DEN-14007 merged staging and resolved the merge conflicts

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 18 more files in changeset.
DEN-14307: Removed obsolete build switches.

  1. … 5 more files in changeset.
DEN-14307 addressed code review comments

  1. … 2 more files in changeset.
DEN-14007 Code Review comments update 2.

DEN-14007 Code Review comments update 1.

  1. … 3 more files in changeset.
DEN-14307 updated the code to get the heat disinfect temperature sensors (THd) and checks it value prior to starting a treatment

  1. … 3 more files in changeset.
Merge branch 'staging' into DEN-14307-dg_hd_dev-hd_dg_dvt-update-part-6

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-14458 addressed a code review comment

DEN-14458 addressed the code review comments

  1. … 3 more files in changeset.
DEN-14307 added more treatment blockers and added a message to override the NV record CRC. Changed some of the NV record alarms to software fault

  1. … 4 more files in changeset.
DEN-14458 cleaned up the code

DEN-14458 cleaned up the code and removed the commented lines

DEN-14464 updated a code with a possible fix to make sure the active reservoir starts with 1500 mL upon transitioning to treatment from pretreatment

DEN-14470 updated the dry self test to be able to command the fill in pretreatment if the syringe pump is enabled

DEN-14458 lowered the self tests' target fill volume

DEN-14417 reversed the direction of the reservoirs' displacement in the second displacement in the wet self test

DEN-14458 updated the wet self test events payload

DEN-14464 changed the fill targets in pre-treatment to 1100 mL and changed the directions of the reservoir's fill from R1 to R2 to R2 to R1 in the wet self tests second displacement

DEN-14458 fixed the post treatment reservoir management

DEN-14458 updated the code

DEN-14458, DEN-14417 updated the wet self test displacement check

DEN-14458 updated the post treatment reservoir management

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14007 De-prime DG lines in Post Treat mode before cartridge removal.

DEN-14344: code review comments.

  1. … 2 more files in changeset.
DEN_14252: Addressed code review comment.

DEN-14252: Another fix to pre-tx reservor mgmt state machine.

DEN-14252: Additional fix to reservoir mgmt state machine.