- Removed unused imports - Updated end_pos field in temperature sensors

  1. … 1 more file in changeset.
D98 has been renamed

D98 has been renamed

No reference of conductivity here. it should be a level sensor.

No reference of conductivity here. it should be a level sensor.

like D63 level sensor, please add D98 level sensor throughout the units. Keep naming as D98_LEVEL ( instead of D98_COND)

like D63 level sensor, please add D98 level sensor throughout the units. Keep naming as D98_LEVEL ( instead of D98_COND)

END_POS_FIELD_19?

END_POS_FIELD_19?

Moved

Moved

- Moved D98 to levels from conductivity sensors

  1. … 2 more files in changeset.
Please move this to level sensor enum. this is not a conductivity sensor; it is a conductive Level sensor.

Please move this to level sensor enum. this is not a conductivity sensor; it is a conductive Level sensor.

- Added new sensors and valves to DD

  1. … 5 more files in changeset.
LEAHI-DIALIN-LDT-2998_Update Dialin to support any B1.9 FW and/or HW changes (higher priority)
LEAHI-DIALIN-LDT-2998_Update Dialin to support any B1.9 FW and/or HW changes (higher priority)
Done

Done

Done

Done

LDT-1886 updated the DD level sensors sync function

LDT-2656 updated the variable names

  1. … 3 more files in changeset.
I don't think we have motor current. We're getting torque instead. Just replace current with torque.

I don't think we have motor current. We're getting torque instead. Just replace current with torque.

I think we should replace motor current with torque instead of adding torque as new field.

I think we should replace motor current with torque instead of adding torque as new field.

LEAHI-DIALIN-LDT-1886_Blood Set Auto-Load and Auto-Eject - TD
LEAHI-DIALIN-LDT-1886_Blood Set Auto-Load and Auto-Eject - TD