Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./components/SliderCreateTreatment.qml
    • -1
    • +1
    ./pages/settings/SettingsExportLogs.qml
  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-16656: UI: No Treatment log export success is displayed when exporting logs to a USB

added the export status to the notification text.

- started (in progress)

    • -1
    • +6
    ./pages/settings/SettingsExportLogs.qml
DEN-16491: UI: During Rinseback setup, user cannot go back to treatment until they have viewed all Rinseback setup instructions

- Removed the restriction for Back Button on the Rinseback Setup screen, to let user go back without going through the entire instruction.

DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

Added the functionality to the TreatmentCreate to reset the scroll to top on clear the parameters.

DEN-16444: UI: UI Logs do not Capture TDi and TRo Moving Average Values

Updated the application to incorporate the two new float values in the DG temperature sensors payload.

  1. … 6 more files in changeset.
DENBUG-183: Ultrafiltration resolution is not 0.1 when sliding from 0.0-0.1

DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

- After fixing the issue it was noticed that the border of the sliders with OFF button will remain highlighted.

    • -1
    • +1
    ./components/SliderCreateTreatment.qml
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./components/ConfirmTreatmentSubTable.qml
    • -1
    • +1
    ./components/ConfirmTreatmentTable.qml
    • -2
    • +2
    ./components/ConfirmTreatmentTableEntry.qml
  1. … 598 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 1 more file in changeset.
DENBUG-204 aligned the commented out code in the diagnostics qml

DENBUG-204 updated the diagnostics screens qml with the latest dialysate outlet pump broadcast changes

DENBUG-182: Phantom Alarm Screen Appears After Clearing Last Alarm

- the order of the QML PROPERTYs cannot exactly predicted therefore setting them if not needed has to be avoided.

- set the alarmID to 0 to trigger the next change since we skip the property updated if the alarmID is 0.

  1. … 2 more files in changeset.
DEN-12728: UI: UI does not clean up Treatment Parameter settings for different patient

- the reset is added to the treatment parameters

- the foundation of the pre-population has been added but still there is parts need to be added to store the sliders' On/Off switch to be restored which is needed for the slider to set properly.

    • -0
    • +11
    ./components/SliderCreateTreatment.qml
DENBUG-179, DENBUG-79: Export dialog files count issues.

DENBUG-179: Export Logs text does not update after Factory Reset

DENBUG-79: UI: SD-Card file count increments when scrolling up and down the log file window

OPEN

    • -11
    • +7
    ./pages/settings/SettingsExportLogs.qml
DENBUG-136: Main treatment screen fails to prompt user with vital measurements entry after set interval

- the issue cannot be reproduced.

- added a count down timer for the vital entry to better understand how and when the timer works on each interval.

    • -0
    • +14
    ./pages/treatment/TreatmentHome.qml
  1. … 2 more files in changeset.
CS-3: CS: Heparin OFF o the HCP portal Treatment report

- Fixed some issues in the create treatment screen on heparin type

- Combined the separated values for log and GUI into one values for both.

- Added the functionality for the TreatmentConfirm (Pre-Tx), TreatmentReview (Ppost-Tx), TreatmentReport (log) for the heparin values.

  1. … 4 more files in changeset.
CS-5: @ UI: TxCode generation on device

~ Added QRCode for the TxCode

- Added a Third-Party Open-Source MIT Licence QRCode generator from the following link:

* Copyright (c) Project Nayuki. (MIT License)

* https://www.nayuki.io/page/qr-code-generator-library

- Created a QRCode component

- Added the QRCode generation functionality in the GuiView

- Added QRCode for both the Pre-Tx-Confirmation (UI Generated) and Post-Tx-Review (CS received)

- Moved the CS TxCode Receive wait "..." from C++ to QML to keep the TxCode empty while wait to avoid displaying the "..." as a QRCode.

- Animated the CS TxCode Receive wait "..."

~ Added development testability for the CloudSync communication

~ Fixed the date change buf file watch to wait for the CS to get the checkin and create the out buf file.

    • -0
    • +61
    ./components/QRCode.qml
  1. … 10 more files in changeset.
Addressing code review comments

Addressing Code Review comments

DENBUG-182 , DENBUG-185

DENBUG-185: UI shows version without period in between (ECO-11648, ECO-11616)

DENBUG-182: Phantom Alarm Screen Appears After Clearing Last Alarm

  1. … 1 more file in changeset.
CS-5: UI: TxCode generation on device

    • -1
    • +1
    ./components/ConfirmTreatmentTableEntry.qml
  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./components/SliderCreateTreatment.qml
    • -1
    • +1
    ./pages/settings/SettingsDecommission.qml
    • -1
    • +1
    ./pages/settings/SettingsFactoryReset.qml
  1. … 10 more files in changeset.
DEN-16450: UI: Adjustment arrows on the UI sliders set the initial minimum value to OFF for heparin bolus, heparin rate, and blood pressure interval

    • -0
    • +13
    ./components/SliderCreateTreatment.qml
DEN-16435: Factory Reset deletes service password from System.conf

- Fixed the Confirmation Dialog confirm call which caused both the Decommissioning and Factory Reset to be called.

- Fixed the PlantUML section of the VDuetRoWaterDG class.

    • -1
    • +3
    ./pages/settings/SettingsDecommission.qml
    • -1
    • +3
    ./pages/settings/SettingsFactoryReset.qml
  1. … 1 more file in changeset.
DEN-16256: Updated - UI.P1B : Display UF Rate

- minor clean up, layout fix and typo fixes

  1. … 1 more file in changeset.
DEN-16443: UI: The concentration units shall be displayed as mEq/L

- Concentrations unit typo fix.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./pages/settings/SettingsExportLogs.qml
    • -1
    • +1
    ./pages/treatment/TreatmentBloodPrime.qml
  1. … 12 more files in changeset.
DEN-16285: UI: Cybersecurity: Restrict the SD-Card Symbolic Link usage

    • -1
    • +1
    ./pages/settings/SettingsExportLogs.qml
  1. … 3 more files in changeset.
DEN-15626: UI: Arterial and Venous Pressure Monitor Limits

    • -1
    • +0
    ./pages/treatment/TreatmentBloodPrime.qml