DEN-14860 Fill time calculation maximum calculation 10% extra margin. Move Depletion time calculation to work in all reservoir states.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14906-dialyzer-de-prime' into develop

DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

Auto stash before merge of "DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off" and "staging"

Merge branch 'staging' into DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off

DEN-14557 updated the code set the valves status to position unknown if the tranistion to a position timedout.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14906-dialyzer-de-prime' into develop

DEN-14906: Fix response to Dialin requiring ACK when it should not.

DEN-14906: restored 30 min interval for dialyzer re-prime.

Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14788: code review comments.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-14906: Increased pump speed during dialyzer re-prime and reduce time in re-prime state in attempt to address over-current alarm.

DEN-14999: Reduced BP occlusion clear threshold from 5500 to 5000 above baseline.

Merge branch 'staging' into DEN-14788-DIAL-242-air-bubble-sensor

    • -6
    • +134
    ./App/Services/SystemCommMessages.c
DEN-14788: fixed UI check-in.

DEN-14860 Fix config bug. If Dry tests disabled, got stuck in Prime because DG Allow Flush/Fill signals not set.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Updated signalAllowDGFlushFills(),

signalAllowDGFillRes[12]() to be sure that the signal to continue occurs even when both thr air pump and dry tests are disabled.

Aligned memory with sector boundaries

Aligned memory with sector boundaries

Aligned memory with sector boundaries

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo