Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15101: disable ac power loss alarms until cpld change gets propagated.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
Merge branch 'staging' into DEN-14999-alarm-table-updates-part-2

DEN-14999-2: Restored a/c power loss alarm detection.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14999 updated the alarms table

  1. … 6 more files in changeset.
DEN-14999-2: Addressed code review comment.

DEN-14830: fixed alarm mgmt faulting on clear.

DEN-14999-2: Temporarily disable AC power loss alarm until CPLDs are updated in lab.

DEN-14830: Fixed alarm mgmt not clearing alarm ids under 30. Fixed Dial temps

DEN-14830: fixes

DEN-14830: simplifying temperature control logic.

DEN-14999-2: Increased UF alarm thresholds to 100 mL for volume and 10 mL/min for rate.

DEN-14999-2: Fixed bug in UF rate alarm detection.

DEN-14999 updated the code based on the alarms table

DEN-14999-2: Removed alarm escalation feature.

DEN-14830: compiler fixes.

DEN-14830: Reworked dialysate temperature logic so persistence triggers do not cancel each other out. Fixed air pump not turning off when dry self tests are user stopped.

DEN-14999-2: prevent treatment resumption from rinseback/recirc states if an alarm that blocks resume has been triggered previously in the treatment.

DEN-14999-2: ignore venous occlusion alarm if venous pressure alarms disabled.

DEN-14830: removed 2 dialysate temperature alarms.

DEN-14830: PRS alignments

Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

DEN-14999 removed the depletion alarm (alarm 168)

DEN-14999-2: Removed alarms 214, 215 and 217.

DEN-14999-2: Shortened a few alarm enum names.