UI-DEN-12182: UI-DEN-12182-2

- During code review of my code, cleaned up the code.

  1. … 6 more files in changeset.
DEN-10153: Treatment Log Comments Investigations

- Added the missing prescribed UF Volume, Rate to the log and review.

  1. … 2 more files in changeset.
DEN-12180: INT: Alarm dialog body space [RC: 15*80]

- reduced the bottom margin of the Alarm Dialog bottoms.

- removed the space from the legacy files which caused issue with DevOps scripts.

    • -0
    • +577
    ./VTreatmentCreate(legacy).cpp
    • -0
    • +258
    ./VTreatmentCreate(legacy).h
  1. … 2 more files in changeset.
DEN-10974: Syringe pump installation screen allows user to proceed before syringe installed

- Added the response message 0xAD00(173), for the request message of 0x6200(98):Pre-treatmetn disposables confirm.

  1. … 10 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 570 more files in changeset.
Updated the DG POST index like the HD

  1. … 4 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 12 more files in changeset.
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- A typo caused the BP/HR entry timer disabled.

  1. … 1 more file in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-10422: Dialysate Flow Rate maximum rate selectable in Treatment does not meet PRS resolution

DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- Added a script to later be used for the log export on the device, but it now only works on the VM since the device doesn't have the zip tool.

- Added a mimic function to simulate the BCuff measurement received and assigned F11 to it and it will only be enabled if the macro DEBUG_BCUFF_MIMIC is defined.

- Added opened and closed signals to the EntryDialog.qml to make it do the auto focus on the firstInput object when opened.

- Added setFocus() function in the BPHREntry.qml

- Removed the enabled option from doReset of the VTreatmentVitals

- And created two separate enableDialog for vital entry, and enable update for the auto update the properties.

- Added a constructor with systolic, diastolic, heartRate parameter to set the data on creation to be used in the message mimic.

- Added systolic, diastolic, heartRate parameters to the doTrigger of the VCommonAdjustmentVitals to open the vitals entry dialog with default values.

    • -3
    • +4
    ./hd/adjustment/common/VCommonAdjustmentVitals.h
  1. … 13 more files in changeset.
Code Cleanup

- DEN-11807: Treatment Confirmation - Rinseback Rate displays "min" instead of "mL/min"

- Renamed the Keyboard component name to VirtualKeyboard to resolve the InputPanel's Keyboard object/type conflict (There was no issue is was only a name conflict.)

- Preventing the Patient Id from being empty space string, and disabling the confirm in that case.

- Mode the create_update_folder.sh to the Scripts/local repository's folder.

- Added Qt logo and Diality Qt Colored log.

- Updated the alarmMapping.sh script to always checkout to common/staging and also print out the log on console.

-

  1. … 12 more files in changeset.
Finishing up the story for Firmware and adding remaining other tasks

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

DEN-11320: KeborMed [ CloudSync Development ]

- Added the message handling for the Device status

- Adding Error handling

  1. … 8 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

  1. … 8 more files in changeset.
DEN-10205: BCuff [Missing Feature]: Try to reconnect to the previously paired

- Stop try to reconnect while still is trying the initial connect.

- Changed the BCuff wait state notification.

- Updated the Heparin section wording.

  1. … 4 more files in changeset.
DEN-10205: BCuff [Missing Feature]: Try to reconnect to the previously paired

Bluetooth Code improvements:

- stopping the agent from scanning if user selects a device to connect.

- Rewording of the parred device not found

- Added "ZERO" check for the address

- Added real-time Bluetooth status on top of the main screen.

- Not The UI will always updated but the TreatmentLog still needs UI to be in a proper screen/state to log the Vitals measurements.

Other Requested modifications:

Chris.Y:

- Added back for the pre-treatment BP/HR screen to the Ultrafiltration screen.

Eliza.P:

- Added notification for the Heparin Dispensing:"Heparin Active"

- Changed the button text to "OFF" when the heparin is off.

- Changed the button text to "COMPLETE" when the heparin is complete.

- Shortened the notification text and removed the "Heparin ... " from the beginning.

- Won't show the off notification since the button text changed to imply that.

  1. … 9 more files in changeset.
Code Review: UI-DEN-10206: UI-DEN-10206-1

  1. … 3 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 10 more files in changeset.
UI-DEN-10206: UI-DEN-10206-1: addressing the code reviews.

  1. … 8 more files in changeset.
DEN-10607: Implementation [Screen]

- Got the information below from the settings to be consistent with all the other screens and views:

- acidConcentrateOptions

- bicarbonateConcentrateOptions

- dialyzerTypeOptions

- heparinTypeOptions

- Added the heparin to the list as well so later can be used on the create treatment screen.

DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

  1. … 6 more files in changeset.
Improved the POST

- Updated the run.sh

- Enabled the Sha256 in the application

- Fixed the ShaSum issue in the run.sh

- Added a command line argument to disable the alarm-no-minimize

- Set the timezone to UTC int eh setup script.

- Setup the correct greeting on the home screen

  1. … 9 more files in changeset.
DEN-10206: BCuff: Known bugs: Only pair with one BCuff at a time

- Updated some BLE device status messaging due to received feed backs to avoid user confusion.

- Updated to emit the BLE paired devices signal in Device view and let UI decide to run the scan or not to have the failed information in UI as well in case needs to be shown

  1. … 2 more files in changeset.
DEN-10206: BCuff: Known bugs: Only pair with one BCuff at a time

- Added the bluetooth_remove_paired.sh script call

- Added the timeout safety for the script calls.

- Filtered the UI acknowledgment on only the messages intent for the UI only.

  1. … 12 more files in changeset.
DEN-10599: KeborMed [ Start working on the CloudSync infrastructure ]

- Updated/Improved CloudSync

- Figured the gedit changes the file in a way that breaks the Qt watch.

. It has been tested with Python file O/R/W/F/C and all worked fine.

. It has been tested by nano and works fine.

. It has been tested with shell echo command and works fine with >,>> .

  1. … 7 more files in changeset.
PRS 464 Update: Treatment Parameter, Stop Heparin Dispensing Before End of Treatment

- const quint32 heparinStopTime_Max = 470 <- 480

DEN-9968: GUI does not indicate the amount of time the device has been stopped in pre-treatment & treatment

- Added message ID_TreatmentStop = 0x4900, // 73 and its Model, View

- Added a timer on the Alarm Dialog which is visible in Treatment Stop and if the message above has a total value.

    • -0
    • +26
    ./hd/data/treatment/VTreatmentStopData.cpp
    • -0
    • +44
    ./hd/data/treatment/VTreatmentStopData.h
  1. … 12 more files in changeset.
DEN-10399: UI: Alarms block Shutdown screen

DEN-9271: Dialysate Temperature implemented resolution does not meet PRS

  1. … 5 more files in changeset.
Code Review Comments

- QApplication typo

- Added a UI version of the Log Macros.

  1. … 13 more files in changeset.