FALSE was correct (throughout) for last param of sendAckResponseMsg() if message/request found to be invalid in some way, TRUE if message is ok.

Timer variable is actually a time stamp of last command being queued - variable name could be clearer.

Timer variable is actually a time stamp of last command being queued - variable name could be clearer.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

For a release build, looks like we would do nothing and get stuck in this state.

For a release build, looks like we would do nothing and get stuck in this state.

Add doxygen comment for persistence. Try to capture thinking on why this is such a brief period.

Add doxygen comment for persistence. Try to capture thinking on why this is such a brief period.

HD-DEN-14497_DG HD Dev HD DG Dvt Update Part 7
HD-DEN-14497_DG HD Dev HD DG Dvt Update Part 7
Fixed. Thanks!

Fixed. Thanks!

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

The only time we would be adding more states is if we are no longer on HIGH/LOW control, like switching to PWM. That would be such a significant change that this would be reviewed.

The only time we would be adding more states is if we are no longer on HIGH/LOW control, like switching to PWM. That would be such a significant change that this would be reviewed.

Would rather you set a BOOL based on state and then pass the BOOL to persistent alarm function. I'm afraid we could add more states later and forget about this.

Would rather you set a BOOL based on state and then pass the BOOL to persistent alarm function. I'm afraid we could add more states later and forget about this.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.