DEN-11098 Resolved merge conflicts

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Removed an dead alarm

DEN-11098 Removed code related to run blood circuit again because it no longer needed

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 98 more files in changeset.
DEN-11098 Addressed code review comments

DEN-11098 Updated prime time limit for the saline dialyzer state

DEN-11098 Converted from seconds to milliseconds fix

DEN-11098 Corrected state transition and changed blood pump speed from 400 to 300 per system specification

DEN-11543: Unable proceed to the Confirm Treatment when "08-3251-9" Acid Concentrate is selected

The HD FW updated to accept up to 4 acid concentration type index.

Below is the UI DataList.conf for acid concentrate.

[Acid Concentrate]

Fres. Naturalyte

08-1251-1

08-2251-0

08-3251-9

DEN-11098 Incorporated peer review comments

DEN-11098 Updated code per peer review comments

DEN-11098 Updated time out limits

DEN-11098 Implemented code update due to requirements change

Finishing up the story for firmware

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

DEN-11098 Corrected delta volume calculationbased on the past 10 seconds value instead of the past 9 seconds

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Fixed DG_RESERVOIR to DG_RESERVOIRS

Merge branch 'DEN-10153_Treatment_Log_Prescribed_UF' into staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-11098 Implemented steady state volume check

DEN-11098 Updated steady state volume detection logic

DEN-11098 Updated header

DEN-11098 Corrected R1 and R2 in bypass state and dialyzer state

DEN-10247: UI Acknow timeouts on HD side, due to the delay of almost 2 seconds.

- Added another timeout of 5000 ms ( same value as UI_COMM_TIMEOUT_IN_MS ) only for the INIT state in which UI has a delayed response.