Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14007 merged staging and resolved the merge conflicts

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DIAL-219: Fixed bug in broadcast user alarm button blockers.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-14307: partial fix to DEN-14086.

DEN-14307 addressed a code review comment

DEN-14307 addressed the code review comments

DEN-14307: Removed obsolete build switches.

DEN-14307 addressed code review comments

DEN-14007 Code Review comments update 2.

DEN-14007 Code Review comments update 1.

DEN-14307: Fixed polarity issue with reported state based alarm button blockage. Renamed battery capacity remaining variable to correct units. Stop disabling battery comm on build switch.

DEN-14307 updated the code to get the heat disinfect temperature sensors (THd) and checks it value prior to starting a treatment

DEN-14307: HD rotor vs. motor check - changed RPM threshold from 5 to 2 RPM.

Merge branch 'staging' into DEN-14307-dg_hd_dev-hd_dg_dvt-update-part-6

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14458 addressed a code review comment

DEN-14458 addressed the code review comments

DEN-14307 added more treatment blockers and added a message to override the NV record CRC. Changed some of the NV record alarms to software fault

DEN-14458 cleaned up the code

DEN-14307: Added alarm response button block flags to alarm info broadcast message.

DEN-14007 Update all pump speed checks for proper forward / reverse direction checking.

DEN-14458 updated the air trap's broadcast interval

DEN-14458 cleaned up the code and removed the commented lines

DEN-14464 updated a code with a possible fix to make sure the active reservoir starts with 1500 mL upon transitioning to treatment from pretreatment

DEN-14470 updated the dry self test to be able to command the fill in pretreatment if the syringe pump is enabled

DEN-14458 lowered the self tests' target fill volume