DEN-15763: Changed clear persistence behavior for illegal air trap alarm. Additional refinements to UF control start up.

DEN-15763: Addressed code review comment.

DEN-15763: set DG mode to fault if DG comm is lost.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15763-dg-restarted-alarm-107-cannot-be-cleared' into develop

DEN-15763: Tuning of start of UF control per data.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15763-dg-restarted-alarm-107-cannot-be-cleared' into develop

DEN-15763: more updates to start of UF control.

DEN-15763: Addressed DEN-15779.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15763: Addressed DEN-15729.

DEN-15763: Addressed code review comment.

DEN-15763: Additional fix for DEN-15763 to allow post-tx to drain reservoirs if DG was restarted (alarm 107).

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15763: Clear preserved POR bit for logging purposes after logged.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

fixed merge conflict.

DEN-15763: Addressed DEN-15763. Increased max UI boot time from 40 sec to 2 min.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15609-fw-usability-dialysate-can-leak-out-side-of-dg-if-user-has-to-swap-cartridge-during' into develop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflict.

DEN-15764: Fixed issue with recent alarm clear change on resume.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15609 Allow reservoir fill only after dry selft test complete.