Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15556: Addressed code review comments.

DEN-15556: Fix to last.

DEN-15556: Stabilize pressures after UF rate change.

  1. … 3 more files in changeset.
DEN-15556: Addressed issues for DEN-15556 re: not repeating used cartridge test and openeing VBA after door closed after cartridge install to prevent saline from leaking into the line.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-15763: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-15763: increased timeout period for dialysate prime state of pre-tx.

DEN-15763: Addressed DEN-15779.

DEN-15763: Addressed DEN-15729.

DEN-15763: Additional fix for DEN-15763 to allow post-tx to drain reservoirs if DG was restarted (alarm 107).

  1. … 2 more files in changeset.
DEN-15763: Clear preserved POR bit for logging purposes after logged.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15609 Allow reservoir fill only after dry selft test complete.

DEN-15754: Addressed DEN-15751 re: max additional rinseback volume.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15637: bypass disinfect in standby mode after post-tx mode if treatment was ended in pre-tx mode.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15282: Changed rinseback volume from 400 to 350 mL per Systems/Clinical input.

  1. … 1 more file in changeset.
DEN-15282: Reverted pressure stabilization time back to 60s from 30s. During rinseback run and paused states, allow resume and end tx options for alarms - resume will resume rinseback.

  1. … 1 more file in changeset.
DEN-15282: increased timeout for rinseback saline bolus to 25 sec from 20 sec.

DEN-15282: Fixed rinseback bolus rate sent to UI for display.

DEN-15510: Fixed RB volume at 400 mL. Changed RB bolus volume from 10 to 50 mL.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged staging to DEN-15656.

DEN-15656: publish 200 rate for rinseback additional.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15656: Addressed code review comment.

DEN-15656: Addressed DEN-15654 issue where pressure limits did not go to stabilization limits on transition to treatment paused state.

DEN-15656: Addressed DEN-15656 issue where pressure stabilization would not occur on alarm resume if blood was recirculating during alarm. Addressed DEN-15652 where rinseback saline bolus was delivered at 50 mL/min instead of 200 mL/min.