LDT-736: moved set air pump from the driver level to the controller level so the driver FSM doesnt stay in OFF state. Added ADC_Dummy to prevent voltage alarms.

LDT-736: comment out blood prime transition until implemented.

LDT-736: use off button as a stop button while stop button not implemented in CPLD.

LDT-736: alpha 1 integration changes and some new build switches to support running w/o DD and w/o functioning pinch valves.

LDT-736: Added support for alarm list request. Fixed a few alarms.

Bamboo Commit: Updated TDCommon.h with build versions from Bamboo.

Fixed merge conflict w/ version.

Bamboo Commit: Updated TDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated TDCommon.h with build versions from Bamboo.

LDT-469: Merged 2 received message look-up tables into 1 to reduce chances of the 2 tables getting out of sync.

LDT-469: Tuned air pump PWM levels.

LDT-469: ifdef a temporary fix to level sensors.

LDT-469: Added some UT for tx param mode. Added a back off to retract/home operations.

LEAH-282: Added integration tests for Timers, Utilities, and WatchdogMgmt

LDT-469: Alpha1 integration changes.

LDT-469: Updates to ejector functionality. Started Tx Params mode unit testing.

LDT-469: Removed HET edge interrupt - no longer needed for rotor hall sensors. Updated BP rotor speed function.

LDT-469: Updated air pump driver.

LDT-469: Addressed code review comments.

LDT-469: Fixed typo.

LDT-469: minor fixes.

LDT-469: Added ejector unit and message support.

    • -0
    • +452
    ./App/Controllers/Ejector.c
    • -0
    • +71
    ./App/Controllers/Ejector.h
    • -0
    • +262
    ./App/Drivers/EjectorMotor.c
    • -0
    • +47
    ./App/Drivers/EjectorMotor.h
  1. … 3 more files in changeset.
LDT-469: Updated blood pump homing function. Added dialyzer volumes table to Tx params unit.

LDT-469: Started UT for 3way valve driver.

LDT-469: Added Heparin type Tx param.

LDT-469: Increased max dialysate temperature setting from 37 to 38 deg C.

LDT-469: Fixed min/max for rinseback volume Tx param.

LDT-469: Updated treatment parameters list. Implemented Dialin message support. Added support to respond to request for version information.

Bamboo Commit: Updated TDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated TDCommon.h with build versions from Bamboo.