DEN-15763: Addressed code review comments.

DEN-15763: increased timeout period for dialysate prime state of pre-tx.

DEN-15763: adjusted timing of uf rate calculation to clean up logging issues.

DEN-15763: Changed clear persistence behavior for illegal air trap alarm. Additional refinements to UF control start up.

DEN-15763: Addressed code review comment.

DEN-15763: set DG mode to fault if DG comm is lost.

DEN-15763: Tuning of start of UF control per data.

DEN-15763: more updates to start of UF control.

DEN-15763: Addressed DEN-15779.

DEN-15763: Addressed DEN-15729.

DEN-15763: Addressed code review comment.

DEN-15763: Additional fix for DEN-15763 to allow post-tx to drain reservoirs if DG was restarted (alarm 107).

DEN-15763: Clear preserved POR bit for logging purposes after logged.

DEN-15763: Addressed DEN-15763. Increased max UI boot time from 40 sec to 2 min.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15764: Fixed issue with recent alarm clear change on resume.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15609 Allow reservoir fill only after dry selft test complete.

DEN-15754: Addressed DEN-15751 re: max additional rinseback volume.

DEN-15754: Addressed DEN-15754 issue.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15637: Fix for DEN-15654.

DEN-15637: bypass disinfect in standby mode after post-tx mode if treatment was ended in pre-tx mode.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15282: Changed rinseback volume from 400 to 350 mL per Systems/Clinical input.