Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'staging' into DEN-12974-miscellaneous-flush-issues_bb

# Conflicts:

# firmware/App/Controllers/TemperatureSensors.c

  1. … 6 more files in changeset.
DEN-12974 Minor correction.

DEN-12974 Changes per code review. Remove Chemical and Heat Disinfect code. Add separate RO_FLO_MIN_LPM for heaters and uv.

  1. … 3 more files in changeset.
DEN-12974 Updates per code review.

DEN-12974 DEN-13558 Update to fill status functions to support V&V

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 26 more files in changeset.
DEN-12931 addressed the code review comments

  1. … 4 more files in changeset.
DEN-12931 addressed the code review comments

  1. … 10 more files in changeset.
DEN-12974 DEN 13558 Prototype new timer design for drain pump and reservoir status

DEN-12974 DEN 13558 Prototype new timer design for drain pump and reservoir status

DEN-12974 DEN-13556 Correct logic per SRSDG 333.

DEN12974 DEN-13558 Do not reset stateTimer in FlushR2AndDrainR1. Same issue will be in ModeChemicalDisinfect.

DEN-12974 DEN-13560 Do not reset stateTimer in FlushR1AndR2.

DEN-13047 haveDrainParmsBeenInit now any array with one entry for each reservoir. DEN-13534 rsrvrFillStableTimeCounter now an array with one entry for each reservoir. May be causing intermittent functionality (test_dg_flush_R1_to_R2_state_R2_filled_before_R1). DEN-12974, DEN-13425 Restore logic around resetting state timer for getRsrvrFillStatus.

DEN-12974, DEN-13549 Correct reservoir volume logic to match requirements. Changed ">" to ">=" in several modes when checking for R1 being filled before R2

DEN-12931 updated the dialysate flow sensor and made switching to service mode legal from solo mode

  1. … 1 more file in changeset.
DEN-12974, DEN-13425 Remove logic around resetting state timer for getRsrvrFillStatus.

DEN-12974

DEN-12974, DEN-13425 Unable to set default arguement.

DEN-12974, DEN-13425 Correct bool to BOOL

DEN-12974, DEN-13425 Do not reset state timer in FlushR2DrainR1

DEN-12931 updated the drivers with the DVT changes. Added code to run RTC and NVDataMgmt POSTs at fault mode

  1. … 8 more files in changeset.
DEN-12931 updated the drviers with DVT changes

  1. … 16 more files in changeset.
DEN-12974, DEN-13078 correct setting of VPo when entering handleFlushModeDrainR2State.

DEN-12931 updated the heaters driver and DVT

  1. … 13 more files in changeset.
DEN-12974, DEN-13078 correct setting of VPo when entering Drain_R1 and Drain_R2

Revert "DEN-13047 Implement unique haveDrainParamsBeenInit for both reservoirs."

This reverts commit 81872acd6f2ce222c6b9a9b93f65b7cd3665f28f.

  1. … 2 more files in changeset.
DEN-12931 DVT updates and SRS changes

  1. … 6 more files in changeset.
DEN-13047 Implement unique haveDrainParamsBeenInit for both reservoirs. DEN-12882 Unique minimum ro flow rate for uv reactors.

  1. … 2 more files in changeset.
DEN-12549 Correct state of valves before entering DRAIN_R1 from FLUSH_R2AND_DRAIN_R1.