Merged staging to DEN-15788.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15788: Increased open loop DPo start rate for dialysate/dialyzer prime state.

DEN-15788: open VBA and VBV when checking for atmosphere pressure before pressure leak test.

DEN-15788: change DPo control during dialysate/dialyzer prime state to begin immediately instead of after fixed prime time.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15788: Added 10 sec venous pressure filter and publish venous pressure readings with 10 sec filter. Use 10 sec venous pressure filter for setting pressure windows after stabilization. Fix proposed UF rate and Tx duration after a UF volume change. Add simple pressure control to DPo speed during dialysate/dialyzer prime state to prevent pressure build up.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15660 addressed a code review comment

DEN-15628 changed the blood leak zero sequence counter from Z to G since Z is repeated twice in each zero sequence

DEN-15660, DEN-15721 added code to check the blood detection in treatment and service mode both

Merge branch 'staging' into DEN-15660-test_configuration_treatment_recovery

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15556: Addressed code review comments.

DEN-15556: Fix to last.

DEN-15556: Stabilize pressures after UF rate change.

DEN-15556: Addressed issues for DEN-15556 re: not repeating used cartridge test and openeing VBA after door closed after cartridge install to prevent saline from leaking into the line.

DEN-15660, DEN-15809 updated the code

DEN-15660 uncommented the code

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15763: Addressed code review comments.

DEN-15763: increased timeout period for dialysate prime state of pre-tx.

DEN-15763: adjusted timing of uf rate calculation to clean up logging issues.

DEN-15660 disabled alarm 63 for testing

DEN-15763: Changed clear persistence behavior for illegal air trap alarm. Additional refinements to UF control start up.

DEN-15763: Addressed code review comment.

DEN-15763: set DG mode to fault if DG comm is lost.